Yamada T, Nakano H, Tanabe S, Sakai T, Tanabe S, Oka T, Sakai H, Oshikane T, Nakano T, Ohta A, Kanazawa T, Kaidu M, Ishikawa H. Verification of Qfix Encompass™ couch modeling using the Acuros XB algorithm and HypeArc™ using a high-spatial-resolution two-dimensional diode array. Med Dosim 2023; 48:261-266. [PMID: 37455221 DOI: 10.1016/j.meddos.2023.06.002] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/18/2023] [Revised: 05/15/2023] [Accepted: 06/20/2023] [Indexed: 07/18/2023]
Abstract
We modeled the Qfix Encompass™ immobilization system and further verified the calculated dose distribution of the AcurosXB (AXB) dose calculation algorithm using SRS MapCHECKⓇ (SRSMC) in the HyperArc™ (HA) clinical plan. An Encompass system with a StereoPHAN™ QA phantom was scanned by SOMATOM go.Sim and imported to an Eclipse™ treatment planning system to create a treatment plan for Encompass modeling. The Encompass modeling was performed in the StereoPHAN with a pinpoint ion chamber for 6 MV and 6 MV flattening filter free (6 MV FFF), and 2 × 2 cm2, 4 × 4 cm2, and 6 × 6 cm2 irradiation field sizes. The dose calculation algorithm used was AXB ver. 15.5 with a 1.0 mm calculation grid size. The Hounsfield unit (HU) values of the Encompass modeling were set to 400, -100, -200, and -300 for Encompass, and -400, -600, -700, and -800 for the Encompass base. We evaluated the dose distribution after Encompass modeling by SRSMC using gamma analysis in 12 patients. We adopted HU values of -200 for Encompass, -800 for Encompass base for 6 MV, and -200 for Encompass and -700 for Encompass. Base for 6 MV FFF was adopted as the HU values for the Encompass modeling based on the measurement results. The proposed Encompass modeling resulted in a mean pass rate evaluation >98% for both 6 MV and 6 MV FFF when the 1%/1 mm criterion was used, demonstrating that the proposed HU value can be adopted to calculate more accurate dose distributions.
Collapse
Affiliation(s)
- Takumi Yamada
- Section of Radiology, Department of Clinical Support, Niigata University Medical and Dental Hospital, Niigata, 951-8520, Japan
| | - Hisashi Nakano
- Department of Radiation Oncology, Niigata University Medical and Dental Hospital, Niigata 951-8520, Japan; Department of Medical Physics and Engineering, Osaka University Graduate School of Medicine, Osaka, 565-0871, Japan.
| | - Satoshi Tanabe
- Department of Radiation Oncology, Niigata University Medical and Dental Hospital, Niigata 951-8520, Japan
| | - Tatsuya Sakai
- Section of Radiology, Department of Clinical Support, Niigata University Medical and Dental Hospital, Niigata, 951-8520, Japan
| | - Shunpei Tanabe
- Department of Radiation Oncology, Niigata University Medical and Dental Hospital, Niigata 951-8520, Japan
| | - Tetsuya Oka
- Section of Radiology, Department of Clinical Support, Niigata University Medical and Dental Hospital, Niigata, 951-8520, Japan
| | - Hironori Sakai
- Section of Radiology, Department of Clinical Support, Niigata University Medical and Dental Hospital, Niigata, 951-8520, Japan
| | - Tomoya Oshikane
- Department of Radiation Oncology, Niigata University Medical and Dental Hospital, Niigata 951-8520, Japan
| | - Toshimichi Nakano
- Department of Radiology and Radiation Oncology, Niigata University Graduate School of Medical and Dental Sciences, Niigata, 951-8122, Japan
| | - Atsushi Ohta
- Department of Radiation Oncology, Niigata University Medical and Dental Hospital, Niigata 951-8520, Japan
| | - Tsutomu Kanazawa
- Section of Radiology, Department of Clinical Support, Niigata University Medical and Dental Hospital, Niigata, 951-8520, Japan
| | - Motoki Kaidu
- Department of Radiology and Radiation Oncology, Niigata University Graduate School of Medical and Dental Sciences, Niigata, 951-8122, Japan
