Cole AR, Green AA, Osborne GA, Reece GD. Vacuum grating spectrometer for the mid and far infrared.
APPLIED OPTICS 1970;
9:23-29. [PMID:
20076130 DOI:
10.1364/ao.9.000023]
[Citation(s) in RCA: 3] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/28/2023]
Abstract
The construction and performance of a single beam large aperture spectrometer totally enclosed in a vacuum chamber are described. The focal length of the Czerny-Turner monochromator is 90 cm and plane gratings up to 23 cm x 18 cm are used. The grating drive incorporates a Merton nut. With a Golay detector, the resolution achieved is about 0.12 cm(-1) at 1000 cm(-1), 0.20-0.30 cm(-1) in the range 300-27 cm(-1) Illustrative spectra of ammonia, water vapor, and glyoxal are given.