1
|
Ichikawa S, Muto H, Imao M, Nonaka T, Sakekawa K, Sato Y. Low-dose whole-spine imaging using slot-scan digital radiography: a phantom study. BMC Med Imaging 2023; 23:17. [PMID: 36710344 PMCID: PMC9885656 DOI: 10.1186/s12880-023-00971-1] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/24/2022] [Accepted: 01/16/2023] [Indexed: 01/31/2023] Open
Abstract
BACKGROUND Slot-scan digital radiography (SSDR) is equipped with detachable scatter grids and a variable copper filter. In this study, this function was used to obtain parameters for low-dose imaging for whole-spine imaging. METHODS With the scatter grid removed and the beam-hardening (BH) filters (0.0, 0.1, 0.2, or 0.3 mm) inserted, the tube voltage (80, 90, 100, 110, or 120 kV) and the exposure time were adjusted to 20 different parameters that produce equivalent image quality. Slot-scan radiographs of an acrylic phantom were acquired with the set parameters, and the optimal parameters (four types) for each filter were determined using the figure of merit. For the four types of parameters obtained in the previous section, SSDR was performed on whole-spine phantoms by varying the tube current, and the parameter with the lowest radiation dose was determined by visual evaluation. RESULTS The parameters for each filter according to the FOM results were 90 kV, 400 mA, and 2.8 ms for 0.0 mm thickness; 100 kV, 400 mA, and 2.0 ms for 0.1 mm thickness; 100 kV, 400 mA, and 2.8 ms for 0.2 mm thickness; and 110 kV, 400 mA, and 2.2 ms for 0.3 mm thickness. Visual evaluation of the varying tube currents was performed using these four parameters when the BH filter thicknesses were 0.0, 0.1, 0.2, and 0.3 mm. The entrance surface dose was 59.44 µGy at 90 kV, 125 mA, and 2.8 ms; 57.39 µGy at 100 kV, 250 mA, and 2.0 ms; 46.89 µGy at 100 kV, 250 mA, and 2.8 ms; and 39.48 µGy at 110 kV, 250 mA, and 2.2 ms, indicating that the 0.3-mm BH filter was associated with the minimum dose. CONCLUSION Whole-spine SSDR could reduce the dose by 79% while maintaining the image quality.
Collapse
Affiliation(s)
- Shigeji Ichikawa
- grid.412879.10000 0004 0374 1074Suzuka University of Medical Science, Graduate School of Health Science Division of Health Science, 1001-1,Kishioka, Suzuka, Mie 510-0293 Japan ,grid.412879.10000 0004 0374 1074Graduate School of Health Science, Suzuka University of Medical Science, 1001-1, Kishioka, Suzuka, Mie 510-0293 Japan
| | - Hiroe Muto
- grid.412879.10000 0004 0374 1074Suzuka University of Medical Science, Graduate School of Health Science Division of Health Science, 1001-1,Kishioka, Suzuka, Mie 510-0293 Japan
| | - Masashi Imao
- Department of Radiology, Faculty of Health Science, Gunma Paz University, 1-7-1 Tonyamachi, Takasaki, Gunma 370-0006 Japan
| | - Takashi Nonaka
- Department of Radiological Technology, Fussa Hospital, 1-6-1 Kamidaira, Fussa-ku, Tokyo, 197-0012 Japan
| | - Kouji Sakekawa
- Department of Radiological Technology, Fussa Hospital, 1-6-1 Kamidaira, Fussa-ku, Tokyo, 197-0012 Japan
| | - Yasutaka Sato
- Department of Radiological Technology, Fussa Hospital, 1-6-1 Kamidaira, Fussa-ku, Tokyo, 197-0012 Japan
| |
Collapse
|
2
|
Ngaile JE, Msaki PK, Suleiman SA, Chuma FM, Muhogora WE, Jusabani AM. ESTIMATION OF PATIENT RADIATION DOSES FROM MULTI-DETECTOR COMPUTED TOMOGRAPHY ANGIOGRAPHY PROCEDURES IN TANZANIA. RADIATION PROTECTION DOSIMETRY 2021; 197:63-77. [PMID: 34729602 DOI: 10.1093/rpd/ncab159] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 05/02/2021] [Revised: 09/08/2021] [Accepted: 10/05/2021] [Indexed: 06/13/2023]
Abstract
