Schaffer M, Wagner J, Schaffer B, Schmied M, Mulders H. Automated three-dimensional X-ray analysis using a dual-beam FIB.
Ultramicroscopy 2007;
107:587-97. [PMID:
17267131 DOI:
10.1016/j.ultramic.2006.11.007]
[Citation(s) in RCA: 62] [Impact Index Per Article: 3.6] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/18/2006] [Revised: 11/21/2006] [Accepted: 11/28/2006] [Indexed: 10/23/2022]
Abstract
We present a fully automated method for three-dimensional (3D) elemental analysis demonstrated using a ceramic sample of chemistry (Ca)MgTiO(x). The specimen is serially sectioned by a focused ion beam (FIB) microscope, and energy-dispersive X-ray spectrometry (EDXS) is used for elemental analysis of each cross-section created. A 3D elemental model is reconstructed from the stack of two-dimensional (2D) data. This work concentrates on issues arising from process automation, the large sample volume of approximately 17 x 17 x 10 microm(3), and the insulating nature of the specimen. A new routine for post-acquisition data correction of different drift effects is demonstrated. Furthermore, it is shown that EDXS data may be erroneous for specimens containing voids, and that back-scattered electron images have to be used to correct for these errors.