Abstract
BACKGROUND
Absolute nailfold capillary number should be a putative biomarker in selected rheumatic diseases but could be time-consuming and not highly repeatable.
OBJECTIVE
To validate an automated software for absolute nailfold capillary number and density evaluation, on NVC images in SSc.
METHODS
An automated software to count nailfold capillary number (AUTOCAPI) had been constructed, through an exploratory image set. Subsequently, application rules have been created to define the ROI in NVC images, through a training images set. The software reliability was assessed through calculation of the ICC between automatic and manual counting, by four independent observers, on the same NVC images.
RESULTS
The following ICC's were obtained per observer, for the patients with SSc (40 images), the healthy (20 images), and the PRP subgroups (20 images), respectively: 0.94, 0.81, and 0.62 (observer 1); 0.94, 0.91, and 0.67 (observer 2); 0.88, 0.56, and 0.64 (observer 3); and 0.88, 0.85, and 0.85 (observer 4).
CONCLUSIONS
The validation of an automated software for measuring absolute nailfold capillary number and density in SSc was achieved. The integration into the pre-existing imaging software should make the assessment of the capillary number in NVC easier, quicker, and standardized.