The aim of the present study was to estimate the volume CT dose index (CTDIvol), dose length product (DLP) and effective dose (ED) to patients from five multi-detector computed tomography angiography (MDCTA) procedures: brain, carotid, coronary, entire aorta and lower limb from four medical institutions in Tanzania; to compare these doses to those reported in the literature, and to compare the data obtained with ICRP 103 and Monte Carlo software. The radiation doses for 217 patients were estimated using patient demographics, patient-related exposure parameters, the geometry of examination and CT-Expo V 2.4 Monte Carlo-based software. The median values of the CTDIvol, DLP and ED for MDCTA procedures of the brain and carotids were 36.8 mGy, 1481.0 mGy∙cm and 5.2 mSv, and 15.9 mGy, 1224.0 mGy∙cm and 7.8 mSv, respectively; while for the coronary, entire aortic, and lower limbs were 49.4 mGy, 1493.0 mGy∙cm and 30.6 mSv; 16.2 mGy, 2287.0 mGy∙cm and 41.1 mSv; and 6.4 mGy, 1406.0 mGy∙cm and 10.5 mSv, respectively. The ratio of the maximum to minimum ED values to individual patients across the four medical centers were 41.4, 11.1, 4.6, 9.5 and 37.4, respectively, for the brain, carotid, coronary, entire aortic and lower limb CT angiography procedures. The mean values of CTDIvol, DLP and ED in the present study were typically higher than the values reported from Kenya, Korea and Saudi Arabia. The 75th percentile values of the DLP were above the preliminary diagnostic references levels proposed by Kenya, Switzerland and Korea. The observed wide range of examination scanning protocols and patient doses for similar MDCTA procedures within and across hospitals; and the observed relatively high patient doses compared to those reported in the literature, call for the need to standardize scanning protocols and optimise patient dose from MDCTA procedures.
Collapse
Affiliation(s)
- J E Ngaile
- Radiation Control Directorate, Tanzania Atomic Energy Commission, P O Box 743 Arusha, Tanzania
- Department of Physics, University of Dar es Salaam, P O Box 35063 Dar es Salaam, Tanzania
| | - P K Msaki
- Department of Physics, University of Dar es Salaam, P O Box 35063 Dar es Salaam, Tanzania
| | - S A Suleiman
- Radiation Control Directorate, Tanzania Atomic Energy Commission, P O Box 743 Arusha, Tanzania
| | - F M Chuma
- Radiation Control Directorate, Tanzania Atomic Energy Commission, P O Box 743 Arusha, Tanzania
| | - W E Muhogora
- Radiation Control Directorate, Tanzania Atomic Energy Commission, P O Box 743 Arusha, Tanzania
| | - A M Jusabani
- Department of Radiology, The Agha Khan Hospital, P O Box 2289 Dar es Salaam, Tanzania
| |
Collapse
|
3
|
Jung S, Kim JI, Park JM, Wu HG, Choi CH. Gold coated contact lens-type ocular in vivo dosimeter (CLOD) for monitoring of low dose in computed tomography: A Monte Carlo study. Phys Med 2021; 92:1-7. [PMID: 34781119 DOI: 10.1016/j.ejmp.2021.10.016] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 04/04/2021] [Revised: 10/06/2021] [Accepted: 10/28/2021] [Indexed: 10/19/2022] Open
Abstract
PURPOSE This study reports a sensitivity enhancement of gold-coated contact lens-type ocular in vivo dosimeters (CLODs) for low-dose measurements in computed tomography (CT). METHODS Monte Carlo (MC) simulations were conducted to evaluate the dose enhancement from the gold (Au) layers on the CLODs. The human eye and CLODs were modeled, and the X-ray tube voltages were defined as 80, 120, and 140 kVp. The thickness of the Au layer attached to a CLOD ranged from 100 nm to 10 μm. The thickness of the active layer ranged from 20 to 140 μm. The dose ratio between the active layer of the Au-coated CLOD and a CLOD without a layer, i.e., the dose enhancement factor (DEF), was calculated. RESULTS The DEFs of the first 20-μm thick active layer of the 5-μm thick Au-coated CLOD were 18.4, 19.7, 20.2 at 80, 120, and 140 kVp, respectively. The DEFs decreased as the thickness of the active layer increased. The DEFs of 100-nm to 5-μm thick Au layers increased from 1.7 to 5.4 for 120-kVp X-ray tube voltage when the thickness of the active layer was 140 μm. CONCLUSIONS The MC results presented a higher sensitivity of Au-coated CLODs (∼20-times higher than that of CLODs without a gold layer). Au-coated CLODs can be applied to an evaluation of very low doses (a few cGy) delivered to patients during CT imaging.
