1
|
Razinskas G, Schindhelm R, Sauer OA, Wegener S. Sensitivity and specificity of Varian Halcyon's portal dosimetry for plan-specific pre-treatment QA. J Appl Clin Med Phys 2023; 24:e14001. [PMID: 37086428 PMCID: PMC10402680 DOI: 10.1002/acm2.14001] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/21/2022] [Revised: 02/06/2023] [Accepted: 04/03/2023] [Indexed: 04/23/2023] Open
Abstract
PURPOSE Developed as a plan-specific pre-treatment QA tool, Varian portal dosimetry promises a fast, high-resolution, and integrated QA solution. In this study, the agreement between predicted fluence and measured cumulative portal dose was determined for the first 140 patient plans at our Halcyon linear accelerator. Furthermore, the capability of portal dosimetry to detect incorrect plan delivery was compared to that of a common QA phantom. Finally, tolerance criteria for verification of VMAT plan delivery with Varian portal dosimetry were derived. METHODS All patient plans and the corresponding verification plans were generated within the Eclipse treatment planning system. Four representative plans of different treatment sites (prostate, prostate with lymphatic drainage, rectum, and head & neck) were intentionally altered to model incorrect plan delivery. Investigated errors included both systematic and random errors. Gamma analysis was conducted on both portal dose (criteria γ2%/2 mm , γ2%/1 mm , and γ1%/1 mm ) and ArcCHECK measurements (criteria γ3%/3 mm , γ3%/2 mm , and γ2%/2 mm ) with a 10% low-dose threshold. Performance assessment of various acceptance criteria for plan-specific treatment QA utilized receiver operating characteristic (ROC) analysis. RESULTS Predicted and acquired portal dosimetry fluences demonstrated a high agreement evident by average gamma passing rates for the clinical patient plans of 99.90%, 96.64%, and 91.87% for γ2%/2 mm , γ2%/1 mm , and γ1%/1 mm , respectively. The ROC analysis demonstrated a very high capability of detecting erroneous plan delivery for portal dosimetry (area under curve (AUC) > 0.98) and in this regard outperforms QA with the ArcCHECK phantom (AUC ≈ 0.82). With the suggested optimum decision thresholds excellent sensitivity and specificity is simultaneously possible. CONCLUSIONS Owing to the high achievable spatial resolution, portal dosimetry at the Halcyon can reliably be deployed as plan-specific pre-treatment QA tool to screen for errors. It is recommended to support the fluence integrated portal dosimetry QA by independent phantom-based measurements of a random sample survey of treatment plans.
Collapse
Affiliation(s)
- Gary Razinskas
- Department of Radiation OncologyUniversity Hospital WurzburgWurzburgGermany
| | - Robert Schindhelm
- Department of Radiation OncologyUniversity Hospital WurzburgWurzburgGermany
| | - Otto A. Sauer
- Department of Radiation OncologyUniversity Hospital WurzburgWurzburgGermany
| | - Sonja Wegener
- Department of Radiation OncologyUniversity Hospital WurzburgWurzburgGermany
| |
Collapse
|
2
|
Xu Y, Zhang K, Liu Z, Liang B, Ma X, Ren W, Men K, Dai J. Treatment plan prescreening for patient-specific quality assurance measurements using independent Monte Carlo dose calculations. Front Oncol 2022; 12:1051110. [PMID: 36419878 PMCID: PMC9676489 DOI: 10.3389/fonc.2022.1051110] [Citation(s) in RCA: 4] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/22/2022] [Accepted: 10/19/2022] [Indexed: 11/22/2023] Open
Abstract
PURPOSE This study proposes a method to identify plans that failed patient-specific quality assurance (QA) and attempts to establish a criterion to prescreen treatment plans for patient-specific QA measurements with independent Monte Carlo dose calculations. MATERIALS AND METHODS Patient-specific QA results measured with an ArcCHECK diode array of 207 patients (head and neck: 25; thorax: 61; abdomen: 121) were retrospectively analyzed. All patients were treated with the volumetric modulated arc therapy (VMAT) technique and plans were optimized with a Pinnacle v16.2 treatment planning system using an analytical algorithm-based dose engine. Afterwards, phantom verification plans were designed and recalculated by an independent GPU-accelerated Monte Carlo (MC) dose engine, ArcherQA. Moreover, sensitivity and specificity analyzes of gamma passing rates between measurements and MC calculations were carried out to show the ability of MC to monitor failing plans (ArcCHECK 3%/3 mm,<90%), and attempt to determine the appropriate threshold and gamma passing rate criterion utilized by ArcherQA to prescreen treatment plans for ArcCHECK measurements. The receiver operator characteristic (ROC) curve was also utilized to characterize the performance of different gamma passing rate criterion used by ArcherQA. RESULTS The thresholds for 100% sensitivity to detect plans that failed patient-specific QA by independent calculation were 97.0%, 95.4%, and 91.0% for criterion 3%/3 mm, 3%/2 mm, and 2%/2 mm, respectively, which corresponded to specificities of 0.720, 0.528, and 0.585, respectively. It was shown that the 3%/3 mm criterion with 97% threshold for ArcherQA demonstrated perfect sensitivity and the highest specificity compared with other criteria, which may be suitable for prescreening treatment plans treated with the investigated machine to implement measurement-based patient-specific QA of patient plans. In addition, the area under the curve (AUC) calculated from ROC analysis for criterion 3%/3 mm, 3%/2 mm, and 2%/2 mm used by ArcherQA were 0.948, 0.924, and 0.929, respectively. CONCLUSIONS Independent dose calculation with the MC-based program ArcherQA has potential as a prescreen treatment for measurement-based patient-specific QA. AUC values (>0.9) showed excellent classification accuracy for monitoring failing plans with independent MC calculations.
Collapse
Affiliation(s)
| | | | | | | | | | | | - Kuo Men
- Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China
| | - Jianrong Dai
- Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China
| |
Collapse
|
3
|
Goodall SK, Norvill C. Variation in Elekta iView electronic portal imager pixel scale factor with gantry angle, and impact on multi-leaf collimator quality assurance. J Appl Clin Med Phys 2022; 23:e13661. [PMID: 35666629 PMCID: PMC9278680 DOI: 10.1002/acm2.13661] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/29/2021] [Revised: 05/01/2022] [Accepted: 05/10/2022] [Indexed: 12/31/2022] Open
Abstract
For Elekta Agility linear accelerators, the iViewGT electronic portal imaging device (EPID) is positioned at a nominal X‐Ray source‐to‐panel distance of 1600 mm. For display, image registration, and data processing purposes, the image pixels are scaled to spatial units at the treatment isocenter plane. This is achieved by applying a pixel scaling factor (PSF). During this investigation, the dependence of the PSF at cardinal gantry angles was determined along with the resulting effects on the multi‐leaf collimator (MLC) quality assurance (QA) results for three linear accelerators (linacs). The PSF was found to vary by 0.0018–0.0022 mm/pixel during gantry rotation, which resulted in variations in the mean MLC reported error of up to 0.8 mm at 100 mm off‐axis with the gantry rotated to 180°. Measurement and application of a gantry angle–specific PSF is a simple process that can be implemented to improve the accuracy of EPID‐based MLC QA at cardinal gantry angles.
Collapse
Affiliation(s)
- Simon K Goodall
- GenesisCare, Wembley, Western Australia, Australia.,School of Physics, Mathematics, and Computing, Faculty of Engineering and Mathematical Sciences, University of Western Australia, Crawley, Western Australia, Australia
| | | |
Collapse
|
4
|
Ahmad M, Nourzadeh H, Siebers J. A regression-based approach to compute the pixels sensitivity map of linear accelerator portal imaging devices. Med Phys 2021; 48:4598-4609. [PMID: 33774827 DOI: 10.1002/mp.14862] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/17/2020] [Revised: 02/03/2021] [Accepted: 03/10/2021] [Indexed: 11/06/2022] Open
Abstract
PURPOSE To determine the pixel sensitivity map (PSM) for amorphous silicon electronic portal imaging devices (EPIDs) using a single flood field signal. METHOD AND MATERIALS A raw EPID signal results from the incident particle energy fluence, the inherent pixels response, and the background signal. In large open fields, particle energy fluence is a slow-varying signal that is locally considered spatially constant. Pixels response is a fast and abrupt varying behavior. The background signal is due to the EPID panel electronics, which is determined during radiation absence. To determine the PSM, after correcting for the background signal, we apply a model that captures the underlying smooth particle energy fluence-induced signal. This fluence signal-fitted model is then used to determine the PSM. Here, we use a polynomial-based regression surface model in both x and y dimensions. To validate the generated PSM, we measure beams and compute PSMs for multiple beam energies with and without flattening filters and for multiple source-to-imager distances. Since the PSM is a detector characteristic, it should be independent of those variables. We also intercompare measurements of fixed slit fields with the EPID being shifted between measurements. RESULTS The fluence signal of the flattening filter-free (FFF) beams was optimally modeled as a 12th degree polynomial surfaces, which had ≤ 0.1% residuals near the central axis. The 6 and 10 MV FFF PSMs were within ˜0.1%, and independent of the EPID SID, suggesting that the PSM is energy independent. The 6, 10, and 15 MV flattened-beam PSMs were well modeled as 12th degree polynomial surfaces, which were equivalent within ˜0.24% but differed from the FFF PSM by up to 0.5% near the beam central axis. Applying the FFF PSMs to the flattened-beam measurements reduced the central-axis deviation between the raw and corrected signal to < 0.1%, confirming the PSM energy independence hypothesis. When the FFF PSM is utilized, output verification with shifted slit deliveries agreed within ˜0.5% for all beam energies, which is within the radiation delivery uncertainty of ˜0.57%. CONCLUSION PSM for MV EPIDs can be determined by separating out the slowly varying, well-behaved fluence signal from the pixel-to-pixel sensitivity variations. The quality of the PSM is found to be dependent on the quality of the surface fit, which is best for the 6 MV FFF beam measured at SID equal to 180 cm. Within fitting errors, the PSM is independent of beam energy for 6, 10, and 15 MV beams with and without flattening filters. The PSM generation does not require shifting the EPID panel nor multiple EPID panel irradiations and should be usable for linacs with fixed geometry EPIDs.
