Guo HL, Wu WW, Huan Y, Zhang HW. SGRT-based stereotactic body radiotherapy for lung cancer setup accuracy and margin of the PTV.
J Appl Clin Med Phys 2024;
25:e14195. [PMID:
37915300 DOI:
10.1002/acm2.14195]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/13/2023] [Revised: 09/09/2023] [Accepted: 10/18/2023] [Indexed: 11/03/2023] Open
Abstract
OBJECTIVE
Surface-guided radiation therapy (SGRT, AlignRT) was used to analyze motion during stereotactic body radiotherapy (SBRT) in lung cancer patients and to explore the margin of the planning target volume (PTV).
METHODS
The residual errors of the AlignRT were evaluated based on grayscale cone-beam computed tomography registration results before each treatment. AlignRT log file was used to analyze the correlation between the frequency and longest duration of errors larger than 2 mm and lasting longer than 2 s and maximum error with age and treatment duration. The displacement value at the end of treatment, the average displacement value, and the 95% probability density displacement interval were defined as intrafraction errors, and PTV1, PTV2, PTV3 were calculated by Van Herk formula or Z score analysis. Organ dosimetric differences were compared after the experience-based margin was replaced with PTV3.
RESULTS
The interfraction residual errors were Vrt0 , 0.06 ± 0.18 cm; Lng0 , -0.03 ± 0.19 cm; Lat0 , 0.02 ± 0.15 cm; Pitch0 , 0.23 ± 0.7°; Roll0 , 0.1 ± 0.69°; Rtn0 , -0.02 ± 0.79°. The frequency, longest duration and maximum error in vertical direction were correlated with treatment duration (r = 0.404, 0.353, 0.283, p < 0.05, respectively). In the longitudinal direction, the frequency was correlated with age and treatment duration (r = 0.376, 0.283, p < 0.05, respectively), maximum error was correlated with age (r = 0.4, P < 0.05). Vertical, longitudinal, lateral margins of PTV1, PTV2, PTV3 were 2 mm, 4 mm, 2 mm; 2 mm, 2 mm, 2 mm, 3 mm, 5 mm, 3 mm, respectively. After replacing the original PTV, mean lung dose (MLD), 2-cm3 chest wall dose (CD), lung V20 decreased by 0.2 Gy, 2.1 Gy, 0.5%, respectively (p < 0.05).
CONCLUSION
AlignRT can be used for interfraction setup and monitoring intrafraction motion. It is more reasonable to use upper and lower limits of the 95% probability density interval as an intrafraction error.
Collapse