Erchinger F, Engjom T, Tjora E, Aksnes L, Dimcevskir G, Gudbrandsen OA. Analysis of amylase in duodenal juice - Automated kinetic spectrophotometric analysis versus manual colorimetric endpoint assay.
Pancreatology 2017;
17:182-187. [PMID:
28190684 DOI:
10.1016/j.pan.2017.02.001]
[Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 10/28/2016] [Accepted: 02/01/2017] [Indexed: 02/08/2023]
Abstract
OBJECTIVE
The measurement of duodenal amylase by a colorimetric end-point assay has been the most used method for amylase activity analyses. The method is manual, time consuming and dependent on specialized equipment. In this study, we compare an automated kinetic spectrophotometric method for pancreatic amylase measurement in duodenal juice with a standardized colorimetric end-point assay.
METHODS
We used specimen of duodenal juice at random from a biobank obtained by short endoscopic secretin test in patients with suspected exocrine pancreatic failure of different reasons. Duodenal juice was tested for amylase activity with a conservative manual colorimetric endpoint assay (Phadebas Amylase test, Magle AB) and an automated enzymatic kinetic spectrophotometric method using standard reagents for pancreatic amylase activity for Cobas c111 (Roche Diagnostics).
RESULTS
52 samples for assay of amylase were analyzed in pairs. Correlation between measurements with the two methods was r = 0.99 (p < 0.001), linear regression 0.99 (p < 0.001).
CONCLUSION
Quantification of duodenal amylase activity with automated spectrophotometry has excellent correlation to measurements made by the manual method. This allows for standardized, center independent analyses of duodenal amylase for the assessment of acinar pancreatic function.