Nagels LJ, Mush G, Massart DL. Rapid-scan hydrodynamic voltammetry and cyclic voltammetry of pharmaceuticals in flow injection analysis conditions.
J Pharm Biomed Anal 1989;
7:1479-83. [PMID:
2490532 DOI:
10.1016/0731-7085(89)80154-2]
[Citation(s) in RCA: 6] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/01/2023]
Abstract
An on-line rapid-scan electrochemical detector is described for HPLC and FIA systems. Its practical use in qualitative analysis is demonstrated for 19 drug substances. The detector can be operated to record convection/diffusion-controlled (S-shaped), or diffusion-controlled (peak-shaped) voltammograms. In the latter mode, on-line cyclic voltammetry measurements are possible. The cell can also be used as an amperometric detector for conventional, microbore and micro-LC methods. Detection limits are of the order of 10 pg (conventional and microbore HPLC), or at the sub-picogram level (micro-LC). For scanning work, drugs can be analysed at mg l-1 levels in an FIA setup.
Collapse