Krbec J, Háček P, Berta M, Seidl J, Hron M, Pánek R. Fast density reconstruction of Li-BES signal on the COMPASS tokamak.
THE REVIEW OF SCIENTIFIC INSTRUMENTS 2018;
89:113504. [PMID:
30501310 DOI:
10.1063/1.5049894]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Received: 07/26/2018] [Accepted: 10/23/2018] [Indexed: 06/09/2023]
Abstract
This article describes a fast and automatic reconstruction of the edge plasma electron density from the radiation of energetic Li atoms of the diagnostic beam on the COMPASS tokamak. Radiation is detected by using a CCD camera and by using an avalanche photo-diode system with a temporal resolution of 20 ms and 2 μs, respectively. Both systems are equipped with a 670.8 nm optical filter which corresponds to the lithium 1s22s1-1s22p1 transition. A theoretical model and a data processing procedure of a raw signal to obtain the density profile are described. The reconstruction algorithm provides the absolutely calibrated electron density profiles together with the measurement error estimated from relatively calibrated light profiles; the implementation is performed in Python. Time demanding operations of the code were optimized to provide reconstruction of a single profile within less than 10 ms which makes the code applicable for processing of a large amount of data. Thanks to this calculation speed, it is possible to reconstruct electron density profiles between two consecutive shots on the COMPASS tokamak with 2 μs time resolution.