Collapse
Affiliation(s)
- Seongmoon Jung
- Department of Radiation Oncology, Seoul National University Hospital, Seoul 03080, Republic of Korea; Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul 03080, Republic of Korea; Biomedical Research Institute, Seoul National University Hospital, Seoul 03080, Republic of Korea
| | - Jung-In Kim
- Department of Radiation Oncology, Seoul National University Hospital, Seoul 03080, Republic of Korea; Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul 03080, Republic of Korea; Biomedical Research Institute, Seoul National University Hospital, Seoul 03080, Republic of Korea
| | - Jong Min Park
- Department of Radiation Oncology, Seoul National University Hospital, Seoul 03080, Republic of Korea; Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul 03080, Republic of Korea; Biomedical Research Institute, Seoul National University Hospital, Seoul 03080, Republic of Korea; Department of Radiation Oncology, Seoul National University College of Medicine, Seoul 03080, Republic of Korea; Robotics Research Laboratory for Extreme Environments, Advanced Institute of Convergence Technology, Suwon 16229, Republic of Korea
| | - Hong-Gyun Wu
- Department of Radiation Oncology, Seoul National University Hospital, Seoul 03080, Republic of Korea; Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul 03080, Republic of Korea; Biomedical Research Institute, Seoul National University Hospital, Seoul 03080, Republic of Korea; Cancer Research Institute, Seoul National University College of Medicine, Seoul 03080, Republic of Korea; Department of Radiation Oncology, Seoul National University College of Medicine, Seoul 03080, Republic of Korea
| | - Chang Heon Choi
- Department of Radiation Oncology, Seoul National University Hospital, Seoul 03080, Republic of Korea; Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul 03080, Republic of Korea; Biomedical Research Institute, Seoul National University Hospital, Seoul 03080, Republic of Korea.
| |
Collapse
|
4
|
Agrawal AK, Singh B, Singhai P, Kashyap Y, Shukla M. The white beam station at imaging beamline BL-4, Indus-2. JOURNAL OF SYNCHROTRON RADIATION 2021; 28:1639-1648. [PMID: 34475311 DOI: 10.1107/s160057752100775x] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Subscribe] [Scholar Register] [Received: 02/09/2021] [Accepted: 07/28/2021] [Indexed: 06/13/2023]
Abstract
The high flux density of synchrotron white beam offers several advantages in X-ray imaging such as higher resolution and signal-to-noise ratio in 3D/4D micro-tomography, higher frame rate in real-time imaging of transient phenomena, and higher penetration in thick and dense materials especially at higher energies. However, these advantages come with additional challenges to beamline optics, camera and sample due to increased heat load and radiation damage, and to personal safety due to higher radiation dose and ozone gas hazards. In this work, a white beam imaging facility at imaging beamline BL-4, Indus-2, has been developed, while taking care of various instrumental and personal safety challenges. The facility has been tested to achieve 1.5 µm spatial resolution, increased penetration depth up to 900 µm in steel, and high temporal resolutions of ∼10 ms (region of interest 2048 × 2048 pixels) and 70 µs (256 × 2048 pixels). The facility is being used successfully for X-ray imaging, non-destructive testing and dosimetry experiments.
