Ross DH, Lee JY, Gao Y, Hollerbach AL, Bilbao A, Shi T, Ibrahim YM, Smith RD, Zheng X. Evaluation of a Reference-Free Collision Cross Section Calibration Strategy for Proteomics Using SLIM-Based High-Resolution Ion Mobility Spectrometry-Mass Spectrometry. JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY 2024; 35:1539-1549. [PMID: 38864778 DOI: 10.1021/jasms.4c00141] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/13/2024]
Abstract
Ion mobility spectrometry (IMS) is a gas-phase analytical technique that separates ions with different sizes and shapes and is compatible with mass spectrometry (MS) to provide an additional separation dimension. The rapid nature of the IMS separation combined with the high sensitivity of MS-based detection and the ability to derive structural information on analytes in the form of the property collision cross section (CCS) makes IMS particularly well-suited for characterizing complex samples in -omics applications. In such applications, the quality of CCS from IMS measurements is critical to confident annotation of the detected components in the complex -omics samples. However, most IMS instrumentation in mainstream use requires calibration to calculate CCS from measured arrival times, with the most notable exception being drift tube IMS measurements using multifield methods. The strategy for calibrating CCS values, particularly selection of appropriate calibrants, has important implications for CCS accuracy, reproducibility, and transferability between laboratories. The conventional approach to CCS calibration involves explicitly defining calibrants ahead of data acquisition and crucially relies upon availability of reference CCS values. In this work, we present a novel reference-free approach to CCS calibration which leverages trends among putatively identified features and computational CCS prediction to conduct calibrations post-data acquisition and without relying on explicitly defined calibrants. We demonstrated the utility of this reference-free CCS calibration strategy for proteomics application using high-resolution structures for lossless ion manipulations (SLIM)-based IMS-MS. We first validated the accuracy of CCS values using a set of synthetic peptides and then demonstrated using a complex peptide sample from cell lysate.
Collapse
Affiliation(s)
- Dylan H Ross
- Earth and Biological Sciences Directorate, Pacific Northwest National Laboratory, Richland, Washington 99352, United States
| | - Jung Yun Lee
- Earth and Biological Sciences Directorate, Pacific Northwest National Laboratory, Richland, Washington 99352, United States
| | - Yuqian Gao
- Earth and Biological Sciences Directorate, Pacific Northwest National Laboratory, Richland, Washington 99352, United States
| | - Adam L Hollerbach
- Earth and Biological Sciences Directorate, Pacific Northwest National Laboratory, Richland, Washington 99352, United States
| | - Aivett Bilbao
- Earth and Biological Sciences Directorate, Pacific Northwest National Laboratory, Richland, Washington 99352, United States
| | - Tujin Shi
- Earth and Biological Sciences Directorate, Pacific Northwest National Laboratory, Richland, Washington 99352, United States
| | - Yehia M Ibrahim
- Earth and Biological Sciences Directorate, Pacific Northwest National Laboratory, Richland, Washington 99352, United States
| | - Richard D Smith
- Earth and Biological Sciences Directorate, Pacific Northwest National Laboratory, Richland, Washington 99352, United States
| | - Xueyun Zheng
- Earth and Biological Sciences Directorate, Pacific Northwest National Laboratory, Richland, Washington 99352, United States
| |
Collapse
|
2
|
Huntley AP, Hollerbach AL, Norheim RV, Hamid AM, Anderson GA, Garimella SV, Ibrahim YM. Cyclable Variable Path Length Multilevel Structures for Lossless Ion Manipulations (SLIM) Platform for Enhanced Ion Mobility Separations. Anal Chem 2024:10.1021/acs.analchem.3c05594. [PMID: 38336463 PMCID: PMC11310366 DOI: 10.1021/acs.analchem.3c05594] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 02/12/2024]
Abstract
Ion mobility-mass spectrometry (IMS-MS) is used to analyze complex samples and provide structural information on unknown compounds. As the complexity of samples increases, there is a need to improve the resolution of IMS-MS instruments to increase the rate of molecular identification. This work evaluated a cyclable and variable path length (and hence resolving power) multilevel Structures for Lossless Ion Manipulations (SLIM) platform to achieve a higher resolving power than what was previously possible. This new multilevel SLIM platform has eight separation levels connected by ion escalators, yielding a total path length of ∼88 m (∼11 m per level). Our new multilevel SLIM can also be operated in an "ion cycling" mode by utilizing a set of return ion escalators that transport ions from the eighth level back to the first, allowing even extendable path lengths (and higher IMS resolution). The platform has been improved to enhance ion transmission and IMS separation quality by reducing the spacing between SLIM boards. The board thickness was reduced to minimize the ions' escalator residence time. Compared to the previous generation, the new multilevel SLIM demonstrated better transmission for a set of phosphazene ions, especially for the low-mobility ions. For example, the transmission of m/z 2834 ions was improved by a factor of ∼3 in the new multilevel SLIM. The new multilevel SLIM achieved 49% better resolving powers for GRGDS1+ ions in 4 levels than our previous 4-level SLIM. The collision cross-section-based resolving power of the SLIM platform was tested using a pair of reverse sequence peptides (SDGRG1+, GRGDS1+). We achieved 1100 resolving power using 88 m of path length (i.e., 8 levels) and 1400 following an additional pass through the eight levels. Further evaluation of the multilevel SLIM demonstrated enhanced separation for positively and negatively charged brain total lipid extract samples. The new multilevel SLIM enables a tunable high resolving power for a wide range of ion mobilities and improved transmission for low-mobility ions.
Collapse
Affiliation(s)
- Adam P. Huntley
- Biological Sciences Division, Pacific Northwest National Laboratory, P.O. Box 999, Richland, Washington, 99354, United States
| | - Adam L. Hollerbach
- Biological Sciences Division, Pacific Northwest National Laboratory, P.O. Box 999, Richland, Washington, 99354, United States
| | - Randolph V. Norheim
- Biological Sciences Division, Pacific Northwest National Laboratory, P.O. Box 999, Richland, Washington, 99354, United States
| | - Ahmed M. Hamid
- Biological Sciences Division, Pacific Northwest National Laboratory, P.O. Box 999, Richland, Washington, 99354, United States
| | - Gordon A. Anderson
- Gordon A. Anderson Custom Electronics (GAACE), Kennewick, Washington, 99338
| | - Sandilya V.B. Garimella
- Biological Sciences Division, Pacific Northwest National Laboratory, P.O. Box 999, Richland, Washington, 99354, United States
| | - Yehia M. Ibrahim
- Biological Sciences Division, Pacific Northwest National Laboratory, P.O. Box 999, Richland, Washington, 99354, United States
| |
Collapse
|