Albrecht K, Grob K, Lange U, Müller-Ladner U, Strunk J. Reliability of different Doppler ultrasound quantification methods and devices in the assessment of therapeutic response in arthritis.
Rheumatology (Oxford) 2008;
47:1521-6. [PMID:
18676991 DOI:
10.1093/rheumatology/ken318]
[Citation(s) in RCA: 30] [Impact Index Per Article: 1.9] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022] Open
Abstract
OBJECTIVE
To evaluate the metric qualities of power Doppler ultrasound (PDUS) for different quantification methods and devices in the assessment of arthritis during anti-inflammatory treatment.
METHODS
Twenty-four patients with active arthritis and first-time treatment with the TNF-alpha inhibitor adalimumab underwent sequential clinical, laboratory and US examination at baseline, weeks 2, 6 and 12. 2D and 3D PDUS was performed by two independent investigators. The images and movies were scored from 0 to 3 and the amount of colour pixels and voxels was calculated. In addition, the resistance index of a synovial artery was measured. Thirteen patients were examined with a second US machine.
RESULTS
Treatment response was already observed at week 2 with a significant reduction of 2D (P < 0.01) and 3D scores (P < 0.001). A moderate correlation to 28-joint disease activity score was found for 3D voxel count (r(s) = 0.35, P < 0.001). Interobserver agreement was kappa or ICC >or= 0.8 for all methods except the resistance index (ICC = 0.60). Intermachine agreement was kappa = 0.57 for 2D PDUS score.
CONCLUSIONS
The study demonstrates good to excellent interobserver and moderate intermachine reliability of different PDUS assessment methods in a longitudinal open-label study.
Collapse