Malaeb H, Vera MA, Sangal RB, Venkatesh AK, Possick S, Maciejak L, Oberle E, El-Khoury JM. Rapid serum tubes reduce transport hemolysis and false positive rates for high-sensitivity troponin T.
Clin Chim Acta 2023;
551:117630. [PMID:
38420909 DOI:
10.1016/j.cca.2023.117630]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/25/2023] [Revised: 10/31/2023] [Accepted: 11/03/2023] [Indexed: 11/08/2023]
Abstract
INTRODUCTION
Hemolysis in the emergency department (ED) can significantly delay results and appropriate action. We evaluated the main sources of hemolysis during sample collection, and to evaluate the use of rapid serum tubes (RST) as a transport hemolysis-mitigating measure for high-sensitivity troponin T (hs-cTnT) testing.
METHODS
We examined the effect of tube type, tube fill, types of sample draw and collection methods on hemolysis and hs-cTnT in samples (n = 158) from ED patients. We also compared hs-cTnT values in paired RST and plasma separate tube (PST) samples that were hemolysis-free.
RESULTS
The primary source of hemolysis in samples collected in the ED was underfilling tubes. In both tube types, PST and RST, filled tubes showed a median reduction in hemolysis of 69.1 % (p < 0.0001). Blood collected in RST also experienced less hemolysis compared to PST. In hemolysis-free samples, false positive results in PST were noted in patients with hs-cTnT values < 50 ng/l.
CONCLUSION
We suggest that proper tube filling during sample collection and use of RST tubes can significantly reduce the effects of hemolysis. In addition, laboratories should be aware that PST tubes have a non-trivial rate of false positives when hs-cTnT < 50 ng/l.