Tanki N, Goto S, Katsuda T, Gotanda R, Gotanda T, Kuwano T. Backside Irradiation of Ultraviolet-A for Correcting Nonuniformity Error of Gafchromic XR-QA2 Films.
J Med Phys 2024;
49:563-567. [PMID:
39926142 PMCID:
PMC11801077 DOI:
10.4103/jmp.jmp_87_24]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/20/2024] [Revised: 11/02/2024] [Accepted: 11/03/2024] [Indexed: 02/11/2025] Open
Abstract
Purpose
Radiochromic film is used for quality assurance and quality control of X-ray equipment in the diagnostic radiology. In addition, three-dimensional dose distribution of computed tomography (CT) is measured. To correct the nonuniformity and uncertainty of radiochromic films for dose measurement of CT, the films are preirradiated ultraviolet (UV)-A rays. There is a difference in the UV protection strength of radiochromic films. A concern exists about the effects of the UV-A irradiation intensity. We thus irradiated with UV-A rays from the backsides of the films to assess if backside irradiation was possible.
Materials and Methods
Gafchromic XR-QA2 and RTQA2 were used in this study. The UV-A rays were simultaneously irradiated on the front and backsides of each film for 12 h. The yellow layer of each film was scanned and imaged. The average pixel values ± standard deviations (SDs) were compared. In the statistical analysis, a paired t-test was performed. To compare, the active-layer densities engendered by the UV-A rays. Calibration curve was created with 48 h of preirradiation of UV-A.
Results
The mean pixel values ± SD for Gafchromic XR-QA2 on the front and backsides were 130.776 ± 0.812 and 81.015 ± 1.128, respectively. On the other hand, the mean pixel values ± SD for Gafchromic RTQA2 on the front and backsides were 62.299 ± 1.077 and 133.761 ± 1.365, respectively. The statistical results of the paired t-test were significantly different (P < 0.01) between both films. Fitting equation of the calibration curve is shown below. y = -390.47 ± 200 + (443.45 ± 10x80).5068 ± 0.0434.
Conclusion
Based on the relationship between the sensitivity of the active layer to UV-A rays and the strength of UV protection on the surface, we concluded that backside irradiation is recommended for Gafchromic XR-QA2, and frontside irradiation is recommended for Gafchromic RTQA2.