The next generation of current measurement for ionization chambers.
Appl Radiat Isot 2020;
163:109216. [PMID:
32561054 DOI:
10.1016/j.apradiso.2020.109216]
[Citation(s) in RCA: 7] [Impact Index Per Article: 1.8] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/29/2019] [Revised: 03/23/2020] [Accepted: 05/03/2020] [Indexed: 11/18/2022]
Abstract
Re-entrant ionization chambers (ICs) are essential to radionuclide metrology and nuclear medicine for maintaining standards and measuring half-lives. The requirements of top-level metrology demand that systems must be precise and stable to 0.1 % over many years, and linear from 10-14 A to 10-8 A. Thus, laboratories depend on bespoke current measurement systems and often rely on sealed sources to generate reference currents. To maintain and improve present capabilities, metrologists need to overcome two looming challenges: ageing electronics and decreasing availability of sealed sources. Possible solutions using Ultrastable Low-Noise Current Amplifiers (ULCAs), resistive-feedback electrometers, and (quantum) single-electron pumps are reviewed. Broader discussions of IC design and methodology are discussed. ULCAs show promise and resistive-feedback systems which take advantage of standard resistor calibrations offer an alternative.
Collapse