Evaluation of the Panbio™ COVID-19 IgG rapid test device performance.
Heliyon 2023;
9:e22612. [PMID:
38125420 PMCID:
PMC10730567 DOI:
10.1016/j.heliyon.2023.e22612]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/30/2023] [Revised: 11/13/2023] [Accepted: 11/15/2023] [Indexed: 12/23/2023] Open
Abstract
Background
The Panbio™ COVID-19 IgG Rapid Test Device ("Panbio™") detects IgG antibodies against the SARS-CoV-2 spike protein from viral infection or vaccination.
Objectives
To determine the diagnostic sensitivity and specificity of the Panbio™ professional use test, using fingerstick whole blood and venous plasma.
Study design
Fingerstick whole blood and venous plasma from each participant were tested with Panbio™ and compared against the SARS-CoV-2 IgG II assay on the Abbott Architect™ platform (Europe) or the equivalent AdviseDx SARS-CoV-2 IgG II Abbott Alinity i™ platform (US). 447 evaluable participants were enrolled across 6 US and 9 European clinical centers.
Results
For unvaccinated participants with PCR-confirmed infection ≥21 days post-symptom onset, the Panbio™ sensitivity with fingerstick whole blood was 92.6 % (95 % CI: 85.9, 96.7), and the specificity was 97.0 % (95 % CI: 93.1, 99.0). For venous plasma, the sensitivity was 90.0 % (95 % CI: 79.5, 96.2) for participants with PCR-confirmed infection and symptom onset 22-180 days ago; the specificity was 96.3 % (92.2, 98.6). For vaccinated participants, the sensitivity was 98.4 % (95 % CI: 91.2, 100.0) for fingerstick whole blood and 96.7 % (95 % CI: 88.7, 99.6) for venous plasma.
Conclusion
The Panbio™ test had high sensitivity and specificity for detecting IgG against the SARS-CoV-2 spike protein.
Collapse