| | - Hiroyuki Ishikawa
- Department of Radiology and Radiation Oncology, Niigata University Graduate School of Medical and Dental Sciences, Niigata, 951-8122, Japan
| |
Collapse
|
2
|
Yamamoto Y, Ohira S, Kanayama N, Inui S, Ueda Y, Koike Y, Miyazaki M, Nishio T, Koizumi M, Konishi K. Comparison of dosimetric parameters and robustness for rotational errors in fractionated stereotactic irradiation using automated noncoplanar volumetric modulated arc therapy for patients with brain metastases: single- versus multi-isocentric technique. Radiol Phys Technol 2023; 16:310-318. [PMID: 37093409 DOI: 10.1007/s12194-023-00720-7] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/19/2023] [Revised: 04/10/2023] [Accepted: 04/11/2023] [Indexed: 04/25/2023]
Abstract
To compare the dosimetric parameters of automated noncoplanar volumetric modulated arc therapy plans using single-isocentric (SIC) and multi-isocentric (MIC) techniques for patients with two brain metastases (BMs) in stereotactic irradiation and to evaluate the robustness of rotational errors. The SIC and MIC plans were retrospectively generated (35 Gy/five fractions) for 58 patients. Subsequently, a receiver operating characteristic curve analysis between the tumor surface distance (TSD) and V25Gy was performed to determine the thresholds for the brain tissue. The SIC and MIC plans were recalculated based on the rotational images to evaluate the dosimetric impact of rotational error. The MIC plans showed better brain tissue sparing for TSD > 6.6 cm. The SIC plans provided a significantly better conformity index for TSD ≤ 6.6 cm, while significantly lower gradient index was obtained (3.22 ± 0.56vs. 3.30 ± 0.57, p < 0.05) in the MIC plans with TSD > 6.6 cm. For organs at risk (OARs) (brainstem, chiasm, lens, optic nerves, and retinas), D0.1 cc was significantly lower (p < 0.05) in the MIC plans than in the SIC plans. The prescription dose could be delivered (D99%) to the gross tumor volume (GTV) for patients with TSD ≤ 6.6 cm when the rotational error was < 1°, whereas 31% of the D99% of GTV fell below the prescription dose with TSD > 6.6 cm. MIC plans can be an optimal approach for reducing doses to OARs and providing robustness against rotational errors in BMs with TSD > 6.6 cm.
Collapse
Affiliation(s)
- Yuki Yamamoto
- Department of Medical Physics and Engineering, Osaka University Graduate School of Medicine, Suita, Japan
| | - Shingo Ohira
- Department of Radiation Oncology, Osaka International Cancer Institute, 3-1-69 Otemae, Chuo-ku, Osaka, Osaka, 537-8567, Japan.
| | - Naoyuki Kanayama
- Department of Radiation Oncology, Osaka International Cancer Institute, 3-1-69 Otemae, Chuo-ku, Osaka, Osaka, 537-8567, Japan
| | - Shoki Inui
- Department of Radiation Oncology, Osaka International Cancer Institute, 3-1-69 Otemae, Chuo-ku, Osaka, Osaka, 537-8567, Japan
| | - Yoshihiro Ueda
- Department of Radiation Oncology, Osaka International Cancer Institute, 3-1-69 Otemae, Chuo-ku, Osaka, Osaka, 537-8567, Japan
| | - Yuhei Koike
- Department of Radiology, Kansai Medical University, Osaka, Japan
| | - Masayoshi Miyazaki
- Department of Radiation Oncology, Osaka International Cancer Institute, 3-1-69 Otemae, Chuo-ku, Osaka, Osaka, 537-8567, Japan
| | - Teiji Nishio
- Department of Medical Physics and Engineering, Osaka University Graduate School of Medicine, Suita, Japan
| | - Masahiko Koizumi
- Department of Medical Physics and Engineering, Osaka University Graduate School of Medicine, Suita, Japan
| | - Koji Konishi
- Department of Radiation Oncology, Osaka International Cancer Institute, 3-1-69 Otemae, Chuo-ku, Osaka, Osaka, 537-8567, Japan
| |
Collapse
|
3
|
