1
|
Dos Santos DS, Ossenkoppele B, Hopf YM, Soozande M, Noothout E, Vos HJ, Bosch JG, Pertijs MAP, Verweij MD, de Jong N. An Ultrasound Matrix Transducer for High-Frame-Rate 3-D Intra-cardiac Echocardiography. ULTRASOUND IN MEDICINE & BIOLOGY 2024; 50:285-294. [PMID: 38036310 DOI: 10.1016/j.ultrasmedbio.2023.11.001] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 06/28/2023] [Revised: 10/31/2023] [Accepted: 11/01/2023] [Indexed: 12/02/2023]
Abstract
OBJECTIVE Described here is the development of an ultrasound matrix transducer prototype for high-frame-rate 3-D intra-cardiac echocardiography. METHODS The matrix array consists of 16 × 18 lead zirconate titanate elements with a pitch of 160 µm × 160 µm built on top of an application-specific integrated circuit that generates transmission signals and digitizes the received signals. To reduce the number of cables in the catheter to a feasible number, we implement subarray beamforming and digitization in receive and use a combination of time-division multiplexing and pulse amplitude modulation data transmission, achieving an 18-fold reduction. The proposed imaging scheme employs seven fan-shaped diverging transmit beams operating at a pulse repetition frequency of 7.7 kHz to obtain a high frame rate. The performance of the prototype is characterized, and its functionality is fully verified. RESULTS The transducer exhibits a transmit efficiency of 28 Pa/V at 5 cm per element and a bandwidth of 60% in transmission. In receive, a dynamic range of 80 dB is measured with a minimum detectable pressure of 10 Pa per element. The element yield of the prototype is 98%, indicating the efficacy of the manufacturing process. The transducer is capable of imaging at a frame rate of up to 1000 volumes/s and is intended to cover a volume of 70° × 70° × 10 cm. CONCLUSION These advanced imaging capabilities have the potential to support complex interventional procedures and enable full-volumetric flow, tissue, and electromechanical wave tracking in the heart.
Collapse
Affiliation(s)
- Djalma Simões Dos Santos
- Laboratory of Medical Imaging, Department of Imaging Physics, Delft University of Technology, Delft, The Netherlands.
| | - Boudewine Ossenkoppele
- Laboratory of Medical Imaging, Department of Imaging Physics, Delft University of Technology, Delft, The Netherlands
| | - Yannick M Hopf
- Electronic Instrumentation Laboratory, Delft University of Technology, Delft, The Netherlands
| | - Mehdi Soozande
- Department of Biomedical Engineering, Thoraxcenter, Erasmus Medical Center, Rotterdam, The Netherlands
| | - Emile Noothout
- Laboratory of Medical Imaging, Department of Imaging Physics, Delft University of Technology, Delft, The Netherlands
| | - Hendrik J Vos
- Laboratory of Medical Imaging, Department of Imaging Physics, Delft University of Technology, Delft, The Netherlands; Department of Biomedical Engineering, Thoraxcenter, Erasmus Medical Center, Rotterdam, The Netherlands
| | - Johan G Bosch
- Department of Biomedical Engineering, Thoraxcenter, Erasmus Medical Center, Rotterdam, The Netherlands
| | - Michiel A P Pertijs
- Electronic Instrumentation Laboratory, Delft University of Technology, Delft, The Netherlands
| | - Martin D Verweij
- Laboratory of Medical Imaging, Department of Imaging Physics, Delft University of Technology, Delft, The Netherlands; Department of Biomedical Engineering, Thoraxcenter, Erasmus Medical Center, Rotterdam, The Netherlands
| | - Nico de Jong
- Laboratory of Medical Imaging, Department of Imaging Physics, Delft University of Technology, Delft, The Netherlands; Department of Biomedical Engineering, Thoraxcenter, Erasmus Medical Center, Rotterdam, The Netherlands
| |
Collapse
|
2
|
Manuel TJ, Phipps MA, Caskey CF. Design of a 1-MHz Therapeutic Ultrasound Array for Small Volume Blood-Brain Barrier Opening at Cortical Targets in Macaques. IEEE TRANSACTIONS ON ULTRASONICS, FERROELECTRICS, AND FREQUENCY CONTROL 2023; 70:449-459. [PMID: 37028345 DOI: 10.1109/tuffc.2023.3256268] [Citation(s) in RCA: 2] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/16/2023]
Abstract
[[gabstract]][] Focused ultrasound (FUS) can temporarily open the blood-brain barrier (BBB) and increase the delivery of chemotherapeutics, viral vectors, and other agents to the brain parenchyma. To limit FUS BBB opening to a single brain region, the transcranial acoustic focus of the ultrasound transducer must not be larger than the region targeted. In this work, we design and characterize a therapeutic array optimized for BBB opening at the frontal eye field (FEF) in macaques. We used 115 transcranial simulations in four macaques varying f-number and frequency to optimize the design for focus size, transmission, and small device footprint. The design leverages inward steering for focus tightening, a 1-MHz transmit frequency, and can focus to a simulation predicted 2.5- ± 0.3-mm lateral and 9.5- ± 1.0-mm axial full-width at half-maximum spot size at the FEF without aberration correction. The array is capable of steering axially 35 mm outward, 26 mm inward, and laterally 13 mm with 50% the geometric focus pressure. The simulated design was fabricated, and we characterized the performance of the array using hydrophone beam maps in a water tank and through an ex vivo skull cap to compare measurements with simulation predictions, achieving a 1.8-mm lateral and 9.5-mm axial spot size with a transmission of 37% (transcranial, phase corrected). The transducer produced by this design process is optimized for BBB opening at the FEF in macaques.
