Cho MR, Jun CM, Kim KT, Song SK, Choi WK. Results of primary THA using 36 mm femoral heads on first-generation highly cross-linked polyethylene in patients less than 60 years of age: Minimum 10-year follow-ups.
J Orthop Surg (Hong Kong) 2020;
28:2309499019896448. [PMID:
31903855 DOI:
10.1177/2309499019896448]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 11/17/2022] Open
Abstract
PURPOSE
Long-term results of total hip arthroplasty (THA) using highly cross-linked polyethylene (HXLPE) and metal femoral head with more than 10 years of follow-up have already been reported. However, most studies included results with a head size of 28 mm that could affect wear rates. The aim of this study was to evaluate the results of 36-mm metallic femoral heads on first-generation HXLPE in patients less than 60 years of age with a minimum follow-up of 10 years.
METHODS
Retrospective analysis included 54 cases from 47 patients. The mean age at the time of surgery was 47.22 years and the mean follow-up period was 131.04 months. Porous-coated cementless acetabular cups (Trilolgy®; Zimmer Inc., Warsaw, Indiana, USA) and HXLPE acetabular liners (Longevity®; Zimmer Inc.) were used for all cases. Acetabular cup abduction angles, anteversion angles, and wear rates of liner were measured using polyWare pro 3D distal version 5.10.
RESULTS
The average modified Harris hip score at the final follow-up was 88.48 (range 80-96). Average Merle d'Aubigne and Postel score was 15.57 (range 14-18). There was no acetabular cup or femoral stem failing due to aseptic loosening. The average steady-state wear rate determined using radiographs taken at 1 year postoperatively and at the latest follow-up was 0.053 ± 0.025 mm/year. There were no statistically significant differences in liner wear rate with respect to age, variety of the femoral stem, or liner thickness.
CONCLUSION
Results of THA with 36-mm metallic femoral heads on first-generation HXLPE in patients less than 60 years of age were satisfactory.
Collapse