51
|
Arthur DTJ, Khan MM. Quantitative deconvolution of human thermal infrared emittance. IEEE J Biomed Health Inform 2012; 17:205-13. [PMID: 23086533 DOI: 10.1109/titb.2012.2225108] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/08/2022]
Abstract
The bioheat transfer models conventionally employed in etiology of human thermal infrared (TIR) emittance rely upon two assumptions; universal graybody emissivity and significant transmission of heat from subsurface tissue layers. In this work, a series of clinical and laboratory experiments were designed and carried out to conclusively evaluate the validity of the two assumptions. Results obtained from the objective analyses of TIR images of human facial and tibial regions demonstrated significant variations in spectral thermophysical properties at different anatomic locations on human body. The limited validity of the two assumptions signifies need for quantitative deconvolution of human TIR emittance in clinical, psychophysiological and critical applications. A novel approach to joint inversion of the bioheat transfer model is also introduced, levering the deterministic temperature-dependency of proton resonance frequency in low-lipid human soft tissue for characterizing the relationship between subsurface 3D tissue temperature profiles and corresponding TIR emittance.
Collapse
|
52
|
Bruggmoser G, Bauchowitz S, Canters R, Crezee H, Ehmann M, Gellermann J, Lamprecht U, Lomax N, Messmer M, Ott O, Abdel-Rahman S, Schmidt M, Sauer R, Thomsen A, Wessalowski R, van Rhoon G. Guideline for the clinical application, documentation and analysis of clinical studies for regional deep hyperthermia. Strahlenther Onkol 2012; 188 Suppl 2:198-211. [DOI: 10.1007/s00066-012-0176-2] [Citation(s) in RCA: 74] [Impact Index Per Article: 5.7] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 02/04/2023]
|
53
|
Kickhefel A, Rosenberg C, Roland J, Viallon M, Gross P, Schick F, Hosten N, Salomir R. A pilot study for clinical feasibility of the near-harmonic 2D referenceless PRFS thermometry in liver under free breathing using MR-guided LITT ablation data. Int J Hyperthermia 2012; 28:250-66. [PMID: 22515346 DOI: 10.3109/02656736.2012.670834] [Citation(s) in RCA: 20] [Impact Index Per Article: 1.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/12/2023] Open
Abstract
OBJECTIVES The conventional implementations of proton resonance frequency shift (PRFS) magnetic resonance thermometry (MRT) require the subtraction of single or multiple temporal references, a motion sensitive critical feature. A pilot study was conducted here to investigate the clinical feasibility of near-harmonic two-dimensional (2D) referenceless PRFS MRT, using patient data from MR-guided laser ablation of liver malignancies. METHODS PRFS MRT with respiratory-triggered multi-slice gradient-recalled (GRE) acquisition was performed under free breathing in six patients. The precision of the novel referenceless MRT was compared with the reference phase subtraction. Coupling the referenceless MRT with a model-based, real-time compatible regularisation algorithm was also investigated. RESULTS The precision of MRT was improved by a factor of 3.3 when using the referenceless method as compared to the reference phase subtraction. The approach combining referenceless PRFS MRT and model-based regularisation yielded an estimated precision of 0.7° to 2.1°C, resulting in millimetre-range agreement between the calculated thermal dose and the 24 h post-treatment unperfused regions in liver. CONCLUSIONS The application of the near-harmonic 2D referenceless MRT method was feasible in a clinical scenario of MR-guided laser-induced thermal therapy (LITT) ablation in liver and permitted accurate prediction of the thermal lesion under free breathing in conscious patients, obviating the need for a controlled breathing under general anaesthesia.
Collapse
Affiliation(s)
- Antje Kickhefel
- Diagnostic and Interventional Radiology, Eberhard Karls University, Tübingen, Germany.
| | | | | | | | | | | | | | | |
Collapse
|
54
|
Nikitin SM, Khokhlova TD, Pelivanov IM. Temperature dependence of the optoacoustic transformation efficiency in ex vivo tissues for application in monitoring thermal therapies. JOURNAL OF BIOMEDICAL OPTICS 2012; 17:061214. [PMID: 22734744 DOI: 10.1117/1.jbo.17.6.061214] [Citation(s) in RCA: 11] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/08/2023]
Abstract
The calibration dependencies of the optoacoustic (OA) transformation efficiency on tissue temperature are obtained for the application in OA temperature monitoring during thermal therapies. Accurate measurement of the OA signal amplitude versus temperature is performed in different ex vivo tissues in the temperature range 25°C to 80°C. The investigated tissues were selected to represent different structural components: chicken breast (skeletal muscle), porcine lard (fatty tissue), and porcine liver (richly perfused tissue). Backward mode of the OA signal detection and a narrow probe laser beam were used in the experiments to avoid the influence of changes in light scattering with tissue coagulation on the OA signal amplitude. Measurements were performed in heating and cooling regimes. Characteristic behavior of the OA signal amplitude temperature dependences in different temperature ranges were described in terms of changes in different structural components of the tissue samples. The accuracy of temperature reconstruction from the obtained calibration dependencies for the investigated tissue types is evaluated.