Collapse
|
3
|
dos Santos DS, Fool F, Mozaffarzadeh M, Shabanimotlagh M, Noothout E, Kim T, Rozsa N, Vos HJ, Bosch JG, Pertijs MAP, Verweij MD, de Jong N. A Tiled Ultrasound Matrix Transducer for Volumetric Imaging of the Carotid Artery. SENSORS (BASEL, SWITZERLAND) 2022; 22:9799. [PMID: 36560168 PMCID: PMC9784751 DOI: 10.3390/s22249799] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Figures] [Subscribe] [Scholar Register] [Received: 11/14/2022] [Revised: 12/09/2022] [Accepted: 12/12/2022] [Indexed: 06/17/2023]
Abstract
High frame rate three-dimensional (3D) ultrasound imaging would offer excellent possibilities for the accurate assessment of carotid artery diseases. This calls for a matrix transducer with a large aperture and a vast number of elements. Such a matrix transducer should be interfaced with an application-specific integrated circuit (ASIC) for channel reduction. However, the fabrication of such a transducer integrated with one very large ASIC is very challenging and expensive. In this study, we develop a prototype matrix transducer mounted on top of multiple identical ASICs in a tiled configuration. The matrix was designed to have 7680 piezoelectric elements with a pitch of 300 μm × 150 μm integrated with an array of 8 × 1 tiled ASICs. The performance of the prototype is characterized by a series of measurements. The transducer exhibits a uniform behavior with the majority of the elements working within the -6 dB sensitivity range. In transmit, the individual elements show a center frequency of 7.5 MHz, a -6 dB bandwidth of 45%, and a transmit efficiency of 30 Pa/V at 200 mm. In receive, the dynamic range is 81 dB, and the minimum detectable pressure is 60 Pa per element. To demonstrate the imaging capabilities, we acquired 3D images using a commercial wire phantom.
Collapse
Affiliation(s)
- Djalma Simões dos Santos
- Laboratory of Medical Imaging, Department of Imaging Physics, Delft University of Technology, 2628 CJ Delft, The Netherlands
| | - Fabian Fool
- Laboratory of Medical Imaging, Department of Imaging Physics, Delft University of Technology, 2628 CJ Delft, The Netherlands
| | - Moein Mozaffarzadeh
- Laboratory of Medical Imaging, Department of Imaging Physics, Delft University of Technology, 2628 CJ Delft, The Netherlands
| | - Maysam Shabanimotlagh
- Laboratory of Medical Imaging, Department of Imaging Physics, Delft University of Technology, 2628 CJ Delft, The Netherlands
| | - Emile Noothout
- Laboratory of Medical Imaging, Department of Imaging Physics, Delft University of Technology, 2628 CJ Delft, The Netherlands
| | - Taehoon Kim
- Electronic Instrumentation Laboratory, Delft University of Technology, 2628 CD Delft, The Netherlands
| | - Nuriel Rozsa
- Electronic Instrumentation Laboratory, Delft University of Technology, 2628 CD Delft, The Netherlands
| | - Hendrik J. Vos
- Laboratory of Medical Imaging, Department of Imaging Physics, Delft University of Technology, 2628 CJ Delft, The Netherlands
- Department Biomedical Engineering, Thoraxcenter, Erasmus Medical Center, 3015 GD Rotterdam, The Netherlands
| | - Johan G. Bosch
- Department Biomedical Engineering, Thoraxcenter, Erasmus Medical Center, 3015 GD Rotterdam, The Netherlands
| | - Michiel A. P. Pertijs
- Electronic Instrumentation Laboratory, Delft University of Technology, 2628 CD Delft, The Netherlands
| | - Martin D. Verweij
- Laboratory of Medical Imaging, Department of Imaging Physics, Delft University of Technology, 2628 CJ Delft, The Netherlands
- Department Biomedical Engineering, Thoraxcenter, Erasmus Medical Center, 3015 GD Rotterdam, The Netherlands
| | - Nico de Jong
- Laboratory of Medical Imaging, Department of Imaging Physics, Delft University of Technology, 2628 CJ Delft, The Netherlands
- Department Biomedical Engineering, Thoraxcenter, Erasmus Medical Center, 3015 GD Rotterdam, The Netherlands
| |
Collapse
|
4
|
Ramalli A, Boni E, Roux E, Liebgott H, Tortoli P. Design, Implementation, and Medical Applications of 2-D Ultrasound Sparse Arrays. IEEE TRANSACTIONS ON ULTRASONICS, FERROELECTRICS, AND FREQUENCY CONTROL 2022; 69:2739-2755. [PMID: 35333714 DOI: 10.1109/tuffc.2022.3162419] [Citation(s) in RCA: 10] [Impact Index Per Article: 5.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/14/2023]
Abstract
An ultrasound sparse array consists of a sparse distribution of elements over a 2-D aperture. Such an array is typically characterized by a limited number of elements, which in most cases is compatible with the channel number of the available scanners. Sparse arrays represent an attractive alternative to full 2-D arrays that may require the control of thousands of elements through expensive application-specific integrated circuits (ASICs). However, their massive use is hindered by two main drawbacks: the possible beam profile deterioration, which may worsen the image contrast, and the limited signal-to-noise ratio (SNR), which may result too low for some applications. This article reviews the work done for three decades on 2-D ultrasound sparse arrays for medical applications. First, random, optimized, and deterministic design methods are reviewed together with their main influencing factors. Then, experimental 2-D sparse array implementations based on piezoelectric and capacitive micromachined ultrasonic transducer (CMUT) technologies are presented. Sample applications to 3-D (Doppler) imaging, super-resolution imaging, photo-acoustic imaging, and therapy are reported. The final sections discuss the main shortcomings associated with the use of sparse arrays, the related countermeasures, and the next steps envisaged in the development of innovative arrays.
