Malafosse C, Massiot N, Guimo F, Ben Abdallah I, Duprey A. Impact of the Endonaut® Angio-Navigation System on Radiation Exposure in Endovascular Aortic Repair Performed with Mobile C-Arms.
Ann Vasc Surg 2024;
109:143-148. [PMID:
38986839 DOI:
10.1016/j.avsg.2024.04.013]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/24/2023] [Revised: 04/14/2024] [Accepted: 04/29/2024] [Indexed: 07/12/2024]
Abstract
BACKGROUND
Fusion imaging systems have proved to reduce radiation exposure mostly in hybrid rooms but reports with mobile C-arms are few. The aim of this study was to analyze the impact of the Endonaut navigation system on radiation exposure in endovascular aneurysm repair (EVAR) performed with mobile C-arms.
METHODS
All patients undergoing EVAR and/or iliac branched devices implantation between January 2016 and August 2022 were included. All procedures were performed with a mobile C-arm (Siemens Avantic or GE Elite until March 2018, Siemens Cios Alpha thereafter). The Endonaut navigation system has been used since January 2021. Two groups were, therefore, compared: before (control group [CGr]) and after the use of Endonaut. Radiation data including Dose Area Product (DAP) values, Air Kerma (AK) and fluoroscopy time (FT) were collected retrospectively.
RESULTS
Overall, 153 patients were included: CGr, n = 121; Endonaut group (EnGr), n = 32. No significant difference was found between the 2 groups regarding demographic data. DAP values were significantly lower in the EnGr (38 Gy cm2 ± 24) vs. the CGr (76 Gy cm2 ± 51) (P < 0.05) despite a significantly higher number of complex procedures such as iliac branched devices (P < 0.05). AK values were not significantly different between the EnGr and the CGr (196 mGy ±114 vs. 209 mGy ±138) as well as FT (33 minutes ±18 vs. 33 minutes ±16). Technical success was 97% (31/32) in the EnGr vs. 96% (116/121) in the CGr (P = 0.79). The volume of contrast media was significantly lower in the EnGr (94 cc ± 41) vs. the CGr (143 cc ± 66) (P < 0.05).
CONCLUSIONS
In this study, the use of the Endonaut angio-navigation system when performing EVAR with mobile C-arms led to a radiation dose reduction without compromising technical success or procedural time.