Zhang H, Wu T, Wu Q, Chen W, Ye C, Wang M, He X. Measurement of CO
2 Isotopologue Ratios Using a Hollow Waveguide-Based Mid-Infrared Dispersion Spectrometer.
Anal Chem 2023;
95:18479-18486. [PMID:
38054623 DOI:
10.1021/acs.analchem.3c03722]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/07/2023]
Abstract
We demonstrate for the first time the measurement of CO2 isotope ratios (13C/12C and 18O/16O) in a hollow waveguide (HWG) fiber using a mid-infrared heterodyne phase-sensitive dispersion spectrometer (HPSDS). A 4.329 μm interband cascade laser is used to target the absorption lines of three CO2 isotopes (13C16O2, 18O12C16O, and 12C16O2) in a 1 m long and 1 mm inner diameter HWG fiber. The detection limits are 0.29 ppm, 65.78 ppb, and 14.65 ppm with an integration time of 218 s for 13C16O2, 18O12C16O, and 12C16O2, respectively, at a modulation frequency of 160 MHz and a pressure of 230 mbar. The measurement precisions of δ13C and δ18O are 0.89 and 0.88 ‰, respectively, corresponding to an integration time of 167 s. An experimental comparison between a HPSDS and a built wavelength modulation system with second-harmonic detection (WMS-2f) is conducted. The results show that compared to the WMS-2f, the developed HPSDS exhibits a greater linear dynamic range and excellent long-term stability. This work aims to demonstrate a detection technique of CO2 isotope dispersion spectroscopy with a large dynamic range for relevant applications focusing on samples with high concentrations of CO2 (% volume fraction), such as respiratory analysis in medical diagnostics.
Collapse