Collapse
|
5
|
Zhang Q, Mao J, Zhang Y, Lu M, Li R, Liu X, Liu Y, Yang R, Wang X, Geng Y, Qi T, Wan M. Multiple-Focus Patterns of Sparse Random Array Using Particle Swarm Optimization for Ultrasound Surgery. IEEE TRANSACTIONS ON ULTRASONICS, FERROELECTRICS, AND FREQUENCY CONTROL 2022; 69:565-579. [PMID: 34757903 DOI: 10.1109/tuffc.2021.3127222] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/13/2023]
Abstract
This study aims to investigate the feasibility and potential of sparse random arrays driven by the particle swarm optimization (PSO) algorithm to generate multiple-focus patterns and a large scanning range without grating lobes, which extends the scanning range of focused ultrasound in the treatment of brain tumors, opening the blood-brain barrier, and neuromodulation. Operating at 1.1 MHz, a random spherical array with 200 square elements (sparseness 58%) and a sparse random array with 660 square elements (sparseness 41%) driven by PSO are employed to simulate different focus patterns. With the same radius of curvature and diameter of transducer and element size, the scanning range of the off-axis single focus of a random 200-element array is two times that of an ordinary array using symmetric arrangement. The focal volume of multiple-focus patterns of the random array is 18 times that of the single focus. The single focus of the sparse random array with 660 elements could steer up to ±23 mm in the radial direction, without grating lobes. The maximum distance between two foci in a multiple-focus "S"-shaped deflection is approximately 25 mm. Simulation results illustrate the capability of a focused beam steered in 3-D space. Multiple-focus patterns could significantly increase the focal volume and shorten the treatment time for large target volumes. Simulation results show the feasibility and potential of the method combining PSO with a sparse random array to generate flexible focus patterns that can adapt to different needs in different tissue treatments.
Collapse
|
6
|
Zubair M, Dickinson RJ. 3D synthetic aperture imaging with a therapeutic spherical random phased array for transcostal applications. Phys Med Biol 2021; 66:035024. [PMID: 33276351 DOI: 10.1088/1361-6560/abd0d0] [Citation(s) in RCA: 6] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/18/2022]
Abstract
Experimental validation of a synthetic aperture imaging technique using a therapeutic random phased array is described, demonstrating the dual nature of imaging and therapy of such an array. The transducer is capable of generating both continuous wave high intensity beams for ablating the tumor and low intensity ultrasound pulses to image the target area. Pulse-echo data is collected from the elements of the phased array to obtain B-mode images of the targets. Since therapeutic arrays are optimized for therapy only with concave apertures having low f-number and large directive elements often coarsely sampled, imaging can not be performed using conventional beamforming. We show that synthetic aperture imaging is capable of processing the acquired RF data to obtain images of the field of interest. Simulations were performed to compare different synthetic aperture imaging techniques to identify the best algorithm in terms of spatial resolution. Experimental validation was performed using a 1 MHz, 256-elements, spherical random phased array with 130 mm radius of curvature. The array was integrated with a research ultrasound scanner via custom connectors to acquire raw RF data for variety of targets. Imaging was implemented using synthetic aperture beamforming to produce images of a rib phantom and ex vivo ribs. The array was shown to resolve spherical targets within ±15 mm of either side of the axis in the focal plane and obtain 3D images of the rib phantom up to ±40 mm of either side of the central axis and at a depth of 3-9 cm from the array surface. The lateral and axial full width half maximum was 1.15 mm and 2.75 mm, respectively. This study was undertaken to emphasize that both therapy and image guidance with a therapeutic random phased array is possible and such a system has the potential to address some major limitations in the existing high intensity focused ultrasound (HIFU) systems. The 3D images obtained with a therapeutic array can be used to identify and locate strong scattering objects aiding to image guidance and treatment planning of the HIFU procedure.
