Won DIL. Measurements of Endpoint Titers Based on the Fluorescence Intensity Trend in Anti-Nuclear Antibody Testing.
Lab Med 2021;
51:469-477. [PMID:
31872218 DOI:
10.1093/labmed/lmz087]
[Citation(s) in RCA: 2] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022] Open
Abstract
BACKGROUND
Automated systems for antinuclear antibody (ANA) testing provide endpoint titers that are predicted based on the fluorescence intensity (FI) value at a screening dilution (single-well titration [SWT]) showing frequent titration errors (more than plus or minus 1 dilution).
METHODS
Line slope titration (LST) was based on the trend of FI values on dilutions. Three dilutions per specimen were prepared considering a patient's previous titer or FI at the screening dilution. On the XY plot, with the reciprocal of dilution as the X-axis and FI value as the Y-axis, a fitted line was drawn to obtain the endpoint titers.
RESULTS
The titration error rate (no. of errors/total no.) of LST using a regression line was lower than that of SWT (31/710 [4.4%] and 152/674 [22.6%], respectively; P < .000000001), with serial dilution as a reference. When comparing a regression line using 3 dilution points with a line using 2 dilution points, the error rate of the former was not significantly different from that of the latter (31/710 [4.4%] and 31/746 [4.2%], respectively; P = .842).
CONCLUSIONS
This LST method is useful as an accurate, cost-effective, and rapid approach to measure endpoint titers in routine ANA testing.