van den Berg P, Schweitzer DH, van Haard PMM, Geusens PP, van den Bergh JP. The use of pulse-echo ultrasound in women with a recent non-vertebral fracture to identify those without osteoporosis and/or a subclinical vertebral fracture: a pilot study.
Arch Osteoporos 2020;
15:56. [PMID:
32291527 DOI:
10.1007/s11657-020-00730-7]
[Citation(s) in RCA: 6] [Impact Index Per Article: 1.2] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 12/28/2019] [Accepted: 03/23/2020] [Indexed: 02/03/2023]
Abstract
UNLABELLED
A pilot study on the use of P-EU to identify patients without osteoporosis and/or a subclinical vertebral fracture after a recently sustained non-vertebral fracture (NVF).
INTRODUCTION
Screening with portable devices at emergency departments or plaster rooms could be of interest to limit referrals for dual X-ray absorptiometry (DXA) and vertebral fracture assessment (VFA). We calculated the number of negative tests for osteoporosis and/or subclinical vertebral fractures (VFs) using pulse-echo ultrasonometry (P-UE) at different thresholds.
PATIENTS AND METHODS
In this cross-sectional study, 209 consecutive women of 50-70 years with a recent non-vertebral fracture (NVF) were studied at the Fracture Liaison Service (FLS) of one hospital. All women received DXA/VFA and P-EU (Bindex®) assessments. Various P-EU thresholds (based on the density index (DI, g/cm2)) were analyzed to calculate the best balance between true negative (indeed no osteoporosis and/or subclinical VF) and false negative tests (osteoporosis and/or subclinical VF according to DXA/VFA).
RESULTS
Eighty-three women had osteoporosis (40%) and 17 women at least one VF (8%). Applying the manufacturer's recommended P-EU threshold (DI 0.844 g/cm2) being their proposed cut-off for not having hip osteoporosis resulted in 77 negative tests (37%, 31% true negative and 6% false negative tests). A DI of 0.896 g/cm2 resulted in 40 negative tests (19.3%) (38 true negative (18.3%) and 2 false negative tests (1.0%)).
CONCLUSION
The application of P-EU enables the identification of a substantial proportion of women with recent non-vertebral fractures at the FLS who would not need a DXA/VFA referral because they had no osteoporosis and/or subclinical vertebral fractures. The most conservative P-EU threshold resulted in 18.3% true negative tests verified by DXA/VFA against 1% false negative test results.