Collapse
Affiliation(s)
- Muhammad Zubair
- Department of Bioengineering, Imperial College London, United Kingdom
| | | |
Collapse
|
7
|
Qu X, Shen G, Wu N, Wu H, Qiao S, Wang E, Chen Y, Wang H. Suppressing Grating Lobes for Transcranial Focused Ultrasound System by Frequency-Modulated Excitation. IEEE TRANSACTIONS ON ULTRASONICS, FERROELECTRICS, AND FREQUENCY CONTROL 2021; 68:341-351. [PMID: 33382650 DOI: 10.1109/tuffc.2020.3047664] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/12/2023]
Abstract
Transcranial focused ultrasound is a novel noninvasive therapeutic modality for glioblastoma and other disorders of the brain. However, because the phase aberrations caused by the skull need to be corrected with computed tomography (CT) images, the transcranial transducer is tightly fixed on the patient's head to avoid any variation in the relative position, and the focus shifting relies mainly on the capacity for electronic beam steering. Due to the presence of grating lobes and the rapid degradation of the focus quality with increasing focus-shifting distance, transcranial focus-shifting sonication may damage healthy brain tissue unintentionally. To reduce the risks associated with transcranial focused ultrasound therapy, linear frequency-modulated (FM) excitation is proposed. The k-space corrected pseudospectral time domain (PSTD) and acoustic holography approach based on the Rayleigh integral are combined to calculate the distribution of the deposited acoustic power. The corresponding simulation was performed with axial/lateral focus shifting at different distances. The distributions of the deposited acoustic power show that linear FM excitation can effectively suppress undesired prefocal grating lobes without compromising focus quality.
Collapse
|
8
|
Yang Y, Wang C, Li Y, Huang J, Cai F, Xiao Y, Ma T, Zheng H. Development of Scalable 2D Plane Array for Transcranial Ultrasonic Neuromodulation on Non-Human Primates: An Ex Vivo Study. IEEE Trans Neural Syst Rehabil Eng 2019; 28:361-369. [PMID: 31841414 DOI: 10.1109/tnsre.2019.2959436] [Citation(s) in RCA: 5] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/06/2022]
Abstract
Ultrasonic neuromodulation on large animals, like non-human primates, requires the array transducer to have a good steering ability to arbitrarily stimulate various brain locations. Moreover, due to the different sizes of the animal heads, the array is preferred to be conveniently adjustable to different aperture sizes. To meet these requirements, a scalable 2D plane array, which can consist of up to tens of 256-element square modules, has been designed and fabricated in this study. Arrays with large apertures, consisting of 12 and 48 modules (including 3072 and 12288 elements, respectively), have been assembled and tested. The square modules were driven at a center frequency of 1.04 MHz. The transcranial and steering abilities of the modules were verified in a water tank on an ex vivo macaque skull. A focused beam transmitted by single module could pass through the skull with its peak acoustic pressure reduced to 18.1% of that in the free-field, with an acceptable transverse steering range up to 10 mm. Moreover, a much smaller focal spot was obtained by simultaneous transmission using two vertical-placed modules. Compared to the single module, the focal zone axial size was reduced from 29.7 mm to 3.3 mm, and the "stripes" pattern in the focal zone caused by standing waves was eliminated using a difference-frequency transmitting strategy. Such scalable 2D plane array, which can provide a small-size focal spot and has a large steering range and an acceptable transcranial ability, can be useful in research on ultrasonic neuromodulation on non-human primates.
