Bennett MT, Brown ML, Koehler J, Lexcen DR, Cheng A, Cheung JW. Trends in implantable cardioverter-defibrillator programming practices and its impact on therapies: Insights from a North American Remote Monitoring Registry 2007-2018.
Heart Rhythm 2021;
19:219-225. [PMID:
34656774 DOI:
10.1016/j.hrthm.2021.10.010]
[Citation(s) in RCA: 2] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 06/20/2021] [Revised: 09/27/2021] [Accepted: 10/11/2021] [Indexed: 11/19/2022]
Abstract
BACKGROUND
Recent evidence has revealed the utility of prolonged arrhythmia detection duration and increased rate cutoff to reduce implantable cardioverter-defibrillator (ICD) therapies. Data on real-world trends in ICD programming and its impact on outcomes are limited.
OBJECTIVE
The purpose of this study was to evaluate trends in ICD programming and its impact on ICD therapy using a large remote monitoring database.
METHODS
A retrospective analysis of patients with ICD implanted from 2007 to 2018 was conducted using the de-identified Medtronic CareLink database. Data on ICD programming (number of intervals to detection [NID] and therapy rate cutoff) and delivered ICD therapies were collected.
RESULTS
Among 210,810 patients, the proportion programmed to a rate cutoff of ≥188 beats/min increased from 41% to 49% and an NID of ≥30/40 increased from 17% to 67% before May 2013 vs after February 2016. Programming to a rate cutoff of ≥188 beats/min, a ventricular fibrillation (VF) NID of ≥30/40, or a combined rate cutoff of ≥188 beats/min and VF NID of ≥30/40 were associated with reductions in ICD therapy. The largest reductions in ICD therapy occurred when the combination of rate cutoff ≥ 188 beats/min and VF NID ≥ 30/40 was programmed (antitachycardia pacing: hazard ratio [HR] 0.35; 95% confidence interval [CI] 0.34-0.36; P < .001; shocks: HR 0.67; 95% CI 0.65-0.69; P < .001; and antitachycardia pacing/shocks: HR 0.43; 95% CI 0.42-0.44; P < .001).
CONCLUSION
Despite evidence supporting the use of prolonged detection duration and high rate cutoff, implementation of shock reduction programming strategies in real-world clinical practice has been modest. The use of evidence-based ICD programming is associated with reduced ICD shocks over long-term follow-up.