Puchnin V, Jandaliyeva A, Hurshkainen A, Solomakha G, Nikulin A, Petrova P, Lavrenteva A, Andreychenko A, Shchelokova A. Quadrature transceive wireless coil: Design concept and application for bilateral breast MRI at 1.5 T.
Magn Reson Med 2023;
89:1251-1264. [PMID:
36336799 DOI:
10.1002/mrm.29507]
[Citation(s) in RCA: 1] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/21/2022] [Revised: 09/20/2022] [Accepted: 10/09/2022] [Indexed: 11/09/2022]
Abstract
PURPOSE
Development of a novel quadrature inductively driven transceive wireless coil for breast MRI at 1.5 T.
METHODS
A quadrature wireless coil (HHMM-coil) design has been developed as a combination of two linearly polarized coils: a pair of 'metasolenoid' coils (MM-coil) and a pair of Helmholtz-type coils (HH-coil). The MM-coil consisted of an array of split-loop resonators. The HH-coil design included two electrically connected flat spirals. All the wireless coils were coupled to a whole-body birdcage coil. The HHMM-coil was studied and compared to the linear coils in terms of transmit and SAR efficiencies via numerical simulations. A prototype of HHMM-coil was built and tested on a 1.5 T scanner in a phantom and healthy volunteer. We also proposed an extended design of the HHMM-coil and compared its performance to a dedicated breast array.
RESULTS
Numerical simulations of the HHMM-coil with a female voxel model have shown more than a 2.5-fold increase in transmit efficiency and a 1.7-fold enhancement of SAR efficiency compared to the linearly polarized coils. Phantom and in vivo imaging showed good agreement with the numerical simulations. Moreover, the HHMM-coil provided good image quality, visualizing all areas of interest similar to a multichannel breast array with a 32% reduction in signal-to-noise ratio.
CONCLUSION
The proposed quadrature HHMM-coil allows the B 1 + $$ {\mathrm{B}}_1^{+} $$ -field to be significantly better focused in the region-of-interest compared to the linearly polarized coils. Thus, the HHMM-coil provides high-quality breast imaging on a 1.5 T scanner using a whole-body birdcage coil for transmit and receive.