Collapse
Affiliation(s)
- Sergey M Nikitin
- International Laser Center, M.V. Lomonosov Moscow State University, Moscow, Russia
| | | | | |
Collapse
|
55
|
Foxley S, Fan X, River J, Zamora M, Markiewicz E, Sokka S, Karczmar GS. Hyperthermically induced changes in high spectral and spatial resolution MR images of tumor tissue--a pilot study. Phys Med Biol 2012; 57:2653-66. [PMID: 22504096 DOI: 10.1088/0031-9155/57/9/2653] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/06/2023]
Abstract
This pilot study investigated the feasibility of using MRI based on BOLD (blood-oxygen-level-dependent) contrast to detect physiological effects of locally induced hyperthermia in a rodent tumor model. Nude mice bearing AT6.1 rodent prostate tumors inoculated in the hind leg were imaged using a 9.4 T scanner using a multi-gradient echo pulse sequence to acquire high spectral and spatial resolution (HiSS) data. Temperature increases of approximately 6 °C were produced in tumor tissue using fiber-optic-guided light from a 250 W halogen lamp. HiSS data were acquired over three slices through the tumor and leg both prior to and during heating. Water spectra were produced from these datasets for each voxel at each time point. Time-dependent changes in water resonance peak width were measured during 15 min of localized tumor heating. The results demonstrated that hyperthermia produced both significant increases and decreases in water resonance peak width. Average decreases in peak width were significantly larger in the tumor rim than in normal muscle (p = 0.04). The effect of hyperthermia in tumor was spatially heterogeneous, i.e. the standard deviation of the change in peak width was significantly larger in the tumor rim than in normal muscle (p = 0.005). Therefore, mild hyperthermia produces spatially heterogeneous changes in water peak width in both tumor and muscle. This may reflect heterogeneous effects of hyperthermia on local oxygenation. The peak width changes in tumor and muscle were significantly different, perhaps due to abnormal tumor vasculature and metabolism. Response to hyperthermia measured by MRI may be useful for identifying and/or characterizing suspicious lesions as well as guiding the development of new hyperthermia protocols.
Collapse
Affiliation(s)
- Sean Foxley
- Department of Radiology, University of Chicago, Chicago, IL 60637, USA
| | | | | | | | | | | | | |
Collapse
|
56
|
In Vivo Magnetic Resonance Spectroscopic Imaging and Ex Vivo Quantitative Neuropathology by High Resolution Magic Angle Spinning Proton Magnetic Resonance Spectroscopy. ACTA ACUST UNITED AC 2012. [DOI: 10.1007/7657_2011_31] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 04/12/2023]
|
57
|
Stafford RJ, Shetty A, Elliott AM, Schwartz JA, Goodrich GP, Hazle JD. MR temperature imaging of nanoshell mediated laser ablation. Int J Hyperthermia 2012; 27:782-90. [PMID: 22098362 DOI: 10.3109/02656736.2011.614671] [Citation(s) in RCA: 22] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/22/2022] Open
Abstract
Minimally invasive thermal therapy using high-power diode lasers is an active area of clinical research. Gold nanoshells (AuNS) can be tuned to absorb light in the range used for laser ablation and may facilitate more conformal tumor heating and sparing of normal tissue via enhanced tumor specific heating. This concept was investigated in a xenograft model of prostate cancer (PC-3) using MR temperature imaging (MRTI) in a 1.5T scanner to characterize the spatiotemporal temperature distribution resulting from nanoparticle mediated heating. Tumors with and without intravenously injected AuNS were exposed to an external laser tuned to 808 nm for 180 sec at 4 W/cm(2) under real-time monitoring with proton resonance frequency shift based MRTI. Microscopy indicated that these nanoparticles (140-150 nm) accumulated passively in the tumor and remained close to the tumor microvasculature. MRTI measured a statistically significant (p < 0.001) increase in maximum temperature in the tumor cortex (mean = 21 ± 7°C) in +AuNS tumors versus control tumors. Analysis of the temperature maps helped demonstrate that the overall distribution of temperature within +AuNS tumors was demonstrably higher versus control, and resulted in damage visible on histopathology. This research demonstrates that passive uptake of intravenously injected AuNS in PC-3 xenografts converts the tumor vasculature into a potent heating source for nanoparticle mediated ablation at power levels which do not generate significant damage in normal tissue. When used in conjunction with MRTI, this has implications for development and validation of more conformal delivery of therapy for interstitial laser ablations.
Collapse
Affiliation(s)
- R Jason Stafford
- Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, USA.
| | | | | | | | | | | |
Collapse
|
58
|
KURODA KAGAYAKI, MORITA SHUHEI, LAM MIEKEE, OBARA MAKOTO, BARON PAUL, BARTELS WILBERT, HONDA MASATOSHI, HORIE TOMOHIKO, IMAI YUTAKA. Feasibility of Noninvasive Magnetic Resonance Temperature Imaging of Fat and Water Based on Methylene Proton Spin-lattice Relaxation Time and Water Proton Resonance Frequency. ACTA ACUST UNITED AC 2012. [DOI: 10.3191/thermalmed.28.87] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/04/2022]
|
59
|
Taylor BA, Elliott AM, Hwang KP, Hazle JD, Stafford RJ. Correlation between the temperature dependence of intrinsic MR parameters and thermal dose measured by a rapid chemical shift imaging technique. NMR IN BIOMEDICINE 2011; 24:1414-1421. [PMID: 21721063 PMCID: PMC3190595 DOI: 10.1002/nbm.1707] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Received: 09/02/2010] [Revised: 12/23/2010] [Accepted: 02/15/2011] [Indexed: 05/31/2023]
Abstract
In order to investigate simultaneous MR temperature imaging and direct validation of tissue damage during thermal therapy, temperature-dependent signal changes in proton resonance frequency (PRF) shifts, R(2)* values, and T1-weighted amplitudes are measured from one technique in ex vivo tissue. Using a multigradient echo acquisition and the Stieglitz-McBride algorithm, the temperature sensitivity coefficients of these parameters are measured in each tissue at high spatiotemporal resolutions (1.6 x 1.6 x 4 mm 3,≤ 5sec) at the range of 25-61 °C. Non-linear changes in MR parameters are examined and correlated with an Arrhenius rate dose model of thermal damage. Using logistic regression, the probability of changes in these parameters is calculated as a function of thermal dose to determine if changes correspond to thermal damage. Temperature sensitivity of R(2)* and, in some cases, T1-weighted amplitudes are statistically different before and after thermal damage occurred. Significant changes in the slopes of R(2)* as a function of temperature are observed. Logistic regression analysis shows that these changes could be accurately predicted using the Arrhenius rate dose model (Ω = 1.01 ± 0.03), thereby showing that the changes in R(2)* could be direct markers of protein denaturation. Overall, by using a chemical shift imaging technique with simultaneous temperature estimation, R(2)* mapping and T1-W imaging, it is shown that changes in the sensitivity of R(2)* and, to a lesser degree, T1-W amplitudes are measured in ex vivo tissue when thermal damage is expected to occur. These changes could possibly be used for direct validation of thermal damage in contrast to model-based predictions.
