Abstract
BACKGROUND
Appropriate ICD programming is the key to prevent inappropriate shock delivery, that is closely associated to a negative patients' outcome.
METHODS
Review of the literature on ICD therapy to generate ICD programmings that can be applied to the broad population of ICD and CRT-D carriers.
RESULTS
Arrhythmia detection should occur with a detection time ranging 9″-12″ in the VF zone, and 15″-60″ in the VT zone. Discriminator should be applied at least up to 200 bpm. ATP therapy is applied to all VTs up to 250 bpm, with a success rate of 70%. Inappropriate shocks should occur in <3.6% of patients.
CONCLUSION
Tailored ICD programming can be achieved following evidence from large ICD trials. Pre-defined settings that are saved on the programmer and that can be uploaded at device implantation help to ensure optimal programming and to avoid random errors.
Collapse