Collapse
|
9
|
Xu R, O'Reilly MA. Simulating transvertebral ultrasound propagation with a multi-layered ray acoustics model. Phys Med Biol 2018; 63:145017. [PMID: 29947618 DOI: 10.1088/1361-6560/aacf75] [Citation(s) in RCA: 11] [Impact Index Per Article: 1.8] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/12/2022]
Abstract
The simulation accuracy of transvertebral ultrasound propagation using a multi-layered ray acoustics model based on CT-derived vertebral geometry was investigated through comparison with experimental measurements of pressure fields in ex vivo human vertebral foramen. A spherically focused transducer (5 cm diameter, f-number 1.2, 514 kHz) was geometrically focused to the centre of individual thoracic vertebral foramen, through the posterior bony elements. Transducer propagation paths through the laminae and the spinous processes were tested. Simulation transducer-vertebra configurations were registered to experiment transducer-vertebra configurations, and simulation accuracy of the simulation model was evaluated for predicting maximum transmitted pressure to the canal, voxel pressure in the canal, and focal distortion. Accuracy in predicting maximum transmitted pressure was calculated by vertebra, and it is shown that simulation predicts maximum pressure with a greater degree of accuracy than a vertebra-specific insertion loss. Simulation error in voxel pressure was evaluated using root-mean-square error and cross-correlation, and found to be similar to the water-only case. Simulation accuracy in predicting focal distortion was evaluated by comparing experiment and simulation maximum pressure location and weighted >50% focal volume location. Average simulation error across all measurements and simulations in maximum pressure location and weighted >50% focal volume location were 2.3 mm and 1.5 mm, respectively. These errors are small relative to the dimensions of the transducer focus (4.9 mm full width half maximum), the spinal cord (10 mm diameter), and vertebral canal diameter (15-20 mm diameter). These results suggest that ray acoustics can be applied to simulating transvertebral ultrasound propagation.
Collapse
Affiliation(s)
- Rui Xu
- Sunnybrook Research Institute, Sunnybrook Health Sciences Centre, 2075 Bayview Avenue, Toronto, ON, M4N 3M5, Canada. Department of Medical Biophysics, Faculty of Medicine, University of Toronto, 101 College Street Suite 15-701, Toronto, ON M5G 1L7, Canada. Author to whom any correspondence should be addressed
| | | |
Collapse
|
10
|
Chaplin V, Phipps MA, Caskey CF. A random phased-array for MR-guided transcranial ultrasound neuromodulation in non-human primates. Phys Med Biol 2018; 63:105016. [PMID: 29667598 PMCID: PMC6941739 DOI: 10.1088/1361-6560/aabeff] [Citation(s) in RCA: 29] [Impact Index Per Article: 4.8] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
Transcranial focused ultrasound (FUS) is a non-invasive technique for therapy and study of brain neural activation. Here we report on the design and characterization of a new MR-guided FUS transducer for neuromodulation in non-human primates at 650 kHz. The array is randomized with 128 elements 6.6 mm in diameter, radius of curvature 7.2 cm, opening diameter 10.3 cm (focal ratio 0.7), and 46% coverage. Simulations were used to optimize transducer geometry with respect to focus size, grating lobes, and directivity. Focus size and grating lobes during electronic steering were quantified using hydrophone measurements in water and a three-axis stage. A novel combination of optical tracking and acoustic mapping enabled measurement of the 3D pressure distribution in the cortical region of an ex vivo skull to within ~3.5 mm of the surface, and allowed accurate modelling of the experiment via non-homogeneous 3D acoustic simulations. The data demonstrates acoustic focusing beyond the skull bone, with the focus slightly broadened and shifted proximal to the skull. The fabricated design is capable of targeting regions within the S1 sensorimotor cortex of macaques.
Collapse
Affiliation(s)
- Vandiver Chaplin
- Vanderbilt University Institute of Imaging Science, 1161 21st Avenue South, Nashville, TN 37232
| | - Marshal A. Phipps
- Vanderbilt University Institute of Imaging Science, 1161 21st Avenue South, Nashville, TN 37232
| | - Charles F. Caskey
- Vanderbilt University Institute of Imaging Science, 1161 21st Avenue South, Nashville, TN 37232
- Department of Radiology and Radiological Sciences, Vanderbilt University Medical Center, 1161 21st Avenue South, Nashville, TN 37232
| |
Collapse
|
11
|
Wang M, Zhou Y. Numerical evaluation of the effect of electronically steering a phased array transducer: axially post-focal shifting. Int J Hyperthermia 2017; 33:758-769. [PMID: 28540816 DOI: 10.1080/02656736.2017.1309579] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/15/2022] Open
Abstract
PURPOSE HIFU has been emerging as an effective and safe modality for the treatment of solid tumours and cancers. The focus shifting range of phased array HIFU transducer is an important safety concern because of the presence of grating lobe in the pre-focal region. However, previous studies were only based on linear acoustic wave model. MATERIALS AND METHODS The nonlinear wave propagation from a 256-element phased array through multiple layered media was simulated using the angular spectrum approach (ASA) in marching fractional steps with the consideration of diffraction, attenuation and non-linearity effects by a second-order operator splitting scheme. The distribution of acoustic intensities, temperature elevations, lesion sizes and grating lobe levels were calculated at various axially post-focal shifting distances and driving frequencies. RESULTS Axially shifting HIFU focus leads to significant increase of the acoustic intensity at the grating lobe, but decrease at the main lobe. The influences on the acoustic field, thermal field and lesion sizes are determined by the shifting distance and driving frequency, and variations can be fit monotonically and linearly. Prediction accuracies by simple regression models are satisfactory. Irreversible tissue coagulation could be generated by the grating lobe at certain conditions. CONCLUSIONS The established nonlinear wave propagation algorithm allows the accurate description of HIFU field and consequently the evaluation of grating lobe and steerability of focus. The influence of focus shifting may be predicted simply. The treatment planning of phased array HIFU ablation could be optimised by setting the appropriate exposure and focus scanning schemes.