Collapse
Affiliation(s)
- Ashish K Agrawal
- Technical Physics Division, Bhabha Atomic Research Centre, Trombay, Mumbai 400085, India
| | - Balwant Singh
- Technical Physics Division, Bhabha Atomic Research Centre, Trombay, Mumbai 400085, India
| | - Payal Singhai
- Technical Physics Division, Bhabha Atomic Research Centre, Trombay, Mumbai 400085, India
| | - Yogesh Kashyap
- Technical Physics Division, Bhabha Atomic Research Centre, Trombay, Mumbai 400085, India
| | - Mayank Shukla
- Technical Physics Division, Bhabha Atomic Research Centre, Trombay, Mumbai 400085, India
| |
Collapse
|
5
|
Shapira N, Scheuermann J, Perkins AE, Kim J, Liu LP, Karp JS, Noël PB. Quantitative positron emission tomography imaging in the presence of iodinated contrast media using electron density quantifications from dual-energy computed tomography. Med Phys 2020; 48:273-286. [PMID: 33170953 DOI: 10.1002/mp.14589] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/12/2020] [Revised: 07/31/2020] [Accepted: 11/02/2020] [Indexed: 11/10/2022] Open
Abstract
PURPOSE As preparation for future positron emission tomography (PET)/dual-energy computed tomography (DECT)T imaging modality and new possible clinical applications, the study aimed to evaluate the utility of clinically available spectral results from a DECT system for improving attenuation corrections of PET acquisitions in the presence of iodinated contrast media. The dependence of the accuracy of PET quantification values, reconstructed with conventional and spectral-based attenuation corrections, was examined as a function of the amount of iodine content and x-ray radiation exposure. METHODS Measurements were performed on commercial PET/CT and DECT systems, using a semi-anthropomorphic phantom with seven centrifuge tubes in its bore. Five different configurations of tube contents were scanned by both PET/CT and DECT. With the aim of mimicking clinically observed concentrations, in all phantom configurations the center tube contained a high concentration of radionuclide while the peripheral tubes contained a lower concentration of radionuclide. Iodine content was incrementally increased between phantom configurations by replacing iodine-free tubes with tubes that contained the original radionuclide concentration within a 10 mg/ml iodine dilution. DECT-based attenuation correction maps were generated by scaling electron density spectral results into corresponding 511 keV photon linear attenuation coefficients. RESULTS Mean SUV values obtained from the nominal PET reconstruction, using conventional CT images as input for the attenuation correction, demonstrate a monotonic increase of 8.6% when the water and radionuclide mixtures were replaced by iodine, water, and radionuclide (same level of activity) mixture. Mean SUV values obtained from the DECT-based reconstruction, in which the attenuation correction utilizes electron density values as input, demonstrate different, more stable behavior across all iodine insert configurations, with a standard deviation to mean ratio of less than 1%. This observed behavior was independent of the area size used for measurement. A minor radiation dose dependency of the electron density values (below 0.5%) was observed. This resulted in consistent (iodine independent) PET quantification behavior, which persisted even at the lowest radiation dose levels tested in our experiment, that is, 25% of the radiation dose utilized for CT acquisition in the clinical PET/CT protocol. CONCLUSIONS Utilization of DECT-generated electron density estimations for attenuation correction benefit PET quantification consistency in the presence of iodine and at nominal and low DECT radiation exposure levels. The ability to correctly account for iodinated contrast media in PET acquisitions will allow the development of new clinical applications that rely on the quantitative capabilities of spectral CT technologies and modern PET systems.
