Akiyama T, Yasuhara R, Kawahata K, Okajima S, Nakayama K. Dispersion interferometer using modulation amplitudes on LHD (invited).
THE REVIEW OF SCIENTIFIC INSTRUMENTS 2014;
85:11D301. [PMID:
25430161 DOI:
10.1063/1.4886777]
[Citation(s) in RCA: 2] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/04/2023]
Abstract
Since a dispersion interferometer is insensitive to mechanical vibrations, a vibration compensation system is not necessary. The CO2 laser dispersion interferometer with phase modulations on the Large Helical Device utilizes the new phase extraction method which uses modulation amplitudes and can improve a disadvantage of the original dispersion interferometer: measurement errors caused by variations of detected intensities. The phase variation within ±2 × 10(17) m(-3) is obtained without vibration compensation system. The measured line averaged electron density with the dispersion interferometer shows good agreement with that with the existing far infrared laser interferometer. Fringe jump errors in high density ranging up to 1.5 × 10(20) m(-3) can be overcome by a sufficient sampling rate of about 100 kHz.
Collapse