Collapse
Affiliation(s)
- Mahmoud Ahmad
- Vanderbilt University Medical Center, Nashville, TN, 37212, USA.,Radiation Oncology Department, University of Virginia, Charlottesville, VA, 22908, USA
| | - Hamidreza Nourzadeh
- Radiation Oncology Department, Sidney Kimmel Medical College and Cancer Center at Thomas Jefferson University, Philadelphia, PA, 19107, USA
| | - Jeffrey Siebers
- Radiation Oncology Department, University of Virginia, Charlottesville, VA, 22908, USA
| |
Collapse
|
5
|
Hanley J, Dresser S, Simon W, Flynn R, Klein EE, Letourneau D, Liu C, Yin FF, Arjomandy B, Ma L, Aguirre F, Jones J, Bayouth J, Holmes T. AAPM Task Group 198 Report: An implementation guide for TG 142 quality assurance of medical accelerators. Med Phys 2021; 48:e830-e885. [PMID: 34036590 DOI: 10.1002/mp.14992] [Citation(s) in RCA: 47] [Impact Index Per Article: 11.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/09/2021] [Revised: 03/16/2021] [Accepted: 04/28/2021] [Indexed: 11/11/2022] Open
Abstract
The charges on this task group (TG) were as follows: (a) provide specific procedural guidelines for performing the tests recommended in TG 142; (b) provide estimate of the range of time, appropriate personnel, and qualifications necessary to complete the tests in TG 142; and (c) provide sample daily, weekly, monthly, or annual quality assurance (QA) forms. Many of the guidelines in this report are drawn from the literature and are included in the references. When literature was not available, specific test methods reflect the experiences of the TG members (e.g., a test method for door interlock is self-evident with no literature necessary). In other cases, the technology is so new that no literature for test methods was available. Given broad clinical adaptation of volumetric modulated arc therapy (VMAT), which is not a specific topic of TG 142, several tests and criteria specific to VMAT were added.
Collapse
Affiliation(s)
- Joseph Hanley
- Princeton Radiation Oncology, Monroe, New Jersey, 08831, USA
| | - Sean Dresser
- Winship Cancer Institute, Radiation Oncology, Emory University, Atlanta, Georgia, 30322, USA
| | | | - Ryan Flynn
- Department of Radiation Oncology, University of Iowa, Iowa City, Iowa, 52242, USA
| | - Eric E Klein
- Brown university, Rhode Island Hospital, Providence, Rhode Island, 02905, USA
| | | | - Chihray Liu
- University of Florida, Gainesville, Florida, 32610-0385, USA
| | - Fang-Fang Yin
- Department of Radiation Oncology, Duke University Medical Center, Durham, North Carolina, 27710, USA
| | - Bijan Arjomandy
- Karmanos Cancer Institute at McLaren-Flint, Flint, Michigan, 48532, USA
| | - Lijun Ma
- Department of Radiation Oncology, University of California San Francisco, San Francisco, 94143-0226, USA
| | | | - Jimmy Jones
- Department of Radiation Oncology, The University of Colorado Health-Poudre Valley, Fort Collins, Colorado, 80525, USA
| | - John Bayouth
- Department of Human Oncology, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, 53792-0600, USA
| | - Todd Holmes
- Varian Medical Systems, Palo Alto, California, 94304, USA
| |
Collapse
|
6
|
Inui S, Ueda Y, Ono S, Ohira S, Isono M, Nitta Y, Ueda H, Miyazaki M, Koizumi M, Teshima T. Evaluation of two-dimensional electronic portal imaging device using integrated images during volumetric modulated arc therapy for prostate cancer. ACTA ACUST UNITED AC 2021; 26:281-290. [PMID: 34211779 DOI: 10.5603/rpor.a2021.0041] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/24/2020] [Accepted: 02/08/2021] [Indexed: 01/04/2023]
Abstract
Background The aim of the study was to evaluate analysis criteria for the identification of the presence of rectal gas during volumetric modulated arc therapy (VMAT) for prostate cancer patients by using electronic portal imaging device (EPID)-based in vivo dosimetry (IVD). Materials and methods All measurements were performed by determining the cumulative EPID images in an integrated acquisition mode and analyzed using PerFRACTION commercial software. Systematic setup errors were simulated by moving the anthropomorphic phantom in each translational and rotational direction. The inhomogeneity regions were also simulated by the I'mRT phantom attached to the Quasar phantom. The presence of small and large air cavities (12 and 48 cm3) was controlled by moving the Quasar phantom in several timings during VMAT. Sixteen prostate cancer patients received EPID-based IVD during VMAT. Results In the phantom study, no systematic setup error was detected in the range that can happen in clinical (< 5-mm and < 3 degree). The pass rate of 2% dose difference (DD2%) in small and large air cavities was 98.74% and 79.05%, respectively, in the appearance of the air cavity after irradiation three quarter times. In the clinical study, some fractions caused a sharp decline in the DD2% pass rate. The proportion for DD2% < 90% was 13.4% of all fractions. Rectal gas was confirmed in 11.0% of fractions by acquiring kilo-voltage X-ray images after the treatment. Conclusions Our results suggest that analysis criteria of 2% dose difference in EPID-based IVD was a suitable method for identification of rectal gas during VMAT for prostate cancer patients.
Collapse
Affiliation(s)
- Shoki Inui
- Department of Radiation Oncology, Osaka International Cancer Institute, Osaka, Japan.,Department of Medical Physics and Engineering, Osaka University Graduate School of Medicine, Suita, Japan
| | - Yoshihiro Ueda
- Department of Radiation Oncology, Osaka International Cancer Institute, Osaka, Japan
| | - Shunsuke Ono
- Department of Radiation Oncology, Osaka International Cancer Institute, Osaka, Japan
| | - Shingo Ohira
- Department of Radiation Oncology, Osaka International Cancer Institute, Osaka, Japan
| | - Masaru Isono
- Department of Radiation Oncology, Osaka International Cancer Institute, Osaka, Japan
| | - Yuya Nitta
- Department of Radiation Oncology, Osaka International Cancer Institute, Osaka, Japan
| | - Hikari Ueda
- Department of Radiation Oncology, Osaka International Cancer Institute, Osaka, Japan
| | - Masayoshi Miyazaki
- Department of Radiation Oncology, Osaka International Cancer Institute, Osaka, Japan
| | - Masahiko Koizumi
- Department of Medical Physics and Engineering, Osaka University Graduate School of Medicine, Suita, Japan
| | - Teruki Teshima
- Department of Radiation Oncology, Osaka International Cancer Institute, Osaka, Japan
| |
Collapse
|
7
|
Zwan BJ, Caillet V, Booth JT, Colvill E, Fuangrod T, O'Brien R, Briggs A, O'Connor DJ, Keall PJ, Greer PB. Toward real-time verification for MLC tracking treatments using time-resolved EPID imaging. Med Phys 2021; 48:953-964. [PMID: 33354787 DOI: 10.1002/mp.14675] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/16/2020] [Revised: 11/29/2020] [Accepted: 12/03/2020] [Indexed: 12/25/2022] Open
Abstract
PURPOSE In multileaf collimator (MLC) tracking, the MLC positions from the original treatment plan are continuously modified to account for intrafraction tumor motion. As the treatment is adapted in real time, there is additional risk of delivery errors which cannot be detected using traditional pretreatment dose verification. The purpose of this work is to develop a system for real-time geometric verification of MLC tracking treatments using an electronic portal imaging device (EPID). METHODS MLC tracking was utilized during volumetric modulated arc therapy (VMAT). During these deliveries, treatment beam images were taken at 9.57 frames per second using an EPID and frame grabber computer. MLC positions were extracted from each image frame and used to assess delivery accuracy using three geometric measures: the location, size, and shape of the radiation field. The EPID-measured field location was compared to the tumor motion measured by implanted electromagnetic markers. The size and shape of the beam were compared to the size and shape from the original treatment plan, respectively. This technique was validated by simulating errors in phantom test deliveries and by comparison between EPID measurements and treatment log files. The method was applied offline to images acquired during the LIGHT Stereotactic Ablative Body Radiotherapy (SABR) clinical trial, where MLC tracking was performed for 17 lung cancer patients. The EPID-based verification results were subsequently compared to post-treatment dose reconstruction. RESULTS Simulated field location errors were detected during phantom validation tests with an uncertainty of 0.28 mm (parallel to MLC motion) and 0.38 mm (perpendicular), expressed as a root-mean-square error (RMSError ). For simulated field size errors, the RMSError was 0.47 cm2 and field shape changes were detected for random errors with standard deviation ≥ 2.5 mm. For clinical lung SABR deliveries, field location errors of 1.6 mm (parallel MLC motion) and 4.9 mm (perpendicular) were measured (expressed as a full-width-half-maximum). The mean and standard deviation of the errors in field size and shape were 0.0 ± 0.3 cm2 and 0.3 ± 0.1 (expressed as a translation-invariant normalized RMS). No correlation was observed between geometric errors during each treatment fraction and dosimetric errors in the reconstructed dose to the target volume for this cohort of patients. CONCLUSION A system for real-time delivery verification has been developed for MLC tracking using time-resolved EPID imaging. The technique has been tested offline in phantom-based deliveries and clinical patient deliveries and was used to independently verify the geometric accuracy of the MLC during MLC tracking radiotherapy.
