Sung JH, Jeong JS. Dual-/tri-apodization techniques for high frequency ultrasound imaging: a simulation study.
Biomed Eng Online 2014;
13:143. [PMID:
25303996 PMCID:
PMC4271436 DOI:
10.1186/1475-925x-13-143]
[Citation(s) in RCA: 6] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/29/2014] [Accepted: 10/05/2014] [Indexed: 12/04/2022] Open
Abstract
BACKGROUND
In the ultrasound B-mode (Brightness-mode) imaging, high side-lobe level reduces contrast to noise ratio (CNR). A linear apodization scheme by using the window function can suppress the side-lobe level while the main-lobe width is increased resulting in degraded lateral resolution. In order to reduce the side-lobe level without sacrificing the main-lobe width, a non-linear apodization method has been suggested.
METHODS
In this paper, we computationally evaluated the performance of the non-linear apodization method such as dual-/tri-apodization focusing on the high frequency ultrasound image. The rectangular, Dolph-Chebyshev, and Kaiser window functions were employed to implement dual-/tri-apodization algorithms. The point and cyst target simulations were conducted by using a dedicated ultrasound simulation tool called Field-II. The center frequency of the simulated linear array transducer was 40 MHz and the total number of elements was 128. The performance of dual-/tri-apodization was compared with that of the rectangular window function focusing on the side-lobe level and the main-lobe widths (at -6 dB and -35 dB).
RESULTS
In the point target simulation, the main-lobe widths of the dual-/tri-apodization were very similar to that of the rectangular window, and the side-lobe levels of the dual-/tri-apodization were more suppressed by 9~10 dB. In the cyst target simulation, CNR values of the dual-/tri-apodization were improved by 41% and 51%, respectively.
CONCLUSIONS
The performance of the non-linear apodization was numerically investigated. In comparison with the rectangular window function, the non-linear apodization method such as dual- and tri-apodization had low side-lobe level without sacrificing the main-lobe width. Thus, it can be a potential way to increase CNR maintaining the main-lobe width in the high frequency ultrasound imaging.
Collapse