Collapse
Affiliation(s)
- Nadav Shapira
- Department of Radiology, University of Pennsylvania, Philadelphia, PA, USA
| | - Joshua Scheuermann
- Department of Radiology, University of Pennsylvania, Philadelphia, PA, USA
| | | | - Johoon Kim
- Department of Radiology, University of Pennsylvania, Philadelphia, PA, USA
| | - Leening P Liu
- Department of Radiology, University of Pennsylvania, Philadelphia, PA, USA
| | - Joel S Karp
- Department of Radiology, University of Pennsylvania, Philadelphia, PA, USA
| | - Peter B Noël
- Department of Radiology, University of Pennsylvania, Philadelphia, PA, USA.,Department of Diagnostic and Interventional Radiology, School of Medicine & klinikum rechts der Isar, Technical University of Munich, München, Germany
| |
Collapse
|
6
|
Gray T, Bassiri N, David S, Patel DY, Abdul-Moqueet M, Kirby N, Mayer KM. Characterization and comparison of imaging contrast enhancement with PEG-functionalized gold nanoparticles in kV cone beam computed tomography and computed tomography imaging. Biomed Phys Eng Express 2020; 6:047002. [PMID: 33444284 DOI: 10.1088/2057-1976/ab9207] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
The purpose of this study is to define a simplified method to accurately predict and characterize kV cone beam computed tomography (kV CBCT) and computed tomography (CT) image contrast enhancement from gold nanoparticles (GNPs). Parameters of the kV CBCT of a Varian Novalis Tx linear accelerator and of a GE LightSpeed 4 Big Bore CT machine were modeled using the MCNP 6.2 Monte Carlo code. A 0.25 × 0.25 cm2 source, defined with a 100 kVp energy spectrum with appropriate filtration, was implemented in the MCNP6.2 model for kV CBCT, which also contained x- and y-blades and a full bowtie filter. A 1 cm3 cube of GNP solution (modeled as a mass percentage of gold in water) was placed 100 cm below the source. For the CT-simulator model, a source was defined with energy spectra for 80 and 140 kVp x-rays with appropriate filtration and angular spectrum. A 1 cm3 GNP solution was modeled as before and a detector was placed 40 cm below that. Attenuation coefficients of four GNP solutions were computed and Hounsfield unit (HU) values were calculated. The computed HU values were compared against experimentally measured values obtained by scanning batches of GNPs of various sizes and concentrations using a GE LightSpeed 4 Big Bore CT scanner at 80 kVp and 140 kVp energies, as well as the kV CBCT capability of a Varian Novalis Tx linear accelerator. HU analysis was carried out using Velocity Medical Solutions clinical CT image analysis software. The MCNP calculated HU values matched the measured values to within ± 5%. Image contrast enhancement analysis showed a total increase in HU of up to 223. The sample having the highest gold mass percentage tested showed the greatest increase in HU number compared to water.
Collapse
Affiliation(s)
- Tara Gray
- The University of Texas at San Antonio, Department of Physics and Astronomy, San Antonio, TX 78249, United States of America
| | | | | | | | | | | | | |
Collapse
|
7
|
Assessment and comparison of radiation dose and image quality in multi-detector CT scanners in non-contrast head and neck examinations. Pol J Radiol 2019; 84:e61-e67. [PMID: 31019596 PMCID: PMC6479057 DOI: 10.5114/pjr.2019.82743] [Citation(s) in RCA: 13] [Impact Index Per Article: 2.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/12/2018] [Accepted: 01/09/2019] [Indexed: 01/08/2023] Open
Abstract
Purpose To assess and compare radiation dose and image quality from non-contrast head and neck computed tomography (CT) examinations from four different multi-detector CT (MDCT) scanners. Material and methods Four CT scanners with different numbers of detector rows including one 4-MDCT, a 6-MDCT, a 16-MDCT, and a 64-MDCT were investigated. Common CT dose descriptors including volumetric CT dose index (CTDIv), dose length product (DLP), and the effective dose (ED), and image quality parameters include image noise, uniformity, and spatial resolution (SR) were estimated for each CT scanner with standard tools and methods. To have a precise comparison between CT scanners and related doses and image quality parameters, the ImPACT Q-factor was used. Results Minimum and maximum CTDIv, DLP, and ED in the head scan were 18 ± 3 and 49 ± 4 mGy, 242 ± 28 and 692 ± 173 mGy × cm, 0.46 ± 0.4 and 1.31 ± 0.33 mSv for 16-MDCT and 64-MDCT, respectively. And 16 ± 2 to 27 ± 3, 286 ± 127 to 645 ± 79 and 1.46 ± 0.65 to 3.29 ± 0.40 for neck scan, respectively. The Q-factor in head scan was 2.4, 3.3, 4.4 and 5.6 for 4-MDCT, 6-MDCT, 16-MDCT and 64-MDCT, respectively. The Q-factor in neck scan was 3.4, 4.6, 4.7 and 6.0 for 4-MDCT, 6-MDCT, 16-MDCT and 64-MDCT, respectively. Conclusions The results clearly indicate an increasing trend in the Q-factor from 4-MDCT to 64-MDCT units in both head and neck examinations. This increasing trend is due to a better SR and less noise of images taken and/or fewer doses in 64-MDCT.