Collapse
Affiliation(s)
- Brian A. Taylor
- Department of Imaging Physics, The University of Texas M. D. Anderson Cancer Center, Houston, Texas
- The University of Texas Graduate School of Biomedical Sciences, Houston, Texas
| | - Andrew M. Elliott
- Department of Imaging Physics, The University of Texas M. D. Anderson Cancer Center, Houston, Texas
| | - Ken-Pin Hwang
- Department of Imaging Physics, The University of Texas M. D. Anderson Cancer Center, Houston, Texas
- Applied Science Laboratory, GE Healthcare, Waukesha, Wisconsin
| | - John D. Hazle
- Department of Imaging Physics, The University of Texas M. D. Anderson Cancer Center, Houston, Texas
| | - R. Jason Stafford
- Department of Imaging Physics, The University of Texas M. D. Anderson Cancer Center, Houston, Texas
| |
Collapse
|
60
|
Le K, Li X, Figueroa D, Towner RA, Garteiser P, Saunders D, Smith N, Liu H, Hode T, Nordquist RE, Chen WR. Assessment of thermal effects of interstitial laser phototherapy on mammary tumors using proton resonance frequency method. JOURNAL OF BIOMEDICAL OPTICS 2011; 16:128001. [PMID: 22191937 PMCID: PMC3245746 DOI: 10.1117/1.3659200] [Citation(s) in RCA: 7] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Received: 07/13/2011] [Revised: 10/18/2011] [Accepted: 10/19/2011] [Indexed: 05/31/2023]
Abstract
Laser immunotherapy (LIT) uses a synergistic approach to treat cancer systemically through local laser irradiation and immunological stimulation. Currently, LIT utilizes dye-assisted noninvasive laser irradiation to achieve selective photothermal interaction. However, LIT faces difficulties treating deeper tumors or tumors with heavily pigmented overlying skin. To circumvent these barriers, we use interstitial laser irradiation to induce the desired photothermal effects. The purpose of this study is to analyze the thermal effects of interstitial irradiation using proton resonance frequency (PRF). An 805-nm near-infrared laser with an interstitial cylindrical diffuser was used to treat rat mammary tumors. Different power settings (1.0, 1.25, and 1.5 W) were applied with an irradiation duration of 10 min. The temperature distributions of the treated tumors were measured by a 7 T magnetic resonance imager using PRF. We found that temperature distributions in tissue depended on both laser power and time settings, and that variance in tissue composition has a major influence in temperature elevation. The temperature elevations measured during interstitial laser irradiation by PRF and thermocouple were consistent, with some variations due to tissue composition and the positioning of the thermocouple's needle probes. Our results indicated that, for a tissue irradiation of 10 min, the elevation of rat tumor temperature ranged from 8 to 11°C for 1 W and 8 to 15°C for 1.5 W. This is the first time a 7 T magnetic resonance imager has been used to monitor interstitial laser irradiation via PRF. Our work provides a basic understanding of the photothermal interaction needed to control the thermal damage inside a tumor using interstitial laser treatment. Our work may lead to an optimal protocol for future cancer treatment using interstitial phototherapy in conjunction with immunotherapy.
Collapse
Affiliation(s)
- Kelvin Le
- University of Central Oklahoma, Department of Engineering and Physics, Edmond, Oklahoma 73034, USA
| | | | | | | | | | | | | | | | | | | | | |
Collapse
|
61
|
Taylor BA, Elliott AM, Hwang KP, Shetty A, Hazle JD, Stafford RJ. Measurement of temperature dependent changes in bone marrow using a rapid chemical shift imaging technique. J Magn Reson Imaging 2011; 33:1128-35. [PMID: 21509871 DOI: 10.1002/jmri.22537] [Citation(s) in RCA: 9] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/14/2023] Open
Abstract
PURPOSE To provide quantitative temperature monitoring for thermal therapies in bone marrow by measuring temperature-dependent signal changes in the bone marrow of ex vivo canine femurs heated with a 980-nm laser at 1.5T and 3.0T. MATERIALS AND METHODS Using a multi-gradient echo (≤ 16) acquisition and signal modeling with the Stieglitz-McBride algorithm, the temperature sensitivity coefficients (TSC, ppm/°C) of water and multiple lipid components' proton resonance frequency (PRF) values are measured at high spatiotemporal resolutions (1.6 × 1.6 × 4 mm(3) , ≤ 5 seconds). Responses in R(2) * and amplitudes of each peak were also measured as a function of temperature simultaneously. RESULTS Calibrations demonstrate that lipid signal may be used to compensate for B(0) errors to provide accurate temperature readings (<1.0°C). Over a temperature range of 17.2-57.2°C, the TSCs after correction to a bulk methylene reference are -0.87 × 10(-2) ± 4.7 × 10(-4) ppm/°C and -0.87 × 10(-2) ± 4.0 × 10(-4) ppm/°C for 1.5T and 3.0T, respectively. CONCLUSION Overall, we demonstrate that accurate and precise temperature measurements can be made in bone marrow. In addition, the relationship of R(2) * and signal amplitudes with respect to temperature are shown to differ significantly where conformal changes are predicted by Arrhenius rate model analysis.
