Effect of Native Aortic Leaflet Geometry Modification on Transcatheter Aortic Valve Neo-sinus and Aortic Sinus Flow: An In-vitro Study.
Cardiovasc Eng Technol 2022;
14:182-193. [PMID:
36357649 DOI:
10.1007/s13239-022-00647-x]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 07/22/2022] [Accepted: 10/28/2022] [Indexed: 11/12/2022]
Abstract
PURPOSE
Leaflet thrombosis is a potentially fatal complication after transcatheter aortic valve replacement (TAVR). Blood flow stagnation in the neo-sinus and aortic sinuses are associated with increased thrombus severity. Native aortic leaflet modification may be a potential strategy to improve the neo-sinus and aortic sinus fluid dynamics. However, limited data exist on the effect of leaflet geometry modification on the flow within the neo-sinus and aortic sinus regions. We evaluate the effect of aortic leaflet modification on the neo-sinus and aortic sinus flow stagnation after simulated TAVR.
METHODS
Particle image velocimetry measurements were performed under nominal (5 LPM) and low (2.5 LPM) cardiac output conditions for an intact leaflet (control) case, and 3 modified leaflet geometries. Aortic leaflet geometry modification via leaflet splay was simulated with increasing splay geometry (leaflet splay distance: 5 mm-narrow, 10 mm-medium, and 20 mm-wide).
RESULTS
Leaflet geometry modification influenced flow features throughout the cardiac cycle, at both cardiac outputs, and allowed for flow communication between the neo-sinus and aortic sinus regions compared to the control. In the aortic sinus, flow stagnation reduced by over 64% at 5LPM, and over 36% at 2.5LPM for all simulated modified leaflet geometries compared to the control. However, only the medium and wide splay geometries enabled a reduction in neo-sinus flow stagnation compared to the control case.
CONCLUSIONS
These findings suggest that aortic leaflet geometry modification (of at least 10 mm leaflet splay distance) may reduce flow stasis and potentially decrease valve thrombosis risk.