Collapse
|
8
|
Tse JJ, Dunmore-Buyze J, Drangova M, Holdsworth DW. Dual-energy computed tomography using a gantry-based preclinical cone-beam microcomputed tomography scanner. J Med Imaging (Bellingham) 2018; 5:033503. [PMID: 30155511 PMCID: PMC6103383 DOI: 10.1117/1.jmi.5.3.033503] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/30/2017] [Accepted: 07/30/2018] [Indexed: 11/14/2022] Open
Abstract
Dual-energy microcomputed tomography (DECT) can provide quantitative information about specific materials of interest, facilitating automated segmentation, and visualization of complex three-dimensional tissues. It is possible to implement DECT on currently available preclinical gantry-based cone-beam micro-CT scanners; however, optimal decomposition image quality requires customized spectral shaping (through added filtration), optimized acquisition protocols, and elimination of misregistration artifacts. We present a method for the fabrication of customized x-ray filters-in both shape and elemental composition-needed for spectral shaping. Fiducial markers, integrated within the sample holder, were used to ensure accurate co-registration between sequential low- and high-energy image volumes. The entire acquisition process was automated through the use of a motorized filter-exchange mechanism. We describe the design, implementation, and evaluation of a DECT system on a gantry-based-preclinical cone-beam micro-CT scanner.
Collapse
Affiliation(s)
- Justin J Tse
- Western University, Bone and Joint Institute, Imaging Research Laboratories, Robarts Research Institute, London, Ontario, Canada.,Western University, Bone and Joint Institute, Departments of Medical Biophysics and Medical Imaging, London, Ontario, Canada
| | - Joy Dunmore-Buyze
- Western University, Bone and Joint Institute, Imaging Research Laboratories, Robarts Research Institute, London, Ontario, Canada
| | - Maria Drangova
- Western University, Bone and Joint Institute, Imaging Research Laboratories, Robarts Research Institute, London, Ontario, Canada.,Western University, Bone and Joint Institute, Departments of Medical Biophysics and Medical Imaging, London, Ontario, Canada
| | - David W Holdsworth
- Western University, Bone and Joint Institute, Imaging Research Laboratories, Robarts Research Institute, London, Ontario, Canada.,Western University, Bone and Joint Institute, Departments of Medical Biophysics and Medical Imaging, London, Ontario, Canada.,Western University, Bone and Joint Institute, Department of Surgery, London, Ontario, Canada
| |
Collapse
|
9
|
Rui X, Jin Y, FitzGerald PF, Wu M, Alessio AM, Kinahan PE, De Man B. Fast analytical approach of application specific dose efficient spectrum selection for diagnostic CT imaging and PET attenuation correction. Phys Med Biol 2016; 61:7787-7811. [PMID: 27754977 DOI: 10.1088/0031-9155/61/21/7787] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/25/2022]
Abstract
Computed tomography (CT) has been used for a variety of applications, two of which include diagnostic imaging and attenuation correction for PET or SPECT imaging. Ideally, the x-ray tube spectrum should be optimized for the specific application to minimize the patient radiation dose while still providing the necessary information. In this study, we proposed a projection-based analytic approach for the analysis of contrast, noise, and bias. Dose normalized contrast to noise ratio (CNRD), inverse noise normalized by dose (IND) and bias are used as evaluation metrics to determine the optimal x-ray spectrum. Our simulation investigated the dose efficiency of the x-ray spectrum ranging from 40 kVp to 200 kVp. Water cylinders with diameters of 15 cm, 24 cm, and 35 cm were used in the simulation to cover a variety of patient sizes. The effects of electronic noise and pre-patient copper filtration were also evaluated. A customized 24 cm CTDI-like phantom with 