Galjaard S, Baeck S, Ameye L, Bourne T, Timmerman D, Devlieger R. Use of a pocket-sized ultrasound machine (PUM) for routine examinations in the third trimester of pregnancy.
ULTRASOUND IN OBSTETRICS & GYNECOLOGY : THE OFFICIAL JOURNAL OF THE INTERNATIONAL SOCIETY OF ULTRASOUND IN OBSTETRICS AND GYNECOLOGY 2014;
44:64-68. [PMID:
24357339 DOI:
10.1002/uog.13285]
[Citation(s) in RCA: 7] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 07/25/2013] [Revised: 10/15/2013] [Accepted: 12/05/2013] [Indexed: 06/03/2023]
Abstract
OBJECTIVES
To evaluate the application of a pocket-sized ultrasound machine (PUM) in a routine antenatal third-trimester scan compared with a high-specification ultrasound machine (HSUM).
METHODS
This was an observational cohort study of 53 unselected women who came for their routine third-trimester ultrasound scan. The first scan was performed by an experienced ultrasonographer on an HSUM for fetal growth, fetal wellbeing and placental location. A second experienced operator performed the measurements on the PUM. Both operators were blinded to the other's findings and measurement results.
RESULTS
Fifty-one patients were eligible and scanned at a median gestational age of 31 + 1 (range, 27 + 2 to 36 + 0) weeks. Mean pregestational body mass index was 22.9 ± 3.1 (range, 17.6-35.7) kg/m(2) . In the comparison between measurements made using the PUM and HSUM, perfect agreement was found for fetal position, fetal bladder and stomach visualization (all κ 1.0) and very good agreement was found for placental position (κ 0.86). Measurements of deepest vertical pocket correlated moderately (intraclass correlation coefficient (ICC) 0.38 (95% CI, 0.12-0.59); Bland-Altman bias, 2.43 (95% limits of agreement (LoA), -22.65 to 27.51)). For fetal growth measurements there was very good agreement for biparietal diameter (ICC, 0.93 (95% CI, 0.88-0.96), Bland-Altman bias, -1.06 (95% LoA, -5.07 to 2.95)), and good agreement for femur length (ICC, 0.66 (95% CI, 0.48-0.79), Bland-Altman bias 0.56 (95% LoA, -5.97 to 7.08)) and transcerebellar diameter (ICC, 0.65 (95% CI, 0.46-0.78), Bland-Altman bias, -0.84 (95% LoA, -7.77 to 6.09)).
CONCLUSION
A battery-driven PUM can be used in third-trimester obstetrics for routine assessment of fetal growth (biparietal diameter, transcerebellar diameter and femur length) and for assessment of fetal wellbeing.
Collapse