Collapse
Affiliation(s)
- Brian A Taylor
- Department of Imaging Physics, University of Texas M.D. Anderson Cancer Center, Houston, Texas 77030, USA
| | | | | | | | | | | |
Collapse
|
62
|
Kickhefel A, Rosenberg C, Weiss CR, Rempp H, Roland J, Schick F, Hosten N. Clinical evaluation of MR temperature monitoring of laser-induced thermotherapy in human liver using the proton-resonance-frequency method and predictive models of cell death. J Magn Reson Imaging 2011; 33:704-12. [PMID: 21563256 DOI: 10.1002/jmri.22499] [Citation(s) in RCA: 24] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/15/2022] Open
Abstract
PURPOSE To assess the feasibility, precision, and accuracy of real-time temperature mapping (TMap) during laser-induced thermotherapy (LITT) for clinical practice in patients liver with a gradient echo (GRE) sequence using the proton resonance frequency (PRF) method. MATERIALS AND METHODS LITT was performed on 34 lesions in 18 patients with simultaneous real-time visualization of relative temperature changes. Correlative contrast-enhanced T1-weighted magnetic resonance (MR) images of the liver were acquired after treatment using the same slice positions and angulations as TMap images acquired during LITT. For each slice, TMap and follow-up images were registered for comparison. Afterwards, segmentation based on temperature (T) >52°C on TMap and based on necrosis seen on follow-up images was performed. These segmented structures were overlaid and divided into zones where the TMap was found to either over- or underestimate necrosis on the postcontrast images. Regions with T>52°C after 20 minutes were defined as necrotic tissue based on data received from two different thermal dose models. RESULTS The average intersecting region of TMap and necrotic zone was 87% ± 5%, the overestimated 13% ± 4%, and the underestimated 13% ± 5%. CONCLUSION This study demonstrates that MR temperature mapping appears reasonably capable of predicting tissue necrosis on the basis of indicating regions having greater temperatures than 52°C and could be used to monitor and adjust the thermal therapy appropriately during treatment.
Collapse
|
63
|
Covaciu L, Weis J, Bengtsson C, Allers M, Lunderquist A, Ahlström H, Rubertsson S. Brain temperature in volunteers subjected to intranasal cooling. Intensive Care Med 2011; 37:1277-84. [DOI: 10.1007/s00134-011-2264-7] [Citation(s) in RCA: 27] [Impact Index Per Article: 1.9] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/12/2010] [Accepted: 05/07/2011] [Indexed: 02/05/2023]
|
64
|
de Senneville BD, Ries M, Bartels LW, Moonen CTW. MRI-Guided High-Intensity Focused Ultrasound Sonication of Liver and Kidney. INTERVENTIONAL MAGNETIC RESONANCE IMAGING 2011. [DOI: 10.1007/174_2011_394] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 01/15/2023]
|
65
|
Li Z, Vogel M, Maccarini PF, Stakhursky V, Soher BJ, Craciunescu OI, Das S, Arabe OA, Joines WT, Stauffer PR. Improved hyperthermia treatment control using SAR/temperature simulation and PRFS magnetic resonance thermal imaging. Int J Hyperthermia 2010; 27:86-99. [PMID: 21070140 DOI: 10.3109/02656736.2010.501509] [Citation(s) in RCA: 28] [Impact Index Per Article: 1.9] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/20/2023] Open
Abstract
PURPOSE This article explores the feasibility of using coupled electromagnetic and thermodynamic simulations to improve planning and control of hyperthermia treatments for cancer. The study investigates the usefulness of preplanning to improve heat localisation in tumour targets in treatments monitored with PRFS-based magnetic resonance thermal imaging (MRTI). METHODS Heating capabilities of a cylindrical radiofrequency (RF) mini-annular phased array (MAPA) applicator were investigated with electromagnetic and thermal simulations of SAR in homogeneous phantom models and two human leg sarcomas. High frequency structure simulator (HFSS) (Ansoft) was used for electromagnetic simulations and SAR patterns were coupled into EPhysics (Ansoft) for thermal modelling with temperature-dependent variable perfusion. Simulations were accelerated by integrating tumour-specific anatomy into a pre-gridded whole body tissue model. To validate this treatment planning approach, simulations were compared with MR thermal images in both homogenous phantoms and heterogeneous tumours. RESULTS SAR simulations demonstrated excellent agreement with temperature rise distributions obtained with MR thermal imaging in homogeneous phantoms and clinical treatments of large soft-tissue sarcomas. The results demonstrate feasibility of preplanning appropriate relative phases of antennas for localising heat in tumour. CONCLUSIONS Advances in the accuracy of computer simulation and non-invasive thermometry via MR thermal imaging have provided powerful new tools for optimisation of clinical hyperthermia treatments. Simulations agree well with MR thermal images in both homogeneous tissue models and patients with lower leg tumours. This work demonstrates that better quality hyperthermia treatments should be possible when simplified hybrid model simulations are performed routinely as part of the clinical pretreatment plan.