13 mm diameter inserts filled with iodine (10 mg ml-1), tantalum (10 mg ml-1), water, and PMMA was measured with both standard (1.5 mGy) and ultra-low (0.2 mGy) dose to verify the simulation results at tube voltages of 80, 100, 120, and 140 kVp. For contrast-enhanced diagnostic imaging, the simulation results indicated that for high dose without filtration, the optimal kVp for water contrast is approximately 100 kVp for a 15 cm water cylinder. However, the 60 kVp spectrum produces the highest CNRD for bone and iodine. The optimal kVp for tantalum has two selections: approximately 50 and 100 kVp. The kVp that maximizes CNRD increases when the object size increases. The trend in the CTDI phantom measurements agrees with the simulation results, which also agrees with previous studies. Copper filtration improved the dose efficiency for water and tantalum, but reduced the iodine and bone dose efficiency in a clinically-relevant range (70-140 kVp). Our study also shows that for CT-based attenuation correction applications for PET or SPECT, a higher-kVp spectrum with copper filtration is preferable. This method is developed based on filter back projection and does not require image reconstruction or Monte Carlo dose estimates; thus, it could potentially be used for patient-specific and task-based on-the-fly protocol optimization.
Collapse
Affiliation(s)
- Xue Rui
- Image Reconstruction Laboratory, GE Global Research Center, Niskayuna, NY, USA
| | | | | | | | | | | | | |
Collapse
|
10
|
Tian X, Yin Z, De Man B, Samei E. Estimation of Radiation Dose in CT Based on Projection Data. J Digit Imaging 2016; 29:615-21. [PMID: 26893140 DOI: 10.1007/s10278-016-9869-x] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 02/04/2023] Open
Abstract
Managing and optimizing radiation dose has become a core problem for the CT community. As a fundamental step for dose optimization, accurate and computationally efficient dose estimates are crucial. The purpose of this study was to devise a computationally efficient projection-based dose metric. The absorbed energy and object mass were individually modeled using the projection data. The absorbed energy was estimated using the difference between intensity of the primary photon and the exit photon. The mass was estimated using the volume under the attenuation profile. The feasibility of the approach was evaluated across phantoms with a broad size range, various kVp settings, and two bowtie filters, using a simulation tool, the Computer Assisted Tomography SIMulator (CATSIM) software. The accuracy of projection-based dose estimation was validated against Monte Carlo (MC) simulations. The relationship between projection-based dose metric and MC dose estimate was evaluated using regression models. The projection-based dose metric showed a strong correlation with Monte Carlo dose estimates (R (2) > 0.94). The prediction errors for the projection-based dose metric were all below 15 %. This study demonstrated the feasibility of computationally efficient dose estimation requiring only the projection data.
Collapse
Affiliation(s)
- Xiaoyu Tian
- Department of Biomedical Engineering, Carl E. Ravin Advanced Imaging Laboratories, Department of Radiology, Duke University Medical Center, 2424 Erwin Road, Suite 302, Durham, NC, 27705, USA
| | - Zhye Yin
- CT Systems and Applications Laboratory, GE Global Research, Niskayuna, NY, 12309, USA
| | - Bruno De Man
- CT Systems and Applications Laboratory, GE Global Research, Niskayuna, NY, 12309, USA
| | - Ehsan Samei
- Department of Biomedical Engineering, Carl E. Ravin Advanced Imaging Laboratories, Department of Radiology, Duke University Medical Center, 2424 Erwin Road, Suite 302, Durham, NC, 27705, USA. .,Medical Physics Graduate Program, Departments of Radiology, Electrical and Computer Engineering, and Physics, Duke University Medical Center, Durham, NC, 27705, USA.
