Su XL, Yu BS, Tan HW, Yang XR, Nie LH, Yao SZ. Flow-injection determination of total ammonia and total carbon dioxide in blood based on gas-diffusion separation and with a bulk acoustic wave impedance sensor.
J Pharm Biomed Anal 1998;
16:759-69. [PMID:
9535187 DOI:
10.1016/s0731-7085(97)00108-8]
[Citation(s) in RCA: 13] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 02/07/2023]
Abstract
A novel flow-injection (FIA) system, for the rapid and direct determination of both total ammonia (T[NH3]) and total carbon dioxide (T[CO2]) in clinical blood samples, has been developed. Samples were injected into a carrier stream of H2O, then emerged with a reagent stream, where the analyte was converted into a gaseous species and diffused across a PTFE gas-permeable membrane into an acceptor stream. The trapped NH3/CO2 in the acceptor was determined on line by a bulk acoustic wave (BAW) impedance sensor. At a through-put of 20 and 65 h(-1), the proposed system exhibited a linear frequency response up to 200 micromol l(-1) ammonium and 20 mmol l(-1) bicarbonate with a detection limit of 1.0 and 10 micromol l(-1), respectively. Results obtained for T(NH3) in serum and T(CO2) in plasma were in agreement with those obtained by the conventional glutamate dehydrogenase (GDH) method and gas-sensing electrode method, respectively. The effects of composition of acceptor stream, cell constant of conductivity electrode, sample volume, flow rate and potential interferents on the FIA signals were also discussed.