Gyekye PK, Schandorf C, Boadu M, Yeboah J, Amoako JK. Patient dose assessment due to fluoroscopic exposure for some selected fluoroscopic procedures in Ghana.
RADIATION PROTECTION DOSIMETRY 2009;
136:203-208. [PMID:
19770212 DOI:
10.1093/rpd/ncp172]
[Citation(s) in RCA: 3] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/28/2023]
Abstract
Organ and effective doses to 90 patients undergoing some selected fluoroscopic examinations at the Korle-Bu Teaching Hospital were estimated using the Monte Carlo-based program (PCXMC version 1.5). Radiation dose was estimated from free-in-air measurements. The mean effective doses were found to be 0.29 +/- 0.07, 0.84 +/- 0.13, 3.15 +/- 0.44, 6.24 +/- 0.70 and 0.38 +/- 0.05 mSv for urethrogram, barium swallow, barium meal, barium enema and myelogram examinations, respectively. The dose area product was estimated to be 3.55 +/- 0.95, 16.44 +/- 2.60, 50.81 +/- 7.04, 99.69 +/- 10.85 and 9.32 +/- 0.99 Gy cm(2) for urethrogram, barium swallow, barium meal, barium enema and myelogram examinations, respectively. Optimisation of procedures is required for barium enema and barium meal examinations.