Nakano H, Shiinoki T, Tanabe S, Nakano T, Takizawa T, Utsunomiya S, Sakai M, Tanabe S, Ohta A, Kaidu M, Nishio T, Ishikawa H. Multicomponent mathematical model for tumor volume calculation with setup error using single-isocenter stereotactic radiotherapy for multiple brain metastases. Phys Eng Sci Med 2023; 46:945-953. [PMID: 36940064 DOI: 10.1007/s13246-023-01241-8] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/30/2022] [Accepted: 03/06/2023] [Indexed: 03/21/2023]
Abstract
We evaluated the tumor residual volumes considering six degrees-of-freedom (6DoF) patient setup errors in stereotactic radiotherapy (SRT) with multicomponent mathematical model using single-isocenter irradiation for brain metastases. Simulated spherical gross tumor volumes (GTVs) with 1.0 (GTV 1), 2.0 (GTV 2), and 3.0 (GTV 3)-cm diameters were used. The distance between the GTV center and isocenter (d) was set at 0-10 cm. The GTV was simultaneously translated within 0-1.0 mm (T) and rotated within 0°-1.0° (R) in the three axis directions using affine transformation. We optimized the tumor growth model parameters using measurements of non-small cell lung cancer cell lines' (A549 and NCI-H460) growth. We calculated the GTV residual volume at the irradiation's end using the physical dose to the GTV when the GTV size, d, and 6DoF setup error varied. The d-values that satisfy tolerance values (10%, 35%, and 50%) of the GTV residual volume rate based on the pre-irradiation GTV volume were determined. The larger the tolerance value set for both cell lines, the longer the distance to satisfy the tolerance value. In GTV residual volume evaluations based on the multicomponent mathematical model on SRT with single-isocenter irradiation, the smaller the GTV size and the larger the distance and 6DoF setup error, the shorter the distance that satisfies the tolerance value might need to be.
Collapse
Affiliation(s)
- Hisashi Nakano
- Department of Radiation Oncology, Niigata University Medical and Dental Hospital, 1-757 Asahimachi-Dori, Chuo-Ku, Niigata-Shi, Niigata, Japan. .,Department of Medical Physics and Engineering, Osaka University Graduate School of Medicine, 1-7 Yamadaoka, Suita-Shi, Osaka, Japan.
| | - Takehiro Shiinoki
- Department of Radiation Oncology, Yamaguchi University, Minamikogushi 1-1-1 Ube, Yamaguchi, Japan
| | - Satoshi Tanabe
- Department of Radiation Oncology, Niigata University Medical and Dental Hospital, 1-757 Asahimachi-Dori, Chuo-Ku, Niigata-Shi, Niigata, Japan
| | - Toshimichi Nakano
- Department of Radiology and Radiation Oncology, Niigata University Graduate School of Medical and Dental Sciences, 1-757 Asahimachi-Dori, Chuo-Ku, Niigata-Shi, Niigata, Japan
| | - Takeshi Takizawa
- Department of Radiology and Radiation Oncology, Niigata University Graduate School of Medical and Dental Sciences, 1-757 Asahimachi-Dori, Chuo-Ku, Niigata-Shi, Niigata, Japan.,Department of Radiation Oncology, Niigata Neurosurgical Hospital, 3057 Yamada, Nishi-Ku, Niigata-Shi, Niigata, Japan
| | - Satoru Utsunomiya
- Department of Radiological Technology, Niigata University Graduate School of Health Sciences, 2-746 Asahimachi-Dori, Chuo-Ku, Niigata-Shi, Niigata, Japan
| | - Madoka Sakai
- Department of Radiation Oncology, Niigata University Medical and Dental Hospital, 1-757 Asahimachi-Dori, Chuo-Ku, Niigata-Shi, Niigata, Japan
| | - Shunpei Tanabe
- Department of Radiation Oncology, Niigata University Medical and Dental Hospital, 1-757 Asahimachi-Dori, Chuo-Ku, Niigata-Shi, Niigata, Japan
| | - Atsushi Ohta
- Department of Radiation Oncology, Niigata University Medical and Dental Hospital, 1-757 Asahimachi-Dori, Chuo-Ku, Niigata-Shi, Niigata, Japan
| | - Motoki Kaidu
- Department of Radiology and Radiation Oncology, Niigata University Graduate School of Medical and Dental Sciences, 1-757 Asahimachi-Dori, Chuo-Ku, Niigata-Shi, Niigata, Japan
