Vesterberg O. Specific, sensitive and accurate quantification of albumin, retinol binding protein and transferrin in human urine and serum by zone immunoelectrophoresis assay (ZIA).
Electrophoresis 1994;
15:589-93. [PMID:
7925236 DOI:
10.1002/elps.1150150182]
[Citation(s) in RCA: 3] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/27/2023]
Abstract
For zone immunoelectrophoresis assay (ZIA) glass tubes, ID 2 mm and 90 mm high, are filled to 2/3 with buffer containing agarose and antibodies against the protein to be quantified, each sample being pipetted on top of separate agarose gel rods. On electrophoresis at 35-150 V for several hours, the sample proteins enter the gel with resultant immunoprecipitates, visualized by staining. The extension of each immunoprecipitation zone from the upper gel surface (measured with a ruler) is directly proportional to the amount of protein in each sample and can easily be quantitated by comparison with a linear calibration curve. ZIA can be used for quantification of several proteins in blood serum and plasma as well as in urine, as is illustrated for albumin, retinol-binding protein (RBP) and transferrin. The recovery of the pure proteins added to urine is often close to 100%. ZIA has many advantages: (i) simple apparatus and procedure (no gel punching nor cooling), (ii) minimal antiserum consumption (1 mL may allow > 1000 assays), (iii) electrophoresis can be performed within a few hours or overnight, (iv) low coefficient of variation (often < 4%), (v) linear calibration curves, (vi) low detection limit (< 20 ng/mL), (vii) wide concentration ranges, (viii) no kits nor unique antisera preparation are required, and (ix) good agreement with the results from other methods.
Collapse