[The future of bedside chest radiography: Comparative study of mobile flat-panels and needle-image plate storage phosphor systems].
Med Klin Intensivmed Notfmed 2015;
110:603-8. [PMID:
25905891 DOI:
10.1007/s00063-015-0022-2]
[Citation(s) in RCA: 2] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/29/2014] [Revised: 01/31/2015] [Accepted: 02/22/2015] [Indexed: 10/23/2022]
Abstract
BACKGROUND
Chest radiography is the most common diagnostic modality in intensive care units with new mobile flat-panels gaining more attention and availability in addition to the already used storage phosphor plates.
PURPOSE
Comparison of the image quality of mobile flat-panels and needle-image plate storage phosphor system in terms of bedside chest radiography.
MATERIALS AND METHODS
Retrospective analysis of 84 bedside chest radiographs of 42 intensive care patients (20 women, 22 men, average age: 65 years). All images were acquired during daily routine. For each patient, two images were analyzed, one from each system mentioned above. Two blinded radiologists evaluated the image quality based on ten criteria (e.g., diaphragm, heart contour, tracheal bifurcation, thoracic spine, lung structure, consolidations, foreign material, and overall impression) using a 5-point visibility scale (1 = excellent, 5 = not usable).
RESULTS
There was no significant difference between the image quality of the two systems (p < 0.05). Overall some anatomical structures such as the diaphragm, heart, pulmonary consolidations and foreign material were considered of higher diagnostic quality compared to others, e.g., tracheal bifurcation and thoracic spine.
CONCLUSIONS
Mobile flat-panels achieve an image quality which is as good as those of needle-image plate storage phosphor systems. In addition, they allow immediate evaluation of the image quality but in return are much more expensive in terms of purchase and maintenance.