Ding Y, Caucci L, Barrett HH. Charged-particle emission tomography.
Med Phys 2017;
44:2478-2489. [PMID:
28370094 PMCID:
PMC5903440 DOI:
10.1002/mp.12245]
[Citation(s) in RCA: 9] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/29/2016] [Revised: 01/20/2017] [Accepted: 03/21/2017] [Indexed: 01/23/2023] Open
Abstract
Purpose
Conventional charged‐particle imaging techniques — such as autoradiography — provide only two‐dimensional (2D) black ex vivo images of thin tissue slices. In order to get volumetric information, images of multiple thin slices are stacked. This process is time consuming and prone to distortions, as registration of 2D images is required. We propose a direct three‐dimensional (3D) autoradiography technique, which we call charged‐particle emission tomography (CPET). This 3D imaging technique enables imaging of thick tissue sections, thus increasing laboratory throughput and eliminating distortions due to registration. CPET also has the potential to enable in vivo charged‐particle imaging with a window chamber or an endoscope.
Methods
Our approach to charged‐particle emission tomography uses particle‐processing detectors (PPDs) to estimate attributes of each detected particle. The attributes we estimate include location, direction of propagation, and/or the energy deposited in the detector. Estimated attributes are then fed into a reconstruction algorithm to reconstruct the 3D distribution of charged‐particle‐emitting radionuclides. Several setups to realize PPDs are designed. Reconstruction algorithms for CPET are developed.
Results
Reconstruction results from simulated data showed that a PPD enables CPET if the PPD measures more attributes than just the position from each detected particle. Experiments showed that a two‐foil charged‐particle detector is able to measure the position and direction of incident alpha particles.
Conclusions
We proposed a new volumetric imaging technique for charged‐particle‐emitting radionuclides, which we have called charged‐particle emission tomography (CPET). We also proposed a new class of charged‐particle detectors, which we have called particle‐processing detectors (PPDs). When a PPD is used to measure the direction and/or energy attributes along with the position attributes, CPET is feasible.
Collapse