Pandeya GD, Greuter MJW, Schmidt B, Flohr T, Oudkerk M. Assessment of thermal sensitivity of CT during heating of liver: an ex vivo study.
Br J Radiol 2012;
85:e661-5. [PMID:
22919016 DOI:
10.1259/bjr/23942179]
[Citation(s) in RCA: 13] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/19/2022] Open
Abstract
OBJECTIVES
The purpose of this study was to assess the thermal sensitivity of CT during heating of ex-vivo animal liver.
METHODS
Pig liver was indirectly heated from 20 to 90 °C by passage of hot air through a plastic tube. The temperature in the heated liver was measured using calibrated thermocouples. In addition, image acquisition was performed with a multislice CT scanner before and during heating of the liver sample. The reconstructed CT images were then analysed to assess the change of CT number as a function of temperature.
RESULTS
During heating, a decrease in CT numbers was observed as a hypodense area on the CT images. In addition, the hypodense area extended outward from the heat source during heating. The analysis showed a linear decrease of CT number as a function of temperature. From this relationship, we derived a thermal sensitivity of CT for pig liver tissue of -0.54±0.03 HU °C(-1) with an r(2) value of 0.91.
CONCLUSIONS
The assessment of the thermal sensitivity of CT in ex-vivo pig liver tissue showed a linear dependency on temperature ≤90 °C. This result may be beneficial for the application of isotherms or thermal maps in CT images of liver tissue.