Barkmann R, Dencks S, Laugier P, Padilla F, Brixen K, Ryg J, Seekamp A, Mahlke L, Bremer A, Heller M, Glüer CC. Femur ultrasound (FemUS)--first clinical results on hip fracture discrimination and estimation of femoral BMD.
Osteoporos Int 2010;
21:969-76. [PMID:
19693640 DOI:
10.1007/s00198-009-1037-4]
[Citation(s) in RCA: 44] [Impact Index Per Article: 2.9] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 05/27/2009] [Accepted: 07/27/2009] [Indexed: 11/24/2022]
Abstract
SUMMARY
A quantitative ultrasound (QUS) device for measurements at the proximal femur was developed and tested in vivo (Femur Ultrasound Scanner, FemUS). Hip fracture discrimination was as good as for DXA, and a high correlation with hip BMD was achieved. Our results show promise for enhanced QUS-based assessment of osteoporosis.
INTRODUCTION
Dual X-ray absorptiometry (DXA) at the femur is the best predictor of hip fractures, better than DXA measurements at other sites. Calcaneal quantitative ultrasound (QUS) can be used to estimate the general osteoporotic fracture risk, but no femoral QUS measurement has been introduced yet. We developed a QUS scanner for measurements at the femur (Femur Ultrasound Scanner, FemUS) and tested its in vivo performance.
METHODS
Using the FemUS device, we obtained femoral QUS and DXA on 32 women with recent hip fractures and 30 controls. Fracture discrimination and the correlation with femur bone mineral density (BMD) were assessed.
RESULTS
Hip fracture discrimination using the FemUS device was at least as good as with hip DXA and calcaneal QUS. Significant correlations with total hip bone mineral density were found with a correlation coefficient R (2) up to 0.72 and a residual error of about one half of a T-score in BMD.
CONCLUSIONS
QUS measurements at the proximal femur are feasible and show a good performance for hip fracture discrimination. Given the promising results, this laboratory prototype should be reengineered to a clinical applicable instrument. Our results show promise for further enhancement of QUS-based assessment of osteoporosis.
Collapse