Spencer AF, Vernon SA. Repeatability and reproducibility of optic disc measurement with the Zeiss 4-mirror contact lens.
Ophthalmology 1996;
103:163-7. [PMID:
8628548 DOI:
10.1016/s0161-6420(96)30745-8]
[Citation(s) in RCA: 7] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 02/01/2023] Open
Abstract
BACKGROUND
A simple clinical quantitative method of optic disc diameter measurement using a 4-mirror gonioscope contact lens has been described recently.
METHODS
Intraobserver and interobserver variations are assessed on 77 eyes of 41 patients. Of these eyes, 63 had refractive errors within 3 diopters of emmetropia. The disc was measured by projecting a slit beam of known height onto the disc.
RESULTS
Intraobserver variation: the coefficient of variation was 3.04% for observer 1 (experienced) and 4.76% for observer 2 (novice). The mean of the three measurements by the two observers was not significantly different (P=0.2318). For observer 1, the mean variance of the eyes with higher refractive errors did not differ significantly from the eyes with lower refractive errors. Interobserver variation: the coefficient of variation was 4.52%. The measurements for the novice were slightly larger the smallest discs and slightly smaller for the largest discs (regression analysis P=0.0209).
CONCLUSION
Repeatability and reproducibility for this method are comparable with other methods of measurement. Until experienced with the technique, the authors recommend that the mean of three readings is used.