Bochud FO, Laedermann JP, Baechler S, Kosinski M, Bailat CJ. Usefulness of specific calibration coefficients for gamma-emitting sources measured by radionuclide calibrators in nuclear medicine.
Med Phys 2011;
38:4073-80. [PMID:
21859006 DOI:
10.1118/1.3596528]
[Citation(s) in RCA: 8] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
PURPOSE
In nuclear medicine, the activity of a radionuclide is measured with a radionuclide calibrator that often has a calibration coefficient independent of the container type and filling.
METHODS
To determine the effect of the container on the accuracy of measuring the activity injected into a patient, The authors simulated a commercial radionuclide calibrator and 18 container types most typically used in clinical practice. The instrument sensitivity was computed for various container thicknesses and filling levels. Monoenergetic photons and electrons as well as seven common radionuclides were considered.
RESULTS
The quality of the simulation with gamma-emitting sources was validated by an agreement with measurements better than 4% in five selected radionuclides. The results show that the measured activity can vary by more than a factor of 2 depending on the type of container. The filling level and the thickness of the container wall only have a marginal effect for radionuclides of high energy but could induce differences up to 4%.
CONCLUSIONS
The authors conclude that radionuclide calibrators should be tailored to the uncertainty required by clinical applications. For most clinical cases, and at least for the low-energy gamma and x-ray emitters, measurements should be performed with calibration coefficients specific to the container type.