| | - Teiji Nishio
- Department of Medical Physics and Engineering, Osaka University Graduate School of Medicine, 1-7 Yamadaoka, Suita-Shi, Osaka, Japan
| | - Hiroyuki Ishikawa
- Department of Radiology and Radiation Oncology, Niigata University Graduate School of Medical and Dental Sciences, 1-757 Asahimachi-Dori, Chuo-Ku, Niigata-Shi, Niigata, Japan
| |
Collapse
|
4
|
Ono T, Kido T, Nakamura M, Iramina H, Kakino R, Mizowaki T. Automatic measurement of beam-positioning accuracy at off-isocenter positions. J Appl Clin Med Phys 2023; 24:e13844. [PMID: 36420973 PMCID: PMC10018661 DOI: 10.1002/acm2.13844] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/07/2022] [Revised: 10/22/2022] [Accepted: 10/26/2022] [Indexed: 11/25/2022] Open
Abstract
PURPOSE This study performed an automatic measurement of the off-axis beam-positioning accuracy at a single isocenter via the TrueBeam Developer mode and evaluated the beam-positioning accuracy considering the effect of couch rotational errors. METHODS TrueBeam STx and the Winston-Lutz test-dedicated phantom, with a 3 mm diameter steel ball, were used in this study. The phantom was placed on the treatment couch, and the Winston-Lutz test was performed at the isocenter for four gantry angles (0°, 90°, 180°, and 270°) using an electronic portal imaging device. The phantom offset positions were at distances of 0, 25, 50, 75, and 100 mm from the isocenter along the superior-inferior, anterior-posterior, and left-right directions. Seventeen patterns of multileaf collimator-shaped square fields of 10 × 10 mm2 were created at the isocenter and off-axis positions for each gantry angle. The beam-positioning accuracy was evaluated with couch rotation along the yaw-axis (0°, ± 0.5°, and ± 1.0°). RESULTS The mean beam-positioning errors at the isocenter and off-isocenter distances (from the isocenter to ±100 mm) were 0.46-0.60, 0.44-0.91, and 0.42-1.11 mm for the couch angles of 0°, ±0.5°, and ±1°, respectively. The beam-positioning errors increased as the distance from the isocenter and couch rotation increased. CONCLUSION These findings suggest that the beam-positioning accuracy at the isocenter and off-isocenter positions can be evaluated quickly and automatically using the TrueBeam Developer mode. The proposed procedure is expected to contribute to an efficient evaluation of the beam-positioning accuracy at off-isocenter positions.
Collapse
Affiliation(s)
- Tomohiro Ono
- Department of Radiation Oncology and Image‐Applied TherapyKyoto UniversityKyotoJapan
| | - Takahisa Kido
- Department of Information Technology and Medical EngineeringHuman Health SciencesGraduate School of MedicineKyoto UniversityKyotoJapan
| | - Mitsuhiro Nakamura
- Department of Information Technology and Medical EngineeringHuman Health SciencesGraduate School of MedicineKyoto UniversityKyotoJapan
| | - Hiraku Iramina
- Department of Radiation Oncology and Image‐Applied TherapyKyoto UniversityKyotoJapan
| | - Ryo Kakino
- Kansai BNCT Medical Center, OsakaMedical and Pharmaceutical UniversityTakatsukiJapan
| | - Takashi Mizowaki
- Department of Radiation Oncology and Image‐Applied TherapyKyoto UniversityKyotoJapan
| |
Collapse
|
5
|
Tatsumi D. [2. Considerations for Radiation Treatment Planning from Medical Accelerators]. Nihon Hoshasen Gijutsu Gakkai Zasshi 2022; 78:526-530. [PMID: 35598962 DOI: 10.6009/jjrt.2022-2018] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 06/15/2023]
|
6
|
Nakano H, Tanabe S, Sasamoto R, Takizawa T, Utsunomiya S, Sakai M, Nakano T, Ohta A, Kaidu M, Ishikawa H. Radiobiological evaluation considering setup error on single-isocenter irradiation in stereotactic radiosurgery. J Appl Clin Med Phys 2021; 22:266-275. [PMID: 34151498 PMCID: PMC8292684 DOI: 10.1002/acm2.13322] [Citation(s) in RCA: 5] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/06/2020] [Revised: 05/19/2021] [Accepted: 05/24/2021] [Indexed: 11/06/2022] Open