Collapse
Affiliation(s)
- Mingjun Wang
- a School of Mechanical and Aerospace Engineering, Nanyang Technological University , Singapore
| | - Yufeng Zhou
- a School of Mechanical and Aerospace Engineering, Nanyang Technological University , Singapore
| |
Collapse
|
12
|
Ramaekers P, de Greef M, Berriet R, Moonen CTW, Ries M. Evaluation of a novel therapeutic focused ultrasound transducer based on Fermat’s spiral. Phys Med Biol 2017; 62:5021-5045. [DOI: 10.1088/1361-6560/aa716c] [Citation(s) in RCA: 15] [Impact Index Per Article: 2.1] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/26/2023]
|
13
|
Dou Y, Hynynen K, Allen C. To heat or not to heat: Challenges with clinical translation of thermosensitive liposomes. J Control Release 2017; 249:63-73. [DOI: 10.1016/j.jconrel.2017.01.025] [Citation(s) in RCA: 102] [Impact Index Per Article: 14.6] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/06/2016] [Revised: 01/17/2017] [Accepted: 01/17/2017] [Indexed: 12/20/2022]
|
14
|
Hughes A, Hynynen K. A Tikhonov Regularization Scheme for Focus Rotations With Focused Ultrasound-Phased Arrays. IEEE TRANSACTIONS ON ULTRASONICS, FERROELECTRICS, AND FREQUENCY CONTROL 2016; 63:2008-2017. [PMID: 27913323 PMCID: PMC5218824 DOI: 10.1109/tuffc.2016.2606245] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/06/2023]
Abstract
Phased arrays have a wide range of applications in focused ultrasound therapy. By using an array of individually driven transducer elements, it is possible to steer a focus through space electronically and compensate for acoustically heterogeneous media with phase delays. In this paper, the concept of focusing an ultrasound-phased array is expanded to include a method to control the orientation of the focus using a Tikhonov regularization scheme. It is then shown that the Tikhonov regularization parameter used to solve the ill-posed focus rotation problem plays an important role in the balance between quality focusing and array efficiency. Finally, the technique is applied to the synthesis of multiple foci, showing that this method allows for multiple independent spatial rotations.
Collapse
|
15
|
Deng L, O'Reilly MA, Jones RM, An R, Hynynen K. A multi-frequency sparse hemispherical ultrasound phased array for microbubble-mediated transcranial therapy and simultaneous cavitation mapping. Phys Med Biol 2016; 61:8476-8501. [PMID: 27845920 DOI: 10.1088/0031-9155/61/24/8476] [Citation(s) in RCA: 49] [Impact Index Per Article: 6.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/01/2023]
Abstract
Focused ultrasound (FUS) phased arrays show promise for non-invasive brain therapy. However, the majority of them are limited to a single transmit/receive frequency and therefore lack the versatility to expose and monitor the treatment volume. Multi-frequency arrays could offer variable transmit focal sizes under a fixed aperture, and detect different spectral content on receive for imaging purposes. Here, a three-frequency (306, 612, and 1224 kHz) sparse hemispherical ultrasound phased array (31.8 cm aperture; 128 transducer modules) was constructed and evaluated for microbubble-mediated transcranial therapy and simultaneous cavitation mapping. The array is able to perform effective electronic beam steering over a volume spanning (-40, 40) and (-30, 50) mm in the lateral and axial directions, respectively. The focal size at the geometric center is approximately 0.9 (2.1) mm, 1.7 (3.9) mm, and 3.1 (6.5) mm in lateral (axial) pressure full width at half maximum (FWHM) at 1224, 612, and 306 kHz, respectively. The array was also found capable of dual-frequency excitation and simultaneous multi-foci sonication, which enables the future exploration of more complex exposure strategies. Passive acoustic mapping of dilute microbubble clouds demonstrated that the point spread function of the receive array has a lateral (axial) intensity FWHM between 0.8-3.5 mm (1.7-11.7 mm) over a volume spanning (-25, 25) mm in both the lateral and axial directions, depending on the transmit/receive frequency combination and the imaging location. The device enabled both half and second harmonic imaging through the intact skull, which may be useful for improving the contrast-to-tissue ratio or imaging resolution, respectively. Preliminary in vivo experiments demonstrated the system's ability to induce blood-brain barrier opening and simultaneously spatially map microbubble cavitation activity in a rat model. This work presents a tool to investigate optimal strategies for non-thermal FUS brain therapy and concurrent microbubble cavitation monitoring through the availability of multiple frequencies.