Collapse
Affiliation(s)
- Zhen Li
- Department of Electric and Computer Engineering, School of Engineering
| | | | | | | | | | | | | | | | | | | |
Collapse
|
66
|
Ranneberg M, Weiser M, Weihrauch M, Budach V, Gellermann J, Wust P. Regularized antenna profile adaptation in online hyperthermia treatment. Med Phys 2010; 37:5382-94. [DOI: 10.1118/1.3488896] [Citation(s) in RCA: 9] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
|
67
|
Sprinkhuizen SM, Bakker CJG, Bartels LW. Absolute MR thermometry using time-domain analysis of multi-gradient-echo magnitude images. Magn Reson Med 2010; 64:239-48. [DOI: 10.1002/mrm.22429] [Citation(s) in RCA: 27] [Impact Index Per Article: 1.8] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/24/2022]
|
68
|
Covaciu L, Rubertsson S, Ortiz-Nieto F, Ahlström H, Weis J. Human brain MR spectroscopy thermometry using metabolite aqueous-solution calibrations. J Magn Reson Imaging 2010; 31:807-14. [DOI: 10.1002/jmri.22107] [Citation(s) in RCA: 21] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/08/2022] Open
|
69
|
Pan X, Li C, Ying K, Weng D, Qin W, Li K. Model-based PRFS thermometry using fat as the internal reference and the extended Prony algorithm for model fitting. Magn Reson Imaging 2010; 28:418-26. [DOI: 10.1016/j.mri.2009.11.002] [Citation(s) in RCA: 5] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/25/2009] [Revised: 08/05/2009] [Accepted: 11/25/2009] [Indexed: 10/19/2022]
|
70
|
Lüdemann L, Wlodarczyk W, Nadobny J, Weihrauch M, Gellermann J, Wust P. Non-invasive magnetic resonance thermography during regional hyperthermia. Int J Hyperthermia 2010; 26:273-82. [DOI: 10.3109/02656731003596242] [Citation(s) in RCA: 52] [Impact Index Per Article: 3.5] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/13/2022] Open
|
71
|
Wyatt C, Soher B, Maccarini P, Charles HC, Stauffer P, Macfall J. Hyperthermia MRI temperature measurement: evaluation of measurement stabilisation strategies for extremity and breast tumours. Int J Hyperthermia 2010; 25:422-33. [PMID: 19925322 DOI: 10.1080/02656730903133762] [Citation(s) in RCA: 31] [Impact Index Per Article: 2.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/16/2022] Open
Abstract
PURPOSE MR thermometry using the proton resonance frequency shift (PRFS) method has been used to measure temperature changes during clinical hyperthermia treatment. However, frequency drift of the MRI system can add large errors to the measured temperature change. These drifts can be measured and corrected using oil references placed around the treatment region. In this study, the number and position of four or more oil references were investigated to obtain a practical approach to correct frequency drift during PRFS thermometry in phantoms and in vivo. MATERIALS AND METHODS Experiments were performed in a 140 MHz four antenna mini-annular phased array (MAPA) heat applicator (for treatment of extremity tumours) and an applicator for heating of the breast, with symmetric and asymmetric positioning of the oil references, respectively. Temperature change PRFS images were obtained during an hour or more of measurement with no application of heat. Afterwards, errors in calculating temperature change due to system drift were quantified with and without various oil reference correction arrangements. RESULTS Results showed good temperature correction in phantoms and in a human leg, with average errors of 0.28 degrees C and 0.94 degrees C respectively. There was further improvement in the leg when using eight or more oil references, reducing the average error to 0.44 degrees C, while the phantoms showed no significant improvement. CONCLUSIONS These results indicate that oil reference correction performs well in vivo, and that eight references can improve the correction by up to 0.5 degrees C compared to four references.
Collapse
Affiliation(s)
- Cory Wyatt
- Department of Biomedical Engineering, Duke University, Durham, North Carolina 27708, USA.
| | | | | | | | | | | |
Collapse
|
72
|
Kickhefel A, Roland J, Weiss C, Schick F. Accuracy of real-time MR temperature mapping in the brain: a comparison of fast sequences. Phys Med 2010; 26:192-201. [PMID: 20096617 DOI: 10.1016/j.ejmp.2009.11.006] [Citation(s) in RCA: 49] [Impact Index Per Article: 3.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 02/12/2009] [Revised: 10/29/2009] [Accepted: 11/29/2009] [Indexed: 11/15/2022] Open
Abstract
PURPOSE To compare magnetic resonance (MR) thermometry based on the proton resonance frequency (PRF) method using a single shot echoplanar imaging (ss EPI) sequence to both of the standard sequences, gradient echo (GRE) and segmented echoplanar imaging (seg EPI) in the in vivo human brain, at 1.5T and 3T. MATERIAL AND METHODS Repetitive MR thermometry was performed on the brain of six volunteers using GRE, seg EPI, and ss EPI sequences on whole-body 1.5T and 3T clinical systems using comparable acquisition parameters. Phase stability and temperature data precision in the human head were determined over 12 min for the three sequences at both field strengths. An ex-vivo swine skeletal muscle model was used to evaluate temperature accuracy of the ss EPI sequence during heating by high intensity focused ultrasound (HIFU). RESULTS In-vivo examinations of brain revealed an average temperature precision of 0.37 °C/0.39 °C/0.16 °C at 3T for the GRE/seg EPI/ss EPI sequences. At 1.5T, a precision of 0.58 °C/0.63 °C/0.21 °C was achieved. In the ex-vivo swine model, a strong correlation of temperature data derived using ss EPI and GRE sequences was found with a temperature deviation <1 °C. CONCLUSION The ss EPI sequence was the fastest and the most precise sequence for MR thermometry, with significantly higher accuracy compared to GRE.
