Lin WC, Wang HH, Hsu WL, Cao BH, Chen DJ, Tsai CJ. Investigation of an optimized scanning protocol for the dentomaxillofacial region using 320-slice multidetector computed tomography.
Dentomaxillofac Radiol 2017;
46:20160395. [PMID:
28128650 DOI:
10.1259/dmfr.20160395]
[Citation(s) in RCA: 4] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/05/2022] Open
Abstract
OBJECTIVES
To propose an imaging protocol that provides satisfactory image quality for oral examination while minimizing radiation dosage using 320-slice multidetector CT (MDCT).
METHODS
An anthropomorphic head phantom was scanned using 320 MDCT with protocols combining different scanning modes: volume scanning (whole or local) vs helical scanning (80- or 64-slice detectors); tube voltage settings (80 kVp, 120 kVp and 135 kVp); and tube current settings (60 mA, 80 mA, 100 mA and 120 mA). A total of six anatomical bone structures and three anatomical soft-tissue structures were assessed using quantitative and qualitative analysis in the three orthographic planes (axial, sagittal and coronal). A figure of merit (FOM) was used to determine the optimal imaging protocol in terms of tube voltage, tube current and scanning mode.
RESULTS
The 80-kVp setting had the worst quantitative and qualitative results (both p < 0.001) compared with the 135-kVp and 120-kVp settings, especially for soft-tissue structures. A significant difference was noted for the scores obtained using a tube current between 120 mA and 60 mA by quantitative analysis, but not by qualitative analysis. Volume scans using either whole or local modes had a significantly higher FOM than helical scanning of 80 or 64 slices.
CONCLUSIONS
In 320 MDCT, a protocol using 135 kVp, 80 mA and the volume-scanning mode (whole or local) offers adequate visualization of both soft-tissue and bone structures while keeping the radiation dose as low as possible. This may therefore be considered a first choice among a wide selection of scanning protocols for dentomaxillofacial CT.
Collapse