Collapse
Affiliation(s)
- Lulu Deng
- Physical Sciences Platform, Sunnybrook Research Institute, Toronto, ON M4N 3M5, Canada
| | | | | | | | | |
Collapse
|
16
|
Hynynen K, Jones RM. Image-guided ultrasound phased arrays are a disruptive technology for non-invasive therapy. Phys Med Biol 2016; 61:R206-48. [PMID: 27494561 PMCID: PMC5022373 DOI: 10.1088/0031-9155/61/17/r206] [Citation(s) in RCA: 78] [Impact Index Per Article: 9.8] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/31/2022]
Abstract
Focused ultrasound offers a non-invasive way of depositing acoustic energy deep into the body, which can be harnessed for a broad spectrum of therapeutic purposes, including tissue ablation, the targeting of therapeutic agents, and stem cell delivery. Phased array transducers enable electronic control over the beam geometry and direction, and can be tailored to provide optimal energy deposition patterns for a given therapeutic application. Their use in combination with modern medical imaging for therapy guidance allows precise targeting, online monitoring, and post-treatment evaluation of the ultrasound-mediated bioeffects. In the past there have been some technical obstacles hindering the construction of large aperture, high-power, densely-populated phased arrays and, as a result, they have not been fully exploited for therapy delivery to date. However, recent research has made the construction of such arrays feasible, and it is expected that their continued development will both greatly improve the safety and efficacy of existing ultrasound therapies as well as enable treatments that are not currently possible with existing technology. This review will summarize the basic principles, current statures, and future potential of image-guided ultrasound phased arrays for therapy.
Collapse
Affiliation(s)
- Kullervo Hynynen
- Physical Sciences Platform, Sunnybrook Research Institute, Toronto, Canada. Department of Medical Biophysics, University of Toronto, Toronto, Canada. Institute of Biomaterials and Biomedical Engineering, University of Toronto, Toronto, Canada
| | | |
Collapse
|
17
|
Ellens N, Hynynen K. Frequency considerations for deep ablation with high-intensity focused ultrasound: A simulation study. Med Phys 2016; 42:4896-10. [PMID: 26233216 DOI: 10.1118/1.4927060] [Citation(s) in RCA: 14] [Impact Index Per Article: 1.8] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/13/2022] Open
Abstract
PURPOSE The objective of this study was to explore frequency considerations for large-volume, deep thermal ablations with focused ultrasound. Though focal patterns, focal steering rate, and the size of focal clusters have all been explored in this context, frequency studies have generally explored shallower depths and hyperthermia applications. This study examines both treatment efficiency and near-field heating rate as functions of frequency and depth. METHODS Flat, 150 mm transducer arrays were simulated to operate at frequencies of 250, 500, 750, 1000, 1250, and 1500 kHz. Each array had λ2 interelement spacing yielding arrays of 2000-70 000 piston-shaped elements arranged in concentric rings. Depths of 50, 100, and 150 mm were explored, with attenuation (α) values of 2.5-10 (Np/m)/MHz. Ultrasound propagation was simulated with the Rayleigh-Sommerfeld integral over a volume of homogeneous simulated tissue. Absorbed power density was determined from the acoustic pressure which, in turn, was modeled with the Pennes bioheat transfer equation. Using this knowledge of temperature over time, thermal dose function of Sapareto and Dewey was used to model the resulting bioeffect of each simulated sonication. Initially, single foci at each depth, frequency, and α were examined with either fixed peak temperatures or fixed powers. Based on the size of the resulting, single foci lesions, larger compound sonications were designed with foci packed together in multiple layers and rings. For each depth, focal patterns were chosen to produce a similar total ablated volume for each frequency. These compound sonications were performed with a fixed peak temperature at each focus. The resulting energy efficiency (volume ablated per acoustic energy applied), near-field heating rate (temperature increase in the anterior third of the simulation space per unit volume ablated), and near- and far-field margins were assessed. RESULTS Lesions of comparable volume were created with different frequencies at different depths. The results reflect the interconnected nature of frequency as it effects focal size (decreasing with frequency), peak pressure (generally increasing with frequency), and attenuation (also increasing with frequency). The ablation efficiency was the highest for α = 5 (Np/m)/MHz at a frequency of 750 kHz at each depth. For α = 10 (Np/m)/MHz, efficiency was the highest at 750 kHz for a depth of 50 mm, and 500 kHz at depths of 100 and 150 mm. At all sonication depths, near-field heating was minimized with lower frequencies of 250 and 500 kHz. CONCLUSIONS Large-volume ablations are most efficient at frequencies of 500-750 kHz at depths of 100-150 mm. When one considers that near-field heat accumulation tends to be the rate limiting factor in large-volume ablations like uterine fibroid surgery, the results show that frequencies as low as 500 kHz are favored for their ability to reduce heating in the near-field.
