Berkebile JA, Mabrouk SA, Ganti VG, Srivatsa AV, Sanchez-Perez JA, Inan OT. Towards Estimation of Tidal Volume and Respiratory Timings via Wearable-Patch-Based Impedance Pneumography in Ambulatory Settings.
IEEE Trans Biomed Eng 2022;
69:1909-1919. [PMID:
34818186 PMCID:
PMC9199959 DOI:
10.1109/tbme.2021.3130540]
[Citation(s) in RCA: 4] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/08/2022]
Abstract
OBJECTIVE
Evaluating convenient, wearable multi-frequency impedance pneumography (IP)-based respiratory monitoring in ambulatory persons with novel electrode positioning.
METHODS
A wearable multi-frequency IP system was utilized to estimate tidal volume (TV) and respiratory timings in 14 healthy subjects. A 5.1 cm × 5.1 cm tetrapolar electrode array, affixed to the sternum, and a conventional thoracic electrode configuration were employed to measure the respective IP signals, patch and thoracic IP. Data collected during static postures-sitting and supine-and activities-walking and stair-stepping-were evaluated against a simultaneously-obtained spirometer (SP) volume signal.
RESULTS
Across all measurements, estimated TV obtained from the patch and thoracic IP maintained a Pearson correlation coefficient (r) of 0.93 ± 0.05 and 0.95 ± 0.05 to the ground truth TV, respectively, with an associated root-mean-square error (RMSE) of 0.177 L and 0.129 L, respectively. Average respiration rates (RRs) were extracted from 30-second segments with mean-absolute-percentage errors (MAPEs) of 0.93% and 0.74% for patch and thoracic IP, respectively. Likewise, average inspiratory and expiratory timings were identified with MAPEs less than 6% and 4.5% for patch and thoracic IP, respectively.
CONCLUSION
We demonstrated that patch IP performs comparably to traditional, cumbersome IP configurations. We also present for the first time, to the best of our knowledge, that IP can robustly estimate breath-by-breath TV and respiratory timings during ambulation.
SIGNIFICANCE
This work represents a notable step towards pervasive wearable ambulatory respiratory monitoring via the fusion of a compact chest-worn form factor and multi-frequency IP that can be readily adapted for holistic cardiopulmonary monitoring.