Collapse
Affiliation(s)
- A Kickhefel
- Siemens Healthcare, H IM MR PLM AW Oncology, FH5/2008, Allee am Röthelheimpark 2, 91052 Erlangen, Germany.
| | | | | | | |
Collapse
|
73
|
Craciunescu OI, Stauffer PR, Soher BJ, Wyatt CR, Arabe O, Maccarini P, Das SK, Cheng KS, Wong TZ, Jones EL, Dewhirst MW, Vujaskovic Z, MacFall JR. Accuracy of real time noninvasive temperature measurements using magnetic resonance thermal imaging in patients treated for high grade extremity soft tissue sarcomas. Med Phys 2010; 36:4848-58. [PMID: 19994492 DOI: 10.1118/1.3227506] [Citation(s) in RCA: 51] [Impact Index Per Article: 3.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
PURPOSE To establish accuracy of real time noninvasive temperature measurements using magnetic resonance thermal imaging in patients treated for high grade extremity soft tissue sarcomas. METHODS Protocol patients with advanced extremity sarcomas were treated with external beam radiation therapy and hyperthermia. Invasive temperature measures were compared to noninvasive magnetic resonance thermal imaging (MRTI) at 1.5 T performed during hyperthermia. Volumetric temperature rise images were obtained using the proton resonance frequency shift (PRFS) technique during heating in a 140 MHz miniannular phased array applicator. MRTI temperature changes were compared to invasive measurements of temperature with a multisensor fiber optic probe inside a #15 g catheter in the tumor. Since the PRFS technique is sensitive to drifts in the primary imaging magnetic field, temperature change distributions were corrected automatically during treatment using temperature-stable reference materials to characterize field changes in 3D. The authors analyzed MRT images and compared, in evaluable treatments, MR-derived temperatures to invasive temperatures measured in extremity sarcomas. Small regions of interest (ROIs) were specified near each invasive sensor identified on MR images. Temperature changes in the interstitial sensors were compared to the corresponding ROI PRFS-based temperature changes over the entire treatment and over the steady-state period. Nonevaluable treatments (motion/imaging artifacts, noncorrectable drifts) were not included in the analysis. RESULTS The mean difference between MRTI and interstitial probe measurements was 0.91 degrees C for the entire heating time and 0.85 degrees C for the time at steady state. These values were obtained from both tumor and normal tissue ROIs. When the analysis is done on just the tumor ROIs, the mean difference for the whole power on time was 0.74 degrees C and during the period of steady state was 0.62 degrees C. CONCLUSIONS The data show that for evaluable treatments, excellent correlation (deltaT < 1 degrees C) of MRTI-ROI and invasive measurements can be achieved, but that motion and other artifacts are still serious challenges that must be overcome in future work.
Collapse
Affiliation(s)
- Oana I Craciunescu
- Department of Radiation Oncology, Duke University Medical Center, Durham, North Carolina 27710, USA.
| | | | | | | | | | | | | | | | | | | | | | | | | |
Collapse
|
74
|
Taylor BA, Hwang KP, Hazle JD, Stafford RJ. Autoregressive moving average modeling for spectral parameter estimation from a multigradient echo chemical shift acquisition. Med Phys 2009; 36:753-64. [PMID: 19378736 DOI: 10.1118/1.3075819] [Citation(s) in RCA: 11] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/11/2023] Open
Abstract
The authors investigated the performance of the iterative Steiglitz-McBride (SM) algorithm on an autoregressive moving average (ARMA) model of signals from a fast, sparsely sampled, multiecho, chemical shift imaging (CSI) acquisition using simulation, phantom, ex vivo, and in vivo experiments with a focus on its potential usage in magnetic resonance (MR)-guided interventions. The ARMA signal model facilitated a rapid calculation of the chemical shift, apparent spin-spin relaxation time (T2*), and complex amplitudes of a multipeak system from a limited number of echoes (< or equal 16). Numerical simulations of one- and two-peak systems were used to assess the accuracy and uncertainty in the calculated spectral parameters as a function of acquisition and tissue parameters. The measured uncertainties from simulation were compared to the theoretical Cramer-Rao lower bound (CRLB) for the acquisition. Measurements made in phantoms were used to validate the T2* estimates and to validate uncertainty estimates made from the CRLB. We demonstrated application to real-time MR-guided interventions ex vivo by using the technique to monitor a percutaneous ethanol injection into a bovine liver and in vivo to monitor a laser-induced thermal therapy treatment in a canine brain. Simulation results showed that the chemical shift and amplitude uncertainties reached their respective CRLB at a signal-to-noise ratio (SNR) > or =5 for echo train lengths (ETLs) > or =4 using a fixed echo spacing of 3.3 ms. T2* estimates from the signal model possessed higher uncertainties but reached the CRLB at larger SNRs and/or ETLs. Highly accurate estimates for the chemical shift (<0.01 ppm) and amplitude (<1.0%) were obtained with > or =4 echoes and for T2*(<1.0%) with > or =7 echoes. We conclude that, over a reasonable range of SNR, the SM algorithm is a robust estimator of spectral parameters from fast CSI acquisitions that acquire < or =16 echoes for one- and two-peak systems. Preliminary ex vivo and in vivo experiments corroborated the results from simulation experiments and further indicate the potential of this technique for MR-guided interventional procedures with high spatiotemporal resolution approximately 1.6 x 1.6 x 4 mm3 in < or =5 s.