Collapse
Affiliation(s)
- Nicholas Ellens
- Sunnybrook Research Institute, Toronto, Ontario M4N 3M5, Canada and Department of Medical Biophysics, University of Toronto, Toronto, Ontario M5G 1L7, Canada
| | - Kullervo Hynynen
- Sunnybrook Research Institute, Toronto, Ontario M4N 3M5, Canada and Department of Medical Biophysics, University of Toronto, Toronto, Ontario M5G 1L7, Canada
| |
Collapse
|
18
|
Ellens NPK, Lucht BBC, Gunaseelan ST, Hudson JM, Hynynen KH. A novel, flat, electronically-steered phased array transducer for tissue ablation: preliminary results. Phys Med Biol 2015; 60:2195-215. [DOI: 10.1088/0031-9155/60/6/2195] [Citation(s) in RCA: 23] [Impact Index Per Article: 2.6] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/13/2022]
|
19
|
de Bever J, Todd N, Payne A, Christensen DA, Roemer RB. Adaptive model-predictive controller for magnetic resonance guided focused ultrasound therapy. Int J Hyperthermia 2014; 30:456-70. [DOI: 10.3109/02656736.2014.968223] [Citation(s) in RCA: 11] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/13/2022] Open
|
20
|
Pajek D, Hynynen K. The application of sparse arrays in high frequency transcranial focused ultrasound therapy: a simulation study. Med Phys 2014; 40:122901. [PMID: 24320540 DOI: 10.1118/1.4829510] [Citation(s) in RCA: 13] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
PURPOSE Transcranial focused ultrasound is an emerging therapeutic modality that can be used to perform noninvasive neurosurgical procedures. The current clinical transcranial phased array operates at 650 kHz, however the development of a higher frequency array would enable more precision, while reducing the risk of standing waves. However, the smaller wavelength and the skull's increased distortion at this frequency are problematic. It would require an order of magnitude more elements to create such an array. Random sparse arrays enable steering of a therapeutic array with fewer elements. However, the tradeoffs inherent in the use of sparsity in a transcranial phased array have not been systematically investigated and so the objective of this simulation study is to investigate the effect of sparsity on transcranial arrays at a frequency of 1.5 MHz that provides small focal spots for precise exposure control. METHODS Transcranial sonication simulations were conducted using a multilayer Rayleigh-Sommerfeld propagation model. Element size and element population were varied and the phased array's ability to steer was assessed. RESULTS The focal pressures decreased proportionally as elements were removed. However, off-focus hotspots were generated if a high degree of steering was attempted with very sparse arrays. A phased array consisting of 1588 elements 3 mm in size, a 10% population, was appropriate for steering up to 4 cm in all directions. However, a higher element population would be required if near-skull sonication is desired. CONCLUSIONS This study demonstrated that the development of a sparse, hemispherical array at 1.5 MHz could enable more precision in therapies that utilize lower intensity sonications.
Collapse
Affiliation(s)
- Daniel Pajek
- Physical Sciences Platform, Sunnybrook Research Institute, Toronto, Ontario M4N 3M5, Canada and Department of Medical Biophysics, University of Toronto, Toronto, Ontario M4N3M5, Canada
| | | |
Collapse
|
21
|
Ellens N, Hynynen K. Simulation study of the effects of near- and far-field heating during focused ultrasound uterine fibroid ablation using an electronically focused phased array: A theoretical analysis of patient safety. Med Phys 2014; 41:072902. [DOI: 10.1118/1.4883777] [Citation(s) in RCA: 23] [Impact Index Per Article: 2.3] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
|
22
|
Malietzis G, Monzon L, Hand J, Wasan H, Leen E, Abel M, Muhammad A, Price P, Abel P. High-intensity focused ultrasound: advances in technology and experimental trials support enhanced utility of focused ultrasound surgery in oncology. Br J Radiol 2013; 86:20130044. [PMID: 23403455 DOI: 10.1259/bjr.20130044] [Citation(s) in RCA: 37] [Impact Index Per Article: 3.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/20/2023] Open
Abstract
High-intensity focused ultrasound (HIFU) is a rapidly maturing technology with diverse clinical applications. In the field of oncology, the use of HIFU to non-invasively cause tissue necrosis in a defined target, a technique known as focused ultrasound surgery (FUS), has considerable potential for tumour ablation. In this article, we outline the development and underlying principles of HIFU, overview the limitations and commercially available equipment for FUS, then summarise some of the recent technological advances and experimental clinical trials that we predict will have a positive impact on extending the role of FUS in cancer therapy.
Collapse
Affiliation(s)
- G Malietzis
- Department of Surgery and Cancer, Faculty of Medicine, Imperial College, London, UK
| | | | | | | | | | | | | | | | | |
Collapse
|