Strobeck JE, Shen JT, Singh B, Obunai K, Miceli C, Sacher H, Ritucci F, Imhoff M. Comparison of a two-lead, computerized, resting ECG signal analysis device, the MultiFunction-CardioGram or MCG (a.k.a. 3DMP), to quantitative coronary angiography for the detection of relevant coronary artery stenosis (>70%) - a meta-analysis of all published trials performed and analyzed in the US.
Int J Med Sci 2009;
6:143-55. [PMID:
19381351 PMCID:
PMC2669600 DOI:
10.7150/ijms.6.143]
[Citation(s) in RCA: 6] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 01/19/2009] [Accepted: 04/06/2009] [Indexed: 11/25/2022] Open
Abstract
BACKGROUND
Accurate, non-invasive diagnosis of, and screening for, coronary artery disease (CAD) and restenosis after coronary revascularization has been a challenge due to either low sensitivity/specificity or relevant morbidity associated with current diagnostic modalities.
METHODS
To assess sensitivity and specificity of a new computerized, multiphase, resting electrocardiogram analysis device (MultiFunction-CardioGram(sm) or MCG a.k.a. 3DMP) for the detection of relevant coronary stenosis (>70%), a meta-analysis of three published prospective trials performed in the US on patient data collected using the US manufactured device and analyzed using the US-based software and New York data analysis center from patients in the US, Germany, and Asia was completed. A total of 1076 patients from the three trials (US - 136; Germany - 751; Asia - 189) (average age 62 +/- 11.5, 65 for women, 60 for men) scheduled for coronary angiography, were included in the analysis. Patients enrolled in the trials may or may not have had prior angiography and/or coronary intervention. Angiographic results in all studies were classified for hemodynamically relevant stenosis (> 70%) by two US based angiographers independently.
RESULTS
Hemodynamically relevant stenosis was diagnosed in 467 patients (43.4%). The device, after performing a frequency-domain, computational analysis of the resting ECG leads and computer-database comparison, calculated a coronary ischemia "severity" score from 0 to 20 for each patient. The severity score was significantly higher for patients with relevant coronary stenosis (5.4 +/- 1.8 vs. 1.7 +/- 2.1). The study device (using a cut-off score for relevant stenosis of 4.0) correctly classified 941 of the 1076 patients with or without relevant stenosis (sensitivity-91.2%; specificity-84.6%; NPV 0.942, PPV 0.777). Adjusted positive and negative predictive values (PPV and NPV) were 81.9% and 92.6%, respectively (ROC AUC = 0.881 [95% CI: 0.860-0.903]). Subgroup analysis showed no significant influence of sex, age, race/nationality, previous revascularization procedures, resting ECG morphology, or participating center on the device's diagnostic performance.
CONCLUSIONS
The new computerized, multiphase, resting ECG analysis device (MultiFunction-CardioGram(sm)) has been shown in this meta-analysis to safely and accurately identify patients with relevant coronary stenosis (>70%) with high sensitivity and specificity and high negative predictive value. Its potential use in the evaluation of symptomatic patients suspected to suffer from coronary disease/ischemia is discussed.
Collapse