Collapse
Affiliation(s)
- Brian A Taylor
- Department of Imaging Physics, The University of Texas M. D. Anderson Cancer Center, Houston, Texas 77030, USA.
| | | | | | | |
Collapse
|
75
|
Weis J, Covaciu L, Rubertsson S, Allers M, Lunderquist A, Ahlström H. Noninvasive monitoring of brain temperature during mild hypothermia. Magn Reson Imaging 2009; 27:923-32. [PMID: 19282122 DOI: 10.1016/j.mri.2009.01.011] [Citation(s) in RCA: 19] [Impact Index Per Article: 1.2] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/08/2008] [Accepted: 01/11/2009] [Indexed: 10/21/2022]
Abstract
The main purpose of this study was to verify the feasibility of brain temperature mapping with high-spatial- and reduced-spectral-resolution magnetic resonance spectroscopic imaging (MRSI). A secondary goal was to determine the temperature coefficient of water chemical shift in the brain with and without internal spectral reference. The accuracy of the proposed MRSI method was verified using a water and vegetable oil phantom. Selective decrease of the brain temperature of pigs was induced by intranasal cooling. Temperature reductions between 2 degrees C and 4 degrees C were achieved within 20 min. The relative changes in temperature during the cooling process were monitored using MRSI. The reference temperature was measured with MR-compatible fiber-optic probes. Single-voxel (1)H MRS was used for measurement of absolute brain temperature at baseline and at the end of cooling. The temperature coefficient of the water chemical shift of brain tissue measured by MRSI without internal reference was -0.0192+/-0.0019 ppm/degrees C. The temperature coefficients of the water chemical shift relative to N-acetylaspartate, choline-containing compounds and creatine were -0.0096+/-0.0009, -0.0083+/-0.0007 and -0.0091+/-0.0011 ppm/degrees C, respectively. The results of this study indicate that MRSI with high spatial and reduced spectral resolutions is a reliable tool for monitoring long-term temperature changes in the brain.
Collapse
Affiliation(s)
- Jan Weis
- Department of Radiology, MR Unit, Uppsala University Hospital, Uppsala, Sweden.
| | | | | | | | | | | |
Collapse
|
76
|
Peters NH, Bartels LW, Sprinkhuizen SM, Vincken KL, Bakker CJ. Do respiration and cardiac motion induce magnetic field fluctuations in the breast and are there implications for MR thermometry? J Magn Reson Imaging 2009; 29:731-5. [DOI: 10.1002/jmri.21680] [Citation(s) in RCA: 34] [Impact Index Per Article: 2.1] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022] Open
|
77
|
Zhu M, Bashir A, Ackerman JJ, Yablonskiy DA. Improved calibration technique for in vivo proton MRS thermometry for brain temperature measurement. Magn Reson Med 2009; 60:536-41. [PMID: 18727039 DOI: 10.1002/mrm.21699] [Citation(s) in RCA: 41] [Impact Index Per Article: 2.6] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/06/2022]
Abstract
The most common MR-based approach to noninvasively measure brain temperature relies on the linear relationship between the (1)H MR resonance frequency of tissue water and the tissue's temperature. Herein we provide the most accurate in vivo assessment existing thus far of such a relationship. It was derived by acquiring in vivo MR spectra from a rat brain using a high field (11.74 Tesla [T]) MRI scanner and a single-voxel MR spectroscopy technique based on a LASER pulse sequence. Data were analyzed using three different methods to estimate the (1)H resonance frequencies of water and the metabolites NAA, Cho, and Cr, which are used as temperature-independent internal (frequency) references. Standard modeling of frequency-domain data as composed of resonances characterized by Lorentzian line shapes gave the tightest resonance-frequency versus temperature correlation. An analysis of the uncertainty in temperature estimation has shown that the major limiting factor is an error in estimating the metabolite frequency. For example, for a metabolite resonance linewidth of 8 Hz, signal sampling rate of 2 Hz and SNR of 5, an accuracy of approximately 0.5 degrees C can be achieved at a magnetic field of 3T. For comparison, in the current study conducted at 11.74T, the temperature estimation error was approximately 0.1 degrees C.
Collapse
Affiliation(s)
- M Zhu
- Department of Chemistry, Washington University, Saint Louis, Missouri 63110, USA
| | | | | | | |
Collapse
|
78
|
Abstract
Minimally invasive thermal therapy as local treatment of benign and malignant diseases has received increasing interest in recent years. Safety and efficacy of the treatment require accurate temperature measurement throughout the thermal procedure. Noninvasive temperature monitoring is feasible with magnetic resonance (MR) imaging based on temperature-sensitive MR parameters such as the proton resonance frequency (PRF), the diffusion coefficient (D), T1 and T2 relaxation times, magnetization transfer, the proton density, as well as temperature-sensitive contrast agents. In this article the principles of temperature measurements with these methods are reviewed and their usefulness for monitoring in vivo procedures is discussed. Whereas most measurements give a temperature change relative to a baseline condition, temperature-sensitive contrast agents and spectroscopic imaging can provide absolute temperature measurements. The excellent linearity and temperature dependence of the PRF and its near independence of tissue type have made PRF-based phase mapping methods the preferred choice for many in vivo applications. Accelerated MRI imaging techniques for real-time monitoring with the PRF method are discussed. Special attention is paid to acquisition and reconstruction methods for reducing temperature measurement artifacts introduced by tissue motion, which is often unavoidable during in vivo applications.
Collapse
Affiliation(s)
- Viola Rieke
- Department of Radiology, Stanford University, Stanford, CA 94305-5488, USA.
| | | |
Collapse
|
79
|
Taylor BA, Hwang KP, Elliott AM, Shetty A, Hazle JD, Stafford RJ. Dynamic chemical shift imaging for image-guided thermal therapy: analysis of feasibility and potential. Med Phys 2008; 35:793-803. [PMID: 18383702 DOI: 10.1118/1.2831915] [Citation(s) in RCA: 24] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
A fast chemical shift imaging (CSI) technique based on a multiple gradient-recalled acquisition using a small number of echoes with intentional aliasing of the reference lipid peak is studied to determine its feasibility for temperature monitoring. Simulations were implemented to find parameters where the lipid and water peaks can be measured using a Fourier-based peak fitting approach as well as using an innovative autoregressive moving average technique. A phantom consisting of 50% mayonnaise/50% lemon juice was calibrated to temperature and compared to literature values. A porcine kidney was treated ex vivo with an external laser and imaged with the CSI technique with comparisons to temperature readings from a fluoroptic monitoring system and complex phase difference (CPD) calculations. To demonstrate the technique in vivo, a Balb/c mouse with a CT26 xenograft in the subcutaneous lower back was treated using gold-coated, silica-core nanoshells heated with an 808 nm interstitial laser. Compared to standard CPD techniques using a two-dimensional fast spoiled gradient recalled echo, this technique maintains spatiotemporal resolution, has high signal-to-noise ratio and accuracy over a wide range of T2* tissue values, can separate water and lipid signals, and additionally can use the lipid peak, when present, as an internal reference.