| |
Collapse
|
11
|
Rana N, Rawat D, Parmar M, Dhawan DK, Bhati AK, Mittal BR. Evaluation of external beam hardening filters on image quality of computed tomography and single photon emission computed tomography/computed tomography. J Med Phys 2016; 40:198-206. [PMID: 26865755 PMCID: PMC4728890 DOI: 10.4103/0971-6203.170790] [Citation(s) in RCA: 9] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/04/2022] Open
Abstract
This study was undertaken to evaluate the effect of external metal filters on the image quality of computed tomography (CT) and single photon emission computed tomography (SPECT)/CT images. Images of Jaszack phantom filled with water and containing iodine contrast filled syringes were acquired using CT (120 kV, 2.5 mA) component of SPECT/CT system, ensuring fixation of filter on X-ray collimator. Different thickness of filters of Al and Cu (1 mm, 2 mm, 3 mm, and 4 mm) and filter combinations Cu 1 mm, Cu 2 mm, Cu 3 mm each in combination with Al (1 mm, 2 mm, 3 mm, and 4 mm), respectively, were used. All image sets were visually analyzed for streak artifacts and contrast to noise ratio (CNR) was derived. Similar acquisition was done using Philips CT quality control (QC) phantom and CNR were calculated for its lexan, perspex, and teflon inserts. Attenuation corrected SPECT/CT images of Jaszack phantom filled with 444-555 MBq (12-15 mCi) of (99m)Tc were obtained by applying attenuation correction map generated by hardened X-ray beam for different filter combination, on SPECT data. Uniformity, root mean square (rms) and contrast were calculated in all image sets. Less streak artifacts at iodine water interface were observed in images acquired using external filters as compared to those without a filter. CNR for syringes, spheres, and inserts of Philips CT QC phantom was almost similar to Al 2 mm, Al 3 mm, and without the use of filters. CNR decreased with increasing copper thickness and other filter combinations. Uniformity and rms were lower, and value of contrast was higher for SPECT/CT images when CT was acquired with Al 2 mm and 3 mm filter than for images acquired without a filter. The study suggests that for Infinia Hawkeye 4, SPECT/CT system, Al 2 mm, and 3 mm are the optimum filters for improving image quality of SPECT/CT images of Jaszack or Philips CT QC phantom keeping other parameters of CT constant.
Collapse
Affiliation(s)
- Nivedita Rana
- Department of Nuclear Medicine and PET, PGIMER, Chandigarh, India
| | - Dinesh Rawat
- Department of Nuclear Medicine and PET, PGIMER, Chandigarh, India
| | - Madan Parmar
- Department of Nuclear Medicine and PET, PGIMER, Chandigarh, India
| | | | | | | |
Collapse
|
12
|
Hachouf N, Kharfi F, Hachouf M, Boucenna A. New analytical approach for neutron beam-hardening correction. Appl Radiat Isot 2015; 107:353-358. [PMID: 26609685 DOI: 10.1016/j.apradiso.2015.11.024] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/29/2015] [Revised: 11/08/2015] [Accepted: 11/12/2015] [Indexed: 11/30/2022]
Abstract
In neutron imaging, the beam-hardening effect has a significant effect on quantitative and qualitative image interpretation. This study aims to propose a linearization method for beam-hardening correction. The proposed method is based on a new analytical approach establishing the attenuation coefficient as a function of neutron energy. Spectrum energy shift due to beam hardening is studied on the basis of Monte Carlo N-Particle (MCNP) simulated data and the analytical data. Good agreement between MCNP and analytical values has been found. Indeed, the beam-hardening effect is well supported in the proposed method. A correction procedure is developed to correct the errors of beam-hardening effect in neutron transmission, and therefore for projection data correction. The effectiveness of this procedure is determined by its application in correcting reconstructed images.
Collapse
Affiliation(s)
- N Hachouf
- Nuclear Research Centre of Birine, BP 180, Ain Oussera, Djelfa, Algeria.
| | - F Kharfi
- Department of Physics, Faculty of Science, University of Setif, Algeria
| | - M Hachouf
- Nuclear Research Centre of Birine, BP 180, Ain Oussera, Djelfa, Algeria
| | - A Boucenna
- Department of Physics, Faculty of Science, University of Setif, Algeria
| |
Collapse
|