Collapse
Affiliation(s)
- Benjamin J Zwan
- School of Mathematical and Physical Sciences, University of Newcastle, Newcastle, NSW, Australia.,Northern Sydney Cancer Centre, Royal North Shore Hospital, St Leonards, NSW, Australia
| | - Vincent Caillet
- Northern Sydney Cancer Centre, Royal North Shore Hospital, St Leonards, NSW, Australia.,ACRF Image X Institute, School of Health Sciences, University of Sydney, Sydney, NSW, Australia
| | - Jeremy T Booth
- Northern Sydney Cancer Centre, Royal North Shore Hospital, St Leonards, NSW, Australia.,Institute of Medical Physics, School of Physics, University of Sydney, Sydney, Australia
| | - Emma Colvill
- Northern Sydney Cancer Centre, Royal North Shore Hospital, St Leonards, NSW, Australia.,ACRF Image X Institute, School of Health Sciences, University of Sydney, Sydney, NSW, Australia
| | - Todsaporn Fuangrod
- School of Mathematical and Physical Sciences, University of Newcastle, Newcastle, NSW, Australia.,Faculty of Medicine and Public Health HRH Princess Chulabhorn College of Medical Science, Chulabhorn Royal Academy, Bangkok, Thailand
| | - Ricky O'Brien
- ACRF Image X Institute, School of Health Sciences, University of Sydney, Sydney, NSW, Australia
| | - Adam Briggs
- Northern Sydney Cancer Centre, Royal North Shore Hospital, St Leonards, NSW, Australia
| | - Daryl J O'Connor
- School of Mathematical and Physical Sciences, University of Newcastle, Newcastle, NSW, Australia
| | - Paul J Keall
- ACRF Image X Institute, School of Health Sciences, University of Sydney, Sydney, NSW, Australia
| | - Peter B Greer
- School of Mathematical and Physical Sciences, University of Newcastle, Newcastle, NSW, Australia.,Department of Radiation Oncology, Calvary Mater Hospital, Newcastle, NSW, Australia
| |
Collapse
|
8
|
Harris TC, Seco J, Ferguson D, Lehmann M, Huber P, Shi M, Jacobson M, Valencia Lozano I, Myronakis M, Baturin P, Fueglistaller R, Morf D, Berbeco R. Clinical translation of a new flat-panel detector for beam's-eye-view imaging. Phys Med Biol 2020; 65:225004. [PMID: 33284786 PMCID: PMC9142212 DOI: 10.1088/1361-6560/abb571] [Citation(s) in RCA: 5] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/25/2022]
Abstract
Electronic portal imaging devices (EPIDs) lend themselves to beams-eye view clinical applications, such as tumor tracking, but are limited by low contrast and detective quantum efficiency (DQE). We characterize a novel EPID prototype consisting of multiple layers and investigate its suitability for use under clinical conditions. A prototype multi-layer imager (MLI) was constructed utilizing four conventional EPID layers, each consisting of a copper plate, a Gd2O2S:Tb phosphor scintillator, and an amorphous silicon flat panel array detector. We measured the detector's response to a 6 MV photon beam with regards to modulation transfer function, noise power spectrum, DQE, contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR), and the linearity of the detector's response to dose. Additionally, we compared MLI performance to the single top layer of the MLI and the standard Varian AS-1200 detector. Pre-clinical imaging was done on an anthropomorphic phantom, and the detector's CNR, SNR and spatial resolution were assessed in a clinical environment. Images obtained from spine and liver patient treatment deliveries were analyzed to verify CNR and SNR improvements. The MLI has a DQE(0) of 9.7%, about 5.7 times the reference AS-1200 detector. Improved noise performance largely drives the increase. CNR and SNR of clinical images improved three-fold compared to reference. A novel MLI was characterized and prepared for clinical translation. The MLI substantially improved DQE and CNR performance while maintaining the same resolution. Pre-clinical tests on an anthropomorphic phantom demonstrated improved performance as predicted theoretically. Preliminary patient data were analyzed, confirming improved CNR and SNR. Clinical applications are anticipated to include more accurate soft tissue tracking.
Collapse
Affiliation(s)
- T C Harris
- Department of Radiation Oncology, Dana Farber/Brigham and Women's Cancer Center, Harvard Medical school, Boston, MA, United States of America
- BioMedical Physics in Radiation Oncology, DKFZ, Heidelberg, Germany
- Department of Physics, University of Heidelberg, Heidelberg, Germany
| | - J Seco
- BioMedical Physics in Radiation Oncology, DKFZ, Heidelberg, Germany
- Department of Physics, University of Heidelberg, Heidelberg, Germany
| | - D Ferguson
- Department of Radiation Oncology, Dana Farber/Brigham and Women's Cancer Center, Harvard Medical school, Boston, MA, United States of America
| | - M Lehmann
- Varian Medical Systems, Baden-Dattwil, Switzerland
| | - P Huber
- Varian Medical Systems, Baden-Dattwil, Switzerland
| | - M Shi
- Department of Radiation Oncology, Dana Farber/Brigham and Women's Cancer Center, Harvard Medical school, Boston, MA, United States of America
- University of Massachusetts Lowell, Lowell, MA, United States of America
| | - M Jacobson
- Department of Radiation Oncology, Dana Farber/Brigham and Women's Cancer Center, Harvard Medical school, Boston, MA, United States of America
| | - I Valencia Lozano
- Department of Radiation Oncology, Dana Farber/Brigham and Women's Cancer Center, Harvard Medical school, Boston, MA, United States of America
| | - M Myronakis
- Department of Radiation Oncology, Dana Farber/Brigham and Women's Cancer Center, Harvard Medical school, Boston, MA, United States of America
| | - P Baturin
- Varian Medical System, Palo Alto, CA, United States of America
| | | | - D Morf
- Varian Medical Systems, Baden-Dattwil, Switzerland
| | - R Berbeco
- Department of Radiation Oncology, Dana Farber/Brigham and Women's Cancer Center, Harvard Medical school, Boston, MA, United States of America
| |
Collapse
|
9
|
McCulloch J, Pawlowski J, Kirby N, Rasmussen K, Shi Z, Myers P, Stathakis S, Papanikolaou N, Saenz DL. Patient-specific dose quality assurance of single-isocenter multiple brain metastasis stereotactic radiosurgery using PTW Octavius 4D. J Appl Clin Med Phys 2020; 21:107-115. [PMID: 32681753 PMCID: PMC7497916 DOI: 10.1002/acm2.12979] [Citation(s) in RCA: 11] [Impact Index Per Article: 2.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/23/2020] [Revised: 06/18/2020] [Accepted: 06/21/2020] [Indexed: 11/16/2022] Open
Abstract
Purpose Single‐isocenter multiple brain metastasis stereotactic radiosurgery is an efficient treatment modality increasing in clinical practice. The need to provide accurate, patient‐specific quality assurance (QA) for these plans is met by several options. This study reviews some of these options and explores the use of the Octavius 4D as a solution for patient‐specific plan quality assurance. Methods The Octavius 4D Modular Phantom (O4D) with the 1000 SRS array was evaluated in this study. The array consists of 977 liquid‐filled ion chambers. The center 5.5 cm × 5.5 cm area has a detector spacing of 2.5 mm. The ability of the O4D to reconstruct three‐dimensional (3D) dose was validated against a 3D gel dosimeter, ion chamber, and film measurements. After validation, 15 patients with 2–11 targets had their plans delivered to the phantom. The criteria used for the gamma calculation was 3%/1 mm. The portion of targets which were measurable by the phantom was countable. The accompanying software compiled the measured doses allowing each target to be counted from the measured dose distribution. Results Spatial resolution was sufficient to verify the high dose distributions characteristic of SRS. Amongst the 15 patients there were 74 targets. Of the 74 targets, 61 (82%) of them were visible on the measured dose distribution. The average gamma passing rate was 99.3% (with sample standard deviation of 0.68%). Conclusions The high resolution provided by the O4D with 1000 SRS board insert allows for very high‐resolution measurement. This high resolution in turn can allow for high gamma passing rates. The O4D with the 1000 SRS array is an acceptable method of performing quality assurance for single‐isocenter multiple brain metastasis SRS.
Collapse
Affiliation(s)
- James McCulloch
- Department of Radiation Oncology, University of Texas Health San Antonio, San Antonio, TX, USA
| | - Jamie Pawlowski
- Department of Radiation Oncology, University of Texas Health San Antonio, San Antonio, TX, USA
| | - Neil Kirby
- Department of Radiation Oncology, University of Texas Health San Antonio, San Antonio, TX, USA
| | - Karl Rasmussen
- Department of Radiation Oncology, University of Texas Health San Antonio, San Antonio, TX, USA
| | - Zheng Shi
- Department of Radiation Oncology, University of Texas Health San Antonio, San Antonio, TX, USA
| | - Pamela Myers
- Department of Radiation Oncology, University of Texas Health San Antonio, San Antonio, TX, USA
| | - Sotirios Stathakis
- Department of Radiation Oncology, University of Texas Health San Antonio, San Antonio, TX, USA
| | - Niko Papanikolaou
- Department of Radiation Oncology, University of Texas Health San Antonio, San Antonio, TX, USA
| | - Daniel L Saenz
- Department of Radiation Oncology, University of Texas Health San Antonio, San Antonio, TX, USA
| |
Collapse
|
10
|
Alharthi T, Arumugam S, Vial P, Holloway L, Thwaites D. EPID sensitivity to delivery errors for pre-treatment verification of lung SBRT VMAT plans. Phys Med 2019; 59:37-46. [PMID: 30928064 DOI: 10.1016/j.ejmp.2019.02.007] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 12/12/2018] [Revised: 01/31/2019] [Accepted: 02/10/2019] [Indexed: 12/31/2022] Open
Abstract
PURPOSE To study the sensitivity of an Electronic Portal Imaging Device (EPID) in detecting delivery errors for VMAT lung stereotactic body radiotherapy (SBRT) using the Collapsed Arc method. METHODS Baseline VMAT plans and plans with errors intentionally introduced were generated for 15 lung SBRT patients. Three types of errors were introduced by modifying collimator angles and multi-leaf collimator (MLC) field sizes (MLCFS) and MLC shifts by ±5, ±2, and ±1° or millimeters. A total of 103 plans were measured with EPID on an Elekta Synergy Linear Accelerator (Agility MLC) and compared to both the original treatment planning system (TPS) Collapsed Arc dose matrix and the no-error plan baseline EPID measurements. Gamma analysis was performed using the OmniPro-I'mRT (IBA Dosimetry) software and gamma criteria of 1%/1 mm, 2%/1 mm, 2%/2 mm, and 3%/3. RESULTS When the error-introduced EPID measured dose matrices were compared to the TPS matrices, the majority of simulated errors were detected with gamma tolerance of 2%/1 mm and 1%/1 mm. When the error-introduced EPID measured dose matrices were compared to the baseline EPID measurements, all the MLCFS and MLC shift errors, and ±5°collimator errors were detected using 2%/1 mm and 1%/1 mm gamma criteria. CONCLUSION This work demonstrates the feasibility and effectiveness of the collapsed arc technique and EPID for pre-treatment verification of lung SBRT VMAT plans. The EPID was able to detect the majority of MLC and the larger collimator errors with sensitivity to errors depending on the gamma tolerances.