Abstract
Purpose We calculated the dosimetric indices and estimated the tumor control probability (TCP) considering six degree‐of‐freedom (6DoF) patient setup errors in stereotactic radiosurgery (SRS) using a single‐isocenter technique. Methods We used simulated spherical gross tumor volumes (GTVs) with diameters of 1.0 cm (GTV 1), 2.0 cm (GTV 2), and 3.0 cm (GTV 3), and the distance (d) between the target center and isocenter was set to 0, 5, and 10 cm. We created the dose distribution by convolving the blur component to uniform dose distribution. The prescription dose was 20 Gy and the dose distribution was adjusted so that D95 (%) of each GTV was covered by 100% of the prescribed dose. The GTV was simultaneously rotated within 0°–1.0° (δR) around the x‐, y‐, and z‐axes and then translated within 0–1.0 mm (δT) in the x‐, y‐, and z‐axis directions. D95, conformity index (CI), and conformation number (CN) were evaluated by varying the distance from the isocenter. The TCP was estimated by translating the calculated dose distribution into a biological response. In addition, we derived the x‐y‐z coordinates with the smallest TCP reduction rate that minimize the sum of squares of the residuals as the optimal isocenter coordinates using the relationship between 6DoF setup error, distance from isocenter, and GTV size. Results D95, CI, and CN were decreased with increasing isocenter distance, decreasing GTV size, and increasing setup error. TCP of GTVs without 6DoF setup error was estimated to be 77.0%. TCP were 25.8% (GTV 1), 35.0% (GTV 2), and 53.0% (GTV 3) with (d, δT,δR) = (10 cm, 1.0 mm, 1.0°). The TCP was 52.3% (GTV 1), 54.9% (GTV 2), and 66.1% (GTV 3) with (d, δT,δR) = (10 cm, 1.0 mm, 1.0°) at the optimal isocenter position. Conclusion The TCP in SRS for multiple brain metastases with a single‐isocenter technique may decrease with increasing isocenter distance and decreasing GTV size when the 6DoF setup errors are exceeded (1.0 mm, 1.0°). Additionally, it might be possible to better maintain TCP for GTVs with 6DoF setup errors by using the optimal isocenter position.
Collapse
Affiliation(s)
- Hisashi Nakano
- Department of Radiation Oncology, Niigata University Medical and Dental Hospital, Niigata, Japan
| | - Satoshi Tanabe
- Department of Radiation Oncology, Niigata University Medical and Dental Hospital, Niigata, Japan
| | - Ryuta Sasamoto
- Department of Radiological Technology, Niigata University Graduate School of Health Sciences, Niigata, Japan
| | - Takeshi Takizawa
- Department of Radiation Oncology, Niigata University Medical and Dental Hospital, Niigata, Japan.,Department of Radiation Oncology, Niigata Neurosurgical Hospital, Niigata, Japan
| | - Satoru Utsunomiya
- Department of Radiological Technology, Niigata University Graduate School of Health Sciences, Niigata, Japan
| | - Madoka Sakai
- Department of Radiation Oncology, Niigata University Medical and Dental Hospital, Niigata, Japan
| | - Toshimichi Nakano
- Department of Radiology and Radiation Oncology, Niigata University Graduate School of Medical and Dental Sciences, Niigata, Japan
| | - Atsushi Ohta
- Department of Radiation Oncology, Niigata University Medical and Dental Hospital, Niigata, Japan
| | - Motoki Kaidu
- Department of Radiology and Radiation Oncology, Niigata University Graduate School of Medical and Dental Sciences, Niigata, Japan
| | - Hiroyuki Ishikawa
- Department of Radiology and Radiation Oncology, Niigata University Graduate School of Medical and Dental Sciences, Niigata, Japan
| |
Collapse
|