Collapse
Affiliation(s)
- Brian A Taylor
- Department of Imaging Physics, The University of Texas M.D. Anderson Cancer Center, Houston, Texas 77030, USA
| | | | | | | | | | | |
Collapse
|
80
|
Pauly KB, Diederich CJ, Rieke V, Bouley D, Chen J, Nau WH, Ross AB, Kinsey AM, Sommer G. Magnetic resonance-guided high-intensity ultrasound ablation of the prostate. Top Magn Reson Imaging 2007; 17:195-207. [PMID: 17414077 DOI: 10.1097/rmr.0b013e31803774dd] [Citation(s) in RCA: 68] [Impact Index Per Article: 3.8] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/21/2022]
Abstract
OBJECTIVES This paper describes our work in developing techniques and devices for magnetic resonance (MR)-guided high-intensity ultrasound ablation of the prostate and includes review of relevant literature. METHODS Catheter-based high-intensity ultrasound applicators, in interstitial and transurethral configurations, were developed to be used under MR guidance. Magnetic resonance thermometry and the relevant characteristics and artifacts were evaluated during in vivo thermal ablation of the prostate in 10 animals. Contrast-enhanced MR imaging (MRI) and diffusion-weighted MRI were used to assess tissue damage and compared with histology. RESULTS During evaluation of these applicators, MR thermometry was used to monitor the temperature distributions in the prostate in real time. Magnetic resonance-derived maximum temperature thresholds of 52 degrees C and thermal dose thresholds of 240 minutes were used to control the extent of treatment and qualitatively correlated well with posttreatment imaging studies and histology. The directional transurethral devices are selective in their ability to target well-defined regions of the prostate gland and can be rotated in discrete steps to conform treatment to prescribed boundaries. The curvilinear applicator is the most precise of these directional techniques. Multisectored transurethral applicators, with dynamic angular control of heating and no rotation requirements, offer a fast and less complex means of treatment with less selective contouring. CONCLUSIONS The catheter-based ultrasound devices can produce spatially selective regions of thermal destruction in prostate. The MR thermal imaging and thermal dose maps, obtained in multiple slices through the target volume, are useful for controlling therapy delivery (rotation, power levels, duration). Contrast-enhanced T1-weighted MRI and diffusion-weighted imaging are useful tools for assessing treatment.
Collapse
Affiliation(s)
- Kim Butts Pauly
- Department of Radiology, Stanford University, Stanford, CA 94305-5488, USA.
| | | | | | | | | | | | | | | | | |
Collapse
|
81
|
Gellermann J, Hildebrandt B, Issels R, Ganter H, Wlodarczyk W, Budach V, Felix R, Tunn PU, Reichardt P, Wust P. Noninvasive magnetic resonance thermography of soft tissue sarcomas during regional hyperthermia. Cancer 2006; 107:1373-82. [PMID: 16902986 DOI: 10.1002/cncr.22114] [Citation(s) in RCA: 112] [Impact Index Per Article: 5.9] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022]
Abstract
BACKGROUND The objective of this study was to evaluate noninvasive magnetic resonance (MR) thermography for the monitoring of regional hyperthermia (RHT) in patients with soft tissue sarcomas of the lower extremities and pelvis. METHODS Noninvasive MR monitoring during RHT was performed in 9 patients who had high-risk soft tissue sarcomas of the lower extremities or pelvis during neoadjuvant chemotherapy plus RHT in the scope of the European Organization for Research and Treatment of Cancer 62961/European Society for Hyperthermic Oncology RHT-95 study. Anatomic and temperature-sensitive data sets were acquired every 10 minutes before and during RHT (using gradient-echo-sequences with variable echo times). MR temperature distributions were derived from the phase differences by using the proton-resonance frequency shift method. A phase convolution setting phase shifts to zero in the fat tissue was performed as a drift correction. The mean MR temperatures in the tumor and muscles and the index temperatures (e.g., T90, which covers 90% of the target volume) and thermal doses were determined and compared with pathohistologic responses and direct temperature measurements if available. RESULTS Thirty of 72 MR-thermography data sets (>40% of heat sessions) were evaluable. A significant correlation was observed between pathohistologic response (defined as a necrosis rate >or=90%) and standardized thermal parameters, such as thermal dose cumulative equivalent minutes at 43 degrees C to 90% of the target volume (T90) (P = .050), mean T90 (P = .048), or T50 (P = .050). The correlation of 13 conventional temperature measurements performed in selected patients and sessions invasively in the tumor or noninvasively in rectum and bladder revealed an excellent correlation with MR temperatures (R2 = .96). CONCLUSIONS Noninvasive MR thermography of soft tissue sarcoma was feasible and suitable for validating the quality of heating during RHT.
Collapse
Affiliation(s)
- Johanna Gellermann
- Clinic for Radiation Medicine, University Hospital Charite, Campus Virchow Klinikum, Berlin, Germany
| | | | | | | | | | | | | | | | | | | |
Collapse
|