Collapse
Affiliation(s)
- Thahabah Alharthi
- Institute of Medical Physics, School of Physics, The University of Sydney, Sydney, New South Wales, Australia; School of Medicine, Taif University, Taif, Saudi Arabia; Liverpool and Macarthur Cancer Therapy Centers, Liverpool, NSW, Australia; Ingham Institute for Applied Medical Research, Sydney, NSW, Australia.
| | - Sankar Arumugam
- Liverpool and Macarthur Cancer Therapy Centers, Liverpool, NSW, Australia; South Western Sydney Clinical School, University of New South Wales, Sydney, NSW, Australia
| | - Phil Vial
- Liverpool and Macarthur Cancer Therapy Centers, Liverpool, NSW, Australia; Ingham Institute for Applied Medical Research, Sydney, NSW, Australia; South Western Sydney Clinical School, University of New South Wales, Sydney, NSW, Australia
| | - Lois Holloway
- Institute of Medical Physics, School of Physics, The University of Sydney, Sydney, New South Wales, Australia; Liverpool and Macarthur Cancer Therapy Centers, Liverpool, NSW, Australia; Ingham Institute for Applied Medical Research, Sydney, NSW, Australia; South Western Sydney Clinical School, University of New South Wales, Sydney, NSW, Australia; Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW, Australia
| | - David Thwaites
- Institute of Medical Physics, School of Physics, The University of Sydney, Sydney, New South Wales, Australia
| |
Collapse
|
11
|
Hiatt J, Mukwada G, Barnes M, Riis HL, Huynh D, Rowshanfarzad P. MLC positioning verification for small fields: a new investigation into automatic EPID-based verification methods. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE 2018; 41:945-955. [PMID: 30259333 DOI: 10.1007/s13246-018-0690-3] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Subscribe] [Scholar Register] [Received: 03/13/2018] [Accepted: 09/24/2018] [Indexed: 12/31/2022]
Abstract
Multileaf-collimator (MLC) defined small fields in radiotherapy are used in high dose, ultra-conformal techniques such as stereotactic radiotherapy and stereotactic radiosurgery. Proximity to critical structures and irreversible damage arising from inaccurate delivery mean that correct positioning of the MLC system is of the utmost importance. Some of the existing techniques for MLC positioning quality assurance make use of electronic portal imaging device (EPID) images. However, conventional collimation verification algorithms based on the full width at half maximum (FWHM) fail when applied to small field images acquired by an EPID due to overlapping aperture penumbrae, lateral electron disequilibrium and radiation source occlusion. The objective of this study was to investigate sub-pixel edge detection and other techniques with the aim of developing an automatic and autonomous EPID-based method suitable for MLC positional verification of small static fields with arbitrary shapes. Methods investigated included derivative interpolation, Laplacian of Gaussian (LoG) and an algorithm based on the partial area effect hypothesis. None of these methods were found to be suitable for MLC positioning verification in small field conditions. A method is proposed which uses a manufacturer-specific empirically modified FWHM algorithm which shows improvement over the conventional techniques in the small field size range. With a measured mean absolute difference from planned position for Varian linacs of 0.01 ± 0.26 mm, compared with the erroneous FWHM value of 0.70 ± 0.51 mm. For Elekta linacs the proposed algorithm returned 0.26 ± 0.25 mm, in contrast to the FWHM result of 1.79 ± 1.07 mm.
Collapse
Affiliation(s)
- Joshua Hiatt
- Department of Radiation Oncology, Liverpool & Macarthur Cancer Therapy Centres, Liverpool, NSW, 2170, Australia. .,School of Physics, Mathematics and Computing, Faculty of Engineering and Mathematical Sciences, The University of Western Australia, Crawley, WA, 6009, Australia.
| | - Godfrey Mukwada
- Department of Radiation Oncology, Sir Charles Gairdner Hospital, Nedlands, WA, 6009, Australia
| | - Michael Barnes
- Department of Radiation Oncology, Calvary Mater Newcastle Hospital, Newcastle, NSW, 2310, Australia.,University of Newcastle, Newcastle, NSW, 2308, Australia
| | | | - Du Huynh
- School of Physics, Mathematics and Computing, Faculty of Engineering and Mathematical Sciences, The University of Western Australia, Crawley, WA, 6009, Australia
| | - Pejman Rowshanfarzad
- School of Physics, Mathematics and Computing, Faculty of Engineering and Mathematical Sciences, The University of Western Australia, Crawley, WA, 6009, Australia
| |
Collapse
|
12
|
Vazquez-Quino LA, Huerta-Hernandez CI, Rangaraj D. Clinical experience with machine log file software for volumetric-modulated arc therapy techniques. Proc (Bayl Univ Med Cent) 2017; 30:276-279. [PMID: 28670056 DOI: 10.1080/08998280.2017.11929614] [Citation(s) in RCA: 13] [Impact Index Per Article: 1.6] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/18/2022] Open
Abstract
Mobius FX, an add-on software module from Mobius Medical Systems™ for intensity-modulated radiation therapy (IMRT) quality assurance (QA), uses linac treatment logs to efficiently calculate and verify the 3D dose delivered to patients. An advantage of the Mobius FX module is that it does not require device positioning. In this study, we compared the Mobius FX with another QA option, ArcCheck, as well as with the treatment planning system (TPS) using 30 volumetric-modulated arc therapy (VMAT) plans planned and delivered on a Varian TrueBeam linac. The plans, which involved 6 and 10 MV and consisted of 2 to 3 arcs per plan, were selected to provide a clinically relevant sample. The average gamma value for all plans between Mobius FX and the TPS was 99.96% for the criterion of 3%-3 mm and 98.80% for the criterion of 2%-2 mm. Very similar results were found when comparing Mobius FX and the TPS dose calculations with those acquired by traditional methods (i.e., ArcCheck). As the gamma criterion of the analysis was narrowed, discrepancies between Mobius FX and traditional methods appeared. Profile analysis showed the production of comparable results when using the Mobius FX method or traditional QA methods. In conclusion, the Mobius FX method for pretreatment of patient-specific QA is capable of producing results similar to those obtained by traditional methods.
Collapse
Affiliation(s)
- Luis Alberto Vazquez-Quino
- Radiation Oncology Department, Baylor Scott & White Health and Texas A&M Health Science Center College of Medicine, Temple, Texas
| | - Claudia Ivette Huerta-Hernandez
- Radiation Oncology Department, Baylor Scott & White Health and Texas A&M Health Science Center College of Medicine, Temple, Texas
| | - Dharanipathy Rangaraj
- Radiation Oncology Department, Baylor Scott & White Health and Texas A&M Health Science Center College of Medicine, Temple, Texas
| |
Collapse
|
13
|
Noufal MP, Abdullah KK, Niyas P, Subha PAR. Study of impacts of different evaluation criteria on gamma pass rates in VMAT QA using MatriXX and EPID. POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING 2017. [DOI: 10.1515/pjmpe-2017-0017] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/15/2022]
Abstract
Abstract
Aim: This study evaluates the impacts of using different evaluation criteria on gamma pass rates in two commercially available QA methods employed for the verification of VMAT plans using different hypothetical planning target volumes (PTVs) and anatomical regions.
Introduction: Volumetric modulated arc therapy (VMAT) is a widely accepted technique to deliver highly conformal treatment in a very efficient manner. As their level of complexity is high in comparison to intensity-modulated radiotherapy (IMRT), the implementation of stringent quality assurance (QA) before treatment delivery is of paramount importance.
Material and Methods: Two sets of VMAT plans were generated using Eclipse planning systems, one with five different complex hypothetical three-dimensional PTVs and one including three anatomical regions. The verification of these plans was performed using a MatriXX ionization chamber array embedded inside a MultiCube phantom and a Varian EPID dosimetric system attached to a Clinac iX. The plans were evaluated based on the 3%/3 mm, 2%/2 mm, and 1%/1 mm global gamma criteria and with three low-dose threshold values (0%, 10%, and 20%).
Results: The gamma pass rates were above 95% in all VMAT plans, when the 3%/3mm gamma criterion was used and no threshold was applied. In both systems, the pass rates decreased as the criteria become stricter. Higher pass rates were observed when no threshold was applied and they tended to decrease for 10% and 20% thresholds.
Conclusion: The results confirm the suitability of the equipments used and the validity of the plans. The study also confirmed that the threshold settings greatly affect the gamma pass rates, especially for lower gamma criteria.
