Evaluation of absorbable and permanent mesh fixation devices: adhesion formation and mechanical strength.
Hernia 2011;
15:553-8. [PMID:
21594698 DOI:
10.1007/s10029-011-0826-9]
[Citation(s) in RCA: 19] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/20/2010] [Accepted: 04/22/2011] [Indexed: 10/18/2022]
Abstract
PURPOSE
Laparoscopic ventral hernia repair is commonly performed with mesh prostheses; however, there is no standard for fixation devices used to secure mesh to the abdominal wall. This study is a functional comparison of novel, screw-type absorbable and permanent fixation devices with a traditional titanium fixation device.
METHODS
Fifteen pigs each underwent the laparoscopic placement of two 11 × 14-cm mesh prostheses and were randomized for mesh fixation with either titanium spiral tacks (TS), absorbable screw-type fasteners (SF), or permanent screw-type fasteners (PF) (n = 10 mesh prostheses for each fixation group). Adhesions were assessed laparoscopically at 4 weeks. The fixation devices were also embedded in porcine abdominal rectus muscle for ex vivo mechanical testing along with partial thickness polypropylene suture (PR) as a control group (n = 40 for each group). Maximum pull-off forces were measured. All statistical tests were two-tailed, and a P-value < 0.05 was considered to be significant.
RESULTS
The mean tenacity adhesion scores were 1.40 ± 0.52 (PF), 1.7 ± 0.82 (SF), and 2.6 ± 1.07 (TS). Adhesions in the PF group were significantly less tenacious compared with the TS group (P = 0.01). Quantitative adhesion scores were not significantly different among groups. The maximum pull-off forces, measured in Newtons, were 28.61 N ± 4.89 N (TS), 22.71 N ± 7.86 N (SF), 16.98 N ± 7.59 N (PF), and 20.83 N ± 6.25 N (PR). The pull-off force in the TS group was higher than all of the other groups (P < 0.001). The SF group also had a higher pull-off force compared with the PF group (P < 0.001).
CONCLUSIONS
The screw-type absorbable and permanent fixation devices provided adequate fixation and were associated with decreased adhesions in this porcine model.