Collapse
Affiliation(s)
- Manthala Padannayil Noufal
- Department of Medical Physics and Radiotherapy , Baby Memorial Hospital , Calicut , India
- Department of Physics , Farook College , Calicut , India
- University of Calicut , Malapuram , Kerala, India
| | | | - Puzhakkal Niyas
- Department of Medical Physics and Radiotherapy , Baby Memorial Hospital , Calicut , India
- Department of Physics , Farook College , Calicut , India
- University of Calicut , Malapuram , Kerala, India
| | | |
Collapse
|
14
|
Cai B, Goddu SM, Yaddanapudi S, Caruthers D, Wen J, Noel C, Mutic S, Sun B. Normalize the response of EPID in pursuit of linear accelerator dosimetry standardization. J Appl Clin Med Phys 2017; 19:73-85. [PMID: 29125224 PMCID: PMC5768011 DOI: 10.1002/acm2.12222] [Citation(s) in RCA: 7] [Impact Index Per Article: 0.9] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/12/2017] [Revised: 08/06/2017] [Accepted: 09/28/2017] [Indexed: 12/18/2022] Open
Abstract
Normalize the response of electronic portal imaging device (EPID) is the first step toward an EPID‐based standardization of Linear Accelerator (linac) dosimetry quality assurance. In this study, we described an approach to generate two‐dimensional (2D) pixel sensitivity maps (PSM) for EPIDs response normalization utilizing an alternative beam and dark‐field (ABDF) image acquisition technique and large overlapping field irradiations. The automated image acquisition was performed by XML‐controlled machine operation and the PSM was generated based on a recursive calculation algorithm for Varian linacs equipped with aS1000 and aS1200 imager panels. Cross‐comparisons of normalized beam profiles and 1.5%/1.5 mm 1D Gamma analysis was adopted to quantify the improvement of beam profile matching before and after PSM corrections. PSMs were derived for both photon (6, 10, 15 MV) and electron (6, 20 MeV) beams via proposed method. The PSM‐corrected images reproduced a horn‐shaped profile for photon beams and a relative uniform profiles for electrons. For dosimetrically matched linacs equipped with aS1000 panels, PSM‐corrected images showed increased 1D‐Gamma passing rates for all energies, with an average 10.5% improvement for crossline and 37% for inline beam profiles. Similar improvements in the phantom study were observed with a maximum improvement of 32% for 15 MV and 22% for 20 MeV. The PSM value showed no significant change for all energies over a 3‐month period. In conclusion, the proposed approach correct EPID response for both aS1000 and aS1200 panels. This strategy enables the possibility to standardize linac dosimetry QA and to benchmark linac performance utilizing EPID as the common detector.
Collapse
Affiliation(s)
- Bin Cai
- Department of Radiation Oncology, Washington University, St. Louis, MO, USA
| | - S Murty Goddu
- Department of Radiation Oncology, Washington University, St. Louis, MO, USA
| | - Sridhar Yaddanapudi
- Department of Radiation Oncology, University of Iowa Hospitals and Clinics, Iowa City, IA, USA
| | - Douglas Caruthers
- Department of Radiation Oncology, Washington University, St. Louis, MO, USA
| | - Jie Wen
- Mallinckrodt Institute of Radiology, Washington University School of Medicine, St. Louis, MO, USA
| | | | - Sasa Mutic
- Department of Radiation Oncology, Washington University, St. Louis, MO, USA
| | - Baozhou Sun
- Department of Radiation Oncology, Washington University, St. Louis, MO, USA
| |
Collapse
|
15
|
Sibolt P, Andersen CE, Ottosson W, Behrens CF. Time-resolved plastic scintillator dosimetry in a dynamic thorax phantom. RADIAT MEAS 2017. [DOI: 10.1016/j.radmeas.2017.04.016] [Citation(s) in RCA: 6] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/25/2022]
|
16
|
Boutry C, Sors A, Fontaine J, Delaby N, Delpon G. Technical Note: A simple algorithm to convert EPID gray values into absorbed dose to water without prior knowledge. Med Phys 2017; 44:6647-6653. [PMID: 28921931 DOI: 10.1002/mp.12587] [Citation(s) in RCA: 5] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/09/2017] [Revised: 07/12/2017] [Accepted: 09/08/2017] [Indexed: 12/28/2022] Open
Abstract
PURPOSE Full integration of EPID-based dosimetry in a global quality control workflow is still complicated. All the actual solutions are based on a relation between image gray-level signal and total linac-delivered dose. In this study, we propose a simple algorithm relying pixel gray-level of EPID image with average linac delivered dose per acquisition frame. METHODS Calibration models are constructed for Varian and Elekta linacs including scattering conditions and EPID-arm backscatter-specific corrections. Only simple homogeneous fields are required to establish the EPID dose conversion model for each x-ray beam. Then, the model was evaluated by comparing calculated and converted dose distributions for homogeneous and modulated beams using gamma maps. RESULTS To fit average dose per frame (Dfnorm ) vs pixel gray value (Ngnorm ) of each EPID image, a logarithmic curve Dfnorm=A+B∗lnNgnorm-C, has been chosen where A, B and C are constants depending on beam energy. Gamma comparison (2%, 2 mm, threshold 15%) between converted images and calculated dose distributions for linac control and pretreatment patient fields led to a gamma pass rate higher than 97% for all the analyzed fields. CONCLUSIONS Without a prior irradiation settings knowledge except the incident energy beam, we use EPID as a reliable dose to water detector for both homogeneous and modulated beams.
Collapse
Affiliation(s)
- Christine Boutry
- Department of Radiotherapy, Oncorad Garonne, 330 Avenue Marcel UNAL, 82000, Montauban, France
| | - Aurélie Sors
- Department of Radiotherapy, Oncorad Garonne, 330 Avenue Marcel UNAL, 82000, Montauban, France
| | - Jimmy Fontaine
- Department of Medical Physics, Institut de Cancérologie de l'Ouest Centre René Gauducheau, 44805, Saint-Herblain, France
| | - Nolwenn Delaby
- Department of Medical Physics, Centre Eugène Marquis, 35000, Rennes, France
| | - Gregory Delpon
- Department of Medical Physics, Institut de Cancérologie de l'Ouest Centre René Gauducheau, 44805, Saint-Herblain, France.,INSERM UMR 1232, IRS UN, 8 quai Moncousu, 44007, Nantes, France
| |
Collapse
|
17
|
Fuangrod T, Greer PB, Zwan BJ, Barnes MP, Lehmann J. A novel and independent method for time-resolved gantry angle quality assurance for VMAT. J Appl Clin Med Phys 2017; 18:134-142. [PMID: 28703451 PMCID: PMC5874941 DOI: 10.1002/acm2.12129] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/10/2017] [Revised: 05/21/2017] [Accepted: 06/05/2017] [Indexed: 11/30/2022] Open
Abstract
Volumetric‐modulated arc therapy (VMAT) treatment delivery requires three key dynamic components; gantry rotation, dose rate modulation, and multi‐leaf collimator motion, which are all simultaneously varied during the delivery. Misalignment of the gantry angle can potentially affect clinical outcome due to the steep dose gradients and complex MLC shapes involved. It is essential to develop independent gantry angle quality assurance (QA) appropriate to VMAT that can be performed simultaneously with other key VMAT QA testing. In this work, a simple and inexpensive fully independent gantry angle measurement methodology was developed that allows quantitation of the gantry angle accuracy as a function of time. This method is based on the analysis of video footage of a “Double dot” pattern attached to the front cover of the linear accelerator that consists of red and green circles printed on A4 paper sheet. A standard mobile phone is placed on the couch to record the video footage during gantry rotation. The video file is subsequently analyzed and used to determine the gantry angle from each video frame using the relative position of the two dots. There were two types of validation tests performed including the static mode with manual gantry angle rotation and dynamic mode with three complex test plans. The accuracy was 0.26° ± 0.04° and 0.46° ± 0.31° (mean ± 1 SD) for the static and dynamic modes, respectively. This method is user friendly, cost effective, easy to setup, has high temporal resolution, and can be combined with existing time‐resolved method for QA of MLC and dose rate to form a comprehensive set of procedures for time‐resolved QA of VMAT delivery system.
Collapse
Affiliation(s)
- Todsaporn Fuangrod
- Department of Radiation Oncology, Calvary Mater Newcastle Hospital, Newcastle, NSW, Australia
| | - Peter B Greer
- Department of Radiation Oncology, Calvary Mater Newcastle Hospital, Newcastle, NSW, Australia.,School of Mathematical and Physics Sciences, Faculty of Science and IT, the University of Newcastle, Newcastle, NSW, Australia
| | - Benjamin J Zwan
- School of Mathematical and Physics Sciences, Faculty of Science and IT, the University of Newcastle, Newcastle, NSW, Australia.,Central Coast Cancer Centre, Gosford Hospital, Gosford, NSW, Australia
| | - Michael P Barnes
- Department of Radiation Oncology, Calvary Mater Newcastle Hospital, Newcastle, NSW, Australia.,School of Mathematical and Physics Sciences, Faculty of Science and IT, the University of Newcastle, Newcastle, NSW, Australia.,School of Medical Radiation Sciences, University of Newcastle, Newcastle, NSW, Australia
| | - Joerg Lehmann
- Department of Radiation Oncology, Calvary Mater Newcastle Hospital, Newcastle, NSW, Australia.,School of Mathematical and Physics Sciences, Faculty of Science and IT, the University of Newcastle, Newcastle, NSW, Australia
| |
Collapse
|
18
|
Retrospective analysis of portal dosimetry pre-treatment quality assurance of prostate volumetric-modulated arc therapy (VMAT) plans. JOURNAL OF RADIOTHERAPY IN PRACTICE 2017. [DOI: 10.1017/s1460396917000486] [Citation(s) in RCA: 5] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022]
Abstract
AbstractBackgroundElectronic portal imaging device (EPID) offers high-resolution digital image that can be compared with a predicted portal dose image. A very common method to quantitatively compare a measured and calculated dose distribution that is routinely used for quality assurance (QA) of volumetric-modulated arc therapy (VMAT) and intensity-modulated radiation therapy treatment plans is the evaluation of the gamma index. The purpose of this work was to evaluate the gamma passing rate (%GP), maximum gamma (γmax), average gamma (γave), maximum dose difference (DDmax) and the average dose difference (DDave) for various regions of interest using Varian’s implementation of three absolute dose gamma calculation techniques of improved, local, and combined improved and local.Methods and materialsWe analyzed 232 portal dose images from 100 prostate cancer patients’ VMAT plans obtained using the Varian EPID on TrueBeam Linacs.ResultsOur data show that the %GP, γmax and γave depend on the gamma calculation method and the acceptance criteria. Higher %GP values were obtained compared with both our current institutional action level and the American Association of Physicists in Medicine Task Group 119 recommendations.ConclusionsThe results of this study can be used to establish stricter action levels for pre-treatment QA of prostate VMAT plans. A stricter 3%/3 mm improved gamma criterion with a passing rate of 97% or the 2%/2 mm improved gamma criterion with a passing rate of 95% can be achieved without additional measurements or configurations.
Collapse
|
19
|
Zwan BJ, Barnes MP, Hindmarsh J, Lim SB, Lovelock DM, Fuangrod T, O'Connor DJ, Keall PJ, Greer PB. Commissioning and quality assurance for VMAT delivery systems: An efficient time-resolved system using real-time EPID imaging. Med Phys 2017; 44:3909-3922. [PMID: 28564208 DOI: 10.1002/mp.12387] [Citation(s) in RCA: 8] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/01/2017] [Revised: 05/22/2017] [Accepted: 05/22/2017] [Indexed: 11/07/2022] Open
Abstract
PURPOSE An ideal commissioning and quality assurance (QA) program for Volumetric Modulated Arc Therapy (VMAT) delivery systems should assess the performance of each individual dynamic component as a function of gantry angle. Procedures within such a program should also be time-efficient, independent of the delivery system and be sensitive to all types of errors. The purpose of this work is to develop a system for automated time-resolved commissioning and QA of VMAT control systems which meets these criteria. METHODS The procedures developed within this work rely solely on images obtained, using an electronic portal imaging device (EPID) without the presence of a phantom. During the delivery of specially designed VMAT test plans, EPID frames were acquired at 9.5 Hz, using a frame grabber. The set of test plans was developed to individually assess the performance of the dose delivery and multileaf collimator (MLC) control systems under varying levels of delivery complexities. An in-house software tool was developed to automatically extract features from the EPID images and evaluate the following characteristics as a function of gantry angle: dose delivery accuracy, dose rate constancy, beam profile constancy, gantry speed constancy, dynamic MLC positioning accuracy, MLC speed and acceleration constancy, and synchronization between gantry angle, MLC positioning and dose rate. Machine log files were also acquired during each delivery and subsequently compared to information extracted from EPID image frames. RESULTS The largest difference between measured and planned dose at any gantry angle was 0.8% which correlated with rapid changes in dose rate and gantry speed. For all other test plans, the dose delivered was within 0.25% of the planned dose for all gantry angles. Profile constancy was not found to vary with gantry angle for tests where gantry speed and dose rate were constant, however, for tests with varying dose rate and gantry speed, segments with lower dose rate and higher gantry speed exhibited less profile stability. MLC positional accuracy was not observed to be dependent on the degree of interdigitation. MLC speed was measured for each individual leaf and slower leaf speeds were shown to be compensated for by lower dose rates. The test procedures were found to be sensitive to 1 mm systematic MLC errors, 1 mm random MLC errors, 0.4 mm MLC gap errors and synchronization errors between the MLC, dose rate and gantry angle controls systems of 1°. In general, parameters measured by both EPID and log files agreed with the plan, however, a greater average departure from the plan was evidenced by the EPID measurements. CONCLUSION QA test plans and analysis methods have been developed to assess the performance of each dynamic component of VMAT deliveries individually and as a function of gantry angle. This methodology relies solely on time-resolved EPID imaging without the presence of a phantom and has been shown to be sensitive to a range of delivery errors. The procedures developed in this work are both comprehensive and time-efficient and can be used for streamlined commissioning and QA of VMAT delivery systems.
Collapse
Affiliation(s)
- Benjamin J Zwan
- Central Coast Cancer Centre, Gosford Hospital, Gosford, NSW, 2250, Australia
- School of Mathematical and Physical Sciences, University of Newcastle, Newcastle, NSW, 2308, Australia
| | - Michael P Barnes
- Department of Radiation Oncology, Calvary Mater Hospital, Newcastle, NSW, 2298, Australia
- School of Medical Radiation Sciences, University of Newcastle, Newcastle, NSW, 2308, Australia
| | - Jonathan Hindmarsh
- Central Coast Cancer Centre, Gosford Hospital, Gosford, NSW, 2250, Australia
| | - Seng B Lim
- Department of Medical Physics, Memorial Sloan-Kettering Cancer Centre, New York, NY, 10065, USA
| | - Dale M Lovelock
- Department of Medical Physics, Memorial Sloan-Kettering Cancer Centre, New York, NY, 10065, USA
| | - Todsaporn Fuangrod
- Department of Radiation Oncology, Calvary Mater Hospital, Newcastle, NSW, 2298, Australia
| | - Daryl J O'Connor
- School of Mathematical and Physical Sciences, University of Newcastle, Newcastle, NSW, 2308, Australia
| | - Paul J Keall
- Radiation Physics Laboratory, Sydney Medical School, University of Sydney, Sydney, NSW, 2006, Australia
| | - Peter B Greer
- School of Mathematical and Physical Sciences, University of Newcastle, Newcastle, NSW, 2308, Australia
- Department of Radiation Oncology, Calvary Mater Hospital, Newcastle, NSW, 2298, Australia
| |
Collapse
|
20
|
Passarge M, Fix MK, Manser P, Stampanoni MFM, Siebers JV. A Swiss cheese error detection method for real-time EPID-based quality assurance and error prevention. Med Phys 2017; 44:1212-1223. [PMID: 28134989 DOI: 10.1002/mp.12142] [Citation(s) in RCA: 11] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/11/2016] [Revised: 01/15/2017] [Accepted: 01/16/2017] [Indexed: 01/17/2023] Open
Abstract
PURPOSE To develop a robust and efficient process that detects relevant dose errors (dose errors of ≥5%) in external beam radiation therapy and directly indicates the origin of the error. The process is illustrated in the context of electronic portal imaging device (EPID)-based angle-resolved volumetric-modulated arc therapy (VMAT) quality assurance (QA), particularly as would be implemented in a real-time monitoring program. METHODS A Swiss cheese error detection (SCED) method was created as a paradigm for a cine EPID-based during-treatment QA. For VMAT, the method compares a treatment plan-based reference set of EPID images with images acquired over each 2° gantry angle interval. The process utilizes a sequence of independent consecutively executed error detection tests: an aperture check that verifies in-field radiation delivery and ensures no out-of-field radiation; output normalization checks at two different stages; global image alignment check to examine if rotation, scaling, and translation are within tolerances; pixel intensity check containing the standard gamma evaluation (3%, 3 mm) and pixel intensity deviation checks including and excluding high dose gradient regions. Tolerances for each check were determined. To test the SCED method, 12 different types of errors were selected to modify the original plan. A series of angle-resolved predicted EPID images were artificially generated for each test case, resulting in a sequence of precalculated frames for each modified treatment plan. The SCED method was applied multiple times for each test case to assess the ability to detect introduced plan variations. To compare the performance of the SCED process with that of a standard gamma analysis, both error detection methods were applied to the generated test cases with realistic noise variations. RESULTS Averaged over ten test runs, 95.1% of all plan variations that resulted in relevant patient dose errors were detected within 2° and 100% within 14° (<4% of patient dose delivery). Including cases that led to slightly modified but clinically equivalent plans, 89.1% were detected by the SCED method within 2°. Based on the type of check that detected the error, determination of error sources was achieved. With noise ranging from no random noise to four times the established noise value, the averaged relevant dose error detection rate of the SCED method was between 94.0% and 95.8% and that of gamma between 82.8% and 89.8%. CONCLUSIONS An EPID-frame-based error detection process for VMAT deliveries was successfully designed and tested via simulations. The SCED method was inspected for robustness with realistic noise variations, demonstrating that it has the potential to detect a large majority of relevant dose errors. Compared to a typical (3%, 3 mm) gamma analysis, the SCED method produced a higher detection rate for all introduced dose errors, identified errors in an earlier stage, displayed a higher robustness to noise variations, and indicated the error source.
Collapse
Affiliation(s)
- Michelle Passarge
- Division of Medical Radiation Physics and Department of Radiation Oncology, Inselspital Bern University Hospital and University of Bern, Berne, 3010, Switzerland.,Department of Radiation Oncology, University of Virginia Health System, Charlottesville, 22908, Virginia, USA
| | - Michael K Fix
- Division of Medical Radiation Physics and Department of Radiation Oncology, Inselspital Bern University Hospital and University of Bern, Berne, 3010, Switzerland
| | - Peter Manser
- Division of Medical Radiation Physics and Department of Radiation Oncology, Inselspital Bern University Hospital and University of Bern, Berne, 3010, Switzerland
| | - Marco F M Stampanoni
- Institute for Biomedical Engineering, Swiss Federal Institute of Technology (ETH), Zurich, 8092, Switzerland.,Paul Scherrer Institute (PSI), Villigen, 5232, Switzerland
| | - Jeffrey V Siebers
- Department of Radiation Oncology, University of Virginia Health System, Charlottesville, 22908, Virginia, USA
| |
Collapse
|
21
|
Köhn J, Licher J, Mielke M, Loutfi-Krauss B, Blümer N, Heine B, Rödel C, Scherf C, Ramm U. Image movement of the Elekta EPID during gantry rotation: Effects on the verification of dose distributions. Phys Med 2017; 34:72-79. [DOI: 10.1016/j.ejmp.2017.01.017] [Citation(s) in RCA: 5] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 11/03/2016] [Revised: 12/31/2016] [Accepted: 01/26/2017] [Indexed: 11/29/2022] Open
|
22
|
McCowan PM, Asuni G, van Beek T, van Uytven E, Kujanpaa K, McCurdy BMC. A model-based 3D patient-specific pre-treatment QA method for VMAT using the EPID. Phys Med Biol 2017; 62:1600-1612. [PMID: 28079525 DOI: 10.1088/1361-6560/aa590a] [Citation(s) in RCA: 13] [Impact Index Per Article: 1.6] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/20/2023]
Abstract
This study reports the development and validation of a model-based, 3D patient dose reconstruction method for pre-treatment quality assurance using EPID images. The method is also investigated for sensitivity to potential MLC delivery errors. Each cine-mode EPID image acquired during plan delivery was processed using a previously developed back-projection dose reconstruction model providing a 3D dose estimate on the CT simulation data. Validation was carried out using 24 SBRT-VMAT patient plans by comparing: (1) ion chamber point dose measurements in a solid water phantom, (2) the treatment planning system (TPS) predicted 3D dose to the EPID reconstructed 3D dose in a solid water phantom, and (3) the TPS predicted 3D dose to the EPID and our forward predicted reconstructed 3D dose in the patient (CT data). AAA and AcurosXB were used for TPS predictions. Dose distributions were compared using 3%/3 mm (95% tolerance) and 2%/2 mm (90% tolerance) γ-tests in the planning target volume (PTV) and 20% dose volumes. The average percentage point dose differences between the ion chamber and the EPID, AcurosXB, and AAA were 0.73 ± 1.25%, 0.38 ± 0.96% and 1.06 ± 1.34% respectively. For the patient (CT) dose comparisons, seven (3%/3 mm) and nine (2%/2 mm) plans failed the EPID versus AAA. All plans passed the EPID versus Acuros XB and the EPID versus forward model γ-comparisons. Four types of MLC sensitive errors (opening, shifting, stuck, and retracting), of varying magnitude (0.2, 0.5, 1.0, 2.0 mm), were introduced into six different SBRT-VMAT plans. γ-comparisons of the erroneous EPID dose and original predicted dose were carried out using the same criteria as above. For all plans, the sensitivity testing using a 3%/3 mm γ-test in the PTV successfully determined MLC errors on the order of 1.0 mm, except for the single leaf retraction-type error. A 2%/2 mm criteria produced similar results with two more additional detected errors.
Collapse
Affiliation(s)
- P M McCowan
- Department of Physics and Astronomy, University of Manitoba, Winnipeg, Manitoba, R3T 2N2, Canada. Medical Physics Department, CancerCare Manitoba, 675 McDermot Ave., Winnipeg, Manitoba, R3E 0V9, Canada
| | | | | | | | | | | |
Collapse
|
23
|
Zwan BJ, Barnes MP, Fuangord T, Stanton CJ, O'Connor DJ, Keall PJ, Greer PB. An EPID-based system for gantry-resolved MLC quality assurance for VMAT. J Appl Clin Med Phys 2016; 17:348-365. [PMID: 27685132 PMCID: PMC5874117 DOI: 10.1120/jacmp.v17i5.6312] [Citation(s) in RCA: 10] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/31/2016] [Revised: 06/19/2016] [Accepted: 05/13/2016] [Indexed: 11/23/2022] Open
Abstract
Multileaf collimator (MLC) positions should be precisely and independently mea-sured as a function of gantry angle as part of a comprehensive quality assurance (QA) program for volumetric-modulated arc therapy (VMAT). It is also ideal that such a QA program has the ability to relate MLC positional accuracy to patient-specific dosimetry in order to determine the clinical significance of any detected MLC errors. In this work we propose a method to verify individual MLC trajectories during VMAT deliveries for use as a routine linear accelerator QA tool. We also extend this method to reconstruct the 3D patient dose in the treatment planning sys-tem based on the measured MLC trajectories and the original DICOM plan file. The method relies on extracting MLC positions from EPID images acquired at 8.41fps during clinical VMAT deliveries. A gantry angle is automatically tagged to each image in order to obtain the MLC trajectories as a function of gantry angle. This analysis was performed for six clinical VMAT plans acquired at monthly intervals for three months. The measured trajectories for each delivery were compared to the MLC positions from the DICOM plan file. The maximum mean error detected was 0.07 mm and a maximum root-mean-square error was 0.8 mm for any leaf of any delivery. The sensitivity of this system was characterized by introducing random and systematic MLC errors into the test plans. It was demonstrated that the system is capable of detecting random and systematic errors on the range of 1-2mm and single leaf calibration errors of 0.5 mm. The methodology developed in the work has potential to be used for efficient routine linear accelerator MLC QA and pretreatment patient-specific QA and has the ability to relate measured MLC positional errors to 3D dosimetric errors within a patient volume.
Collapse
|
24
|
Katsuta Y, Kadoya N, Fujita Y, Shimizu E, Matsunaga K, Matsushita H, Majima K, Jingu K. Quantification of residual dose estimation error on log file-based patient dose calculation. Phys Med 2016; 32:701-5. [PMID: 27162084 DOI: 10.1016/j.ejmp.2016.04.015] [Citation(s) in RCA: 21] [Impact Index Per Article: 2.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 01/06/2016] [Revised: 04/02/2016] [Accepted: 04/29/2016] [Indexed: 11/30/2022] Open
Abstract
PURPOSE The log file-based patient dose estimation includes a residual dose estimation error caused by leaf miscalibration, which cannot be reflected on the estimated dose. The purpose of this study is to determine this residual dose estimation error. METHODS AND MATERIALS Modified log files for seven head-and-neck and prostate volumetric modulated arc therapy (VMAT) plans simulating leaf miscalibration were generated by shifting both leaf banks (systematic leaf gap errors: ±2.0, ±1.0, and ±0.5mm in opposite directions and systematic leaf shifts: ±1.0mm in the same direction) using MATLAB-based (MathWorks, Natick, MA) in-house software. The generated modified and non-modified log files were imported back into the treatment planning system and recalculated. Subsequently, the generalized equivalent uniform dose (gEUD) was quantified for the definition of the planning target volume (PTV) and organs at risks. RESULTS For MLC leaves calibrated within ±0.5mm, the quantified residual dose estimation errors that obtained from the slope of the linear regression of gEUD changes between non- and modified log file doses per leaf gap are in head-and-neck plans 1.32±0.27% and 0.82±0.17Gy for PTV and spinal cord, respectively, and in prostate plans 1.22±0.36%, 0.95±0.14Gy, and 0.45±0.08Gy for PTV, rectum, and bladder, respectively. CONCLUSIONS In this work, we determine the residual dose estimation errors for VMAT delivery using the log file-based patient dose calculation according to the MLC calibration accuracy.
Collapse
Affiliation(s)
- Yoshiyuki Katsuta
- Department of Radiology, Takeda General Hospital, Aizuwakamatsu, Japan; Department of Radiation Oncology, Tohoku University Graduate School of Medicine, Sendai, Japan
| | - Noriyuki Kadoya
- Department of Radiation Oncology, Tohoku University Graduate School of Medicine, Sendai, Japan.
| | - Yukio Fujita
- Department of Radiation Oncology, Tokai University Graduate School of Medicine, Isehara, Japan
| | - Eiji Shimizu
- Department of Radiology, Takeda General Hospital, Aizuwakamatsu, Japan
| | - Kenichi Matsunaga
- Department of Radiology, Takeda General Hospital, Aizuwakamatsu, Japan
| | - Haruo Matsushita
- Department of Radiation Oncology, Tohoku University Graduate School of Medicine, Sendai, Japan
| | - Kazuhiro Majima
- Department of Radiology, Takeda General Hospital, Aizuwakamatsu, Japan
| | - Keiichi Jingu
- Department of Radiation Oncology, Tohoku University Graduate School of Medicine, Sendai, Japan
| |
Collapse
|
25
|
Ding A, Xing L, Han B. Development of an accurate EPID-based output measurement and dosimetric verification tool for electron beam therapy. Med Phys 2016; 42:4190-8. [PMID: 26133618 DOI: 10.1118/1.4922400] [Citation(s) in RCA: 7] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/13/2022] Open
Abstract
PURPOSE To develop an efficient and robust tool for output measurement and absolute dose verification of electron beam therapy by using a high spatial-resolution and high frame-rate amorphous silicon flat panel electronic portal imaging device (EPID). METHODS The dosimetric characteristics of the EPID, including saturation, linearity, and ghosting effect, were first investigated on a Varian Clinac 21EX accelerator. The response kernels of the individual pixels of the EPID to all available electron energies (6, 9, 12, 16, and 20 MeV) were calculated by using Monte Carlo (MC) simulations, which formed the basis to deconvolve an EPID raw images to the incident electron fluence map. The two-dimensional (2D) dose distribution at reference depths in water was obtained by using the constructed fluence map with a MC simulated pencil beam kernel with consideration of the geometric and structural information of the EPID. Output factor measurements were carried out with the EPID at a nominal source-surface distance of 100 cm for 2 × 2, 3 × 3, 6 × 6, 10 × 10, and 15 × 15 cm(2) fields for all available electron energies, and the results were compared with that measured in a solid water phantom using film and a Farmer-type ion chamber. The dose distributions at a reference depth specific to each energy and the flatness and symmetry of the 10 × 10 cm(2) electron beam were also measured using EPID, and the results were compared with ion chamber array and water scan measurements. Finally, three patient cases with various field sizes and irregular cutout shapes were also investigated. RESULTS EPID-measured dose changed linearly with the monitor units and showed little ghosting effect for dose rate up to 600 MU/min. The flatness and symmetry measured with the EPID were found to be consistent with ion chamber array and water scan measurements. The EPID-measured output factors for standard square fields of 2 × 2, 3 × 3, 6 × 6, 10 × 10, 15 × 15 cm(2) agreed with film and ion chamber measurements. The average discrepancy between EPID and ion chamber/film measurements was 0.81% ± 0.60% (SD) and 1.34% ± 0.75%, respectively. For the three clinical cases, the difference in output between the EPID- and ion chamber array measured values was found to be 1.13% ± 0.11%, 0.54% ± 0.10%, and 0.74% ± 0.11%, respectively. Furthermore, the γ-index analysis showed an excellent agreement between the EPID- and ion chamber array measured dose distributions: 100% of the pixels passed the criteria of 3%/3 mm. When the γ-index was set to be 2%/2 mm, the pass rate was found to be 99.0% ± 0.07%, 98.2% ± 0.14%, and 100% for the three cases. CONCLUSIONS The EPID dosimetry system developed in this work provides an accurate and reliable tool for routine output measurement and dosimetric verification of electron beam therapy. Coupled with its portability and ease of use, the proposed system promises to replace the current film-based approach for fast and reliable assessment of small and irregular electron field dosimetry.
Collapse
Affiliation(s)
- Aiping Ding
- Department of Radiation Oncology, Stanford University School of Medicine, Stanford, Calilfornia 94305
| | - Lei Xing
- Department of Radiation Oncology, Stanford University School of Medicine, Stanford, Calilfornia 94305
| | - Bin Han
- Department of Radiation Oncology, Stanford University School of Medicine, Stanford, Calilfornia 94305
| |
Collapse
|
26
|
Xu Z, Wang IZ, Kumaraswamy LK, Podgorsak MB. Evaluation of dosimetric effect caused by slowing with multi-leaf collimator (MLC) leaves for volumetric modulated arc therapy (VMAT). Radiol Oncol 2016; 50:121-8. [PMID: 27069458 PMCID: PMC4825334 DOI: 10.1515/raon-2016-0008] [Citation(s) in RCA: 7] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/16/2015] [Accepted: 10/19/2015] [Indexed: 01/10/2023] Open
Abstract
BACKGROUND This study is to report 1) the sensitivity of intensity modulated radiation therapy (IMRT) QA method for clinical volumetric modulated arc therapy (VMAT) plans with multi-leaf collimator (MLC) leaf errors that will not trigger MLC interlock during beam delivery; 2) the effect of non-beam-hold MLC leaf errors on the quality of VMAT plan dose delivery. MATERIALS AND METHODS Eleven VMAT plans were selected and modified using an in-house developed software. For each control point of a VMAT arc, MLC leaves with the highest speed (1.87-1.95 cm/s) were set to move at the maximal allowable speed (2.3 cm/s), which resulted in a leaf position difference of less than 2 mm. The modified plans were considered as 'standard' plans, and the original plans were treated as the 'slowing MLC' plans for simulating 'standard' plans with leaves moving at relatively lower speed. The measurement of each 'slowing MLC' plan using MapCHECK®2 was compared with calculated planar dose of the 'standard' plan with respect to absolute dose Van Dyk distance-to-agreement (DTA) comparisons using 3%/3 mm and 2%/2 mm criteria. RESULTS All 'slowing MLC' plans passed the 90% pass rate threshold using 3%/3 mm criteria while one brain and three anal VMAT cases were below 90% with 2%/2 mm criteria. For ten out of eleven cases, DVH comparisons between 'standard' and 'slowing MLC' plans demonstrated minimal dosimetric changes in targets and organs-at-risk. CONCLUSIONS For highly modulated VMAT plans, pass rate threshold (90%) using 3%/3mm criteria is not sensitive in detecting MLC leaf errors that will not trigger the MLC leaf interlock. However, the consequential effects of non-beam hold MLC errors on target and OAR doses are negligible, which supports the reliability of current patient-specific IMRT quality assurance (QA) method for VMAT plans.
Collapse
Affiliation(s)
| | | | - Lalith K Kumaraswamy
- Department of Radiation Medicine, Roswell Park Cancer Institute, Buffalo, NY 14263
| | | |
Collapse
|
27
|
Fenoglietto P, Khodri M, Nguyen D, Josserand-Pietri F, Aillères N. Twin machines validation for VMAT treatments using electronic portal-imaging device: a multicenter study. Radiat Oncol 2016; 11:2. [PMID: 26762179 PMCID: PMC4712460 DOI: 10.1186/s13014-015-0577-3] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/20/2015] [Accepted: 12/22/2015] [Indexed: 11/26/2022] Open
Abstract
Purpose To verify the accuracy of volumetric arc therapy (VMAT) using the RapidArc™ device when switching patients from one single linear accelerator (linac) to a paired energy and mechanics "twin" linac without reoptimization of the original treatment plan. Patients and Methods Four centers using 8 linacs were involved in this study. Seventy-four patients previously treated with the 6MV photon RapidArc™ technique were selected for analysis, using 242 measurements. In each institution, all patients were planned on linac A, and their plans were verified both on linac A and on the twin linac B. Verifications were done using the amorphous silicium electronic portal imager (EPID) of the linacs and were analyzed with the EpiQa software (Epidos, Bratislavia, Slovakia). The gamma index formalism was used for validation with a double threshold of 3 % and 3 mm with a measurement resolution of 0.39 mm/pixel, and a smoothed resolution of approximately 2.5 mm. Results The number of points passing the gamma criteria between the measured and computed doses was 94.79 ± 2.57 % for linac A and 94.61 ± 2.46 % for linac B. Concerning the smoothed measurement analysis, 98.67 ± 1.26 % and 98.59 ± 1.20 % points passing the threshold were obtained for linacs A and B, respectively. The difference between the 2 dose matrices acquired on the EPID was very small, with 99.92 ± 0.06 % of the points passing the criteria. Conclusion For linacs sharing the same mechanical and energy parameters, this study tends to indicate that patients may be safely switched from treatment with one linac to treatment with its twin linac using the same VMAT plan.
Collapse
Affiliation(s)
- P Fenoglietto
- Département d'Oncologie Radiothérapie, Institut régional du Cancer de Montpellier (ICM), 208 rue des Apothicaires, F-34298, Montpellier, Cedex 5, France.
| | - M Khodri
- Département d'Oncologie Radiothérapie, Institut de Cancérologie de la Loire (ICL), Saint Etienne, France.
| | - D Nguyen
- Service de radiothérapie, Groupe ORLAM, Mâcon, France.
| | | | - N Aillères
- Département d'Oncologie Radiothérapie, Institut régional du Cancer de Montpellier (ICM), 208 rue des Apothicaires, F-34298, Montpellier, Cedex 5, France.
| |
Collapse
|
28
|
Van Uytven E, Van Beek T, McCowan PM, Chytyk-Praznik K, Greer PB, McCurdy BMC. Validation of a method for in vivo
3D dose reconstruction for IMRT and VMAT treatments using on-treatment EPID images and a model-based forward-calculation algorithm. Med Phys 2015; 42:6945-54. [DOI: 10.1118/1.4935199] [Citation(s) in RCA: 51] [Impact Index Per Article: 5.1] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/03/2023] Open
|
29
|
Song JH, Kim MJ, Park SH, Lee SR, Lee MY, Lee DS, Suh TS. Gamma analysis dependence on specified low-dose thresholds for VMAT QA. J Appl Clin Med Phys 2015; 16:263-272. [PMID: 26699582 PMCID: PMC5691030 DOI: 10.1120/jacmp.v16i6.5696] [Citation(s) in RCA: 13] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/12/2015] [Revised: 07/20/2015] [Accepted: 07/07/2015] [Indexed: 11/23/2022] Open
Abstract
The American Association of Physicists in Medicine Task Group 119 instructed institutions to use a low-dose threshold of 10% or a region of interest determined by the jaw setting when they collected gamma analysis quality assurance (QA) data for the planar dose distribution. However, there are no clinical data to quantitatively demonstrate the impact of the low-dose threshold on the gamma index. Therefore, we performed a gamma analysis with various low-dose thresholds in the range of 0% to 15% according to both global and local normalization and different acceptance criteria (3%/3 mm, 2%/2 mm, and 1%/1 mm). A total of 30 treatment plans--10 head and neck, 10 brain, and 10 prostate cancer cases--were randomly selected from the Varian Eclipse treatment planning system (TPS). For the gamma analysis, a calculated portal image was acquired through a portal dose calculation algorithm in the Eclipse TPS, and a measured portal image was obtained using an electronic portal-imaging device. Then, the gamma analysis was performed using the Portal Dosimetry software (Varian Medical Systems, Palo Alto, CA). The gamma passing rate (%GP) for the global normalization decreased as the low-dose threshold increased, and all low-dose thresholds led to %GP values above 95% for both the 3%/3 mm and 2%/2 mm criteria. However, for the local normalization, %GP for a low-dose threshold of 10% was 7.47%, 10.23%, and 6.71% greater than the low-dose threshold of 0% for head and neck, brain, and prostate for the 3%/3 mm criteria, respectively. The results indicate that applying the low-dose threshold to global normalization does not have a critical impact on patient-specific QA results. However, in the local normalization, the low-dose threshold level should be carefully selected because the excluded low-dose points could cause the average %GP to increase rapidly.
Collapse
|
30
|
Beierholm A, Behrens C, Andersen C. Studying the potential of point detectors in time-resolved dose verification of dynamic radiotherapy. RADIAT MEAS 2015. [DOI: 10.1016/j.radmeas.2015.09.007] [Citation(s) in RCA: 5] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/28/2022]
|
31
|
Rowshanfarzad P, Riis HL, Zimmermann SJ, Ebert MA. A comprehensive study of the mechanical performance of gantry, EPID and the MLC assembly in Elekta linacs during gantry rotation. Br J Radiol 2015; 88:20140581. [PMID: 25906294 DOI: 10.1259/bjr.20140581] [Citation(s) in RCA: 16] [Impact Index Per Article: 1.6] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/05/2022] Open
Abstract
OBJECTIVE In radiotherapy treatments, it is crucial to monitor the performance of linear accelerator (linac) components, including gantry, collimation system and electronic portal imaging device (EPID) during arc deliveries. In this study, a simple EPID-based measurement method is suggested in conjunction with an algorithm to investigate the stability of these systems at various gantry angles with the aim of evaluating machine-related errors in treatments. METHODS The EPID sag, gantry sag, changes in source-to-detector distance (SDD), EPID and collimator skewness, EPID tilt and the sag in leaf bank assembly owing to linac rotation were separately investigated by acquisition of 37 EPID images of a simple phantom with 5 ball bearings at various gantry angles. A fast and robust software package was developed for automated analysis of the image data. Nine Elekta AB (Stockholm, Sweden) linacs of different models and number of years in service were investigated. RESULTS The average EPID sag was within 2 mm for all tested linacs. Some machines showed >1-mm gantry sag. Changes in the SDD values were within 1.3 cm. EPID skewness and tilt values were <1° in all machines. The maximum sag in multileaf collimator leaf bank assemblies was around 1 mm. A meaningful correlation was found between the age of the linacs and their mechanical performance. Conclusions and Advances in knowledge: The method and software developed in this study provide a simple tool for effective investigation of the behaviour of Elekta linac components with gantry rotation. Such a comprehensive study has been performed for the first time on Elekta machines.
Collapse
Affiliation(s)
- P Rowshanfarzad
- 1 School of Physics, University of Western Australia, Crawley, WA, Australia
| | - H L Riis
- 2 Radiofysisk Laboratorium, Odense University Hospital, Odense, Denmark
| | - S J Zimmermann
- 2 Radiofysisk Laboratorium, Odense University Hospital, Odense, Denmark
| | - M A Ebert
- 1 School of Physics, University of Western Australia, Crawley, WA, Australia.,3 Department of Radiation Oncology, Sir Charles Gairdner Hospital, Nedlands, WA, Australia
| |
Collapse
|
32
|
Hobson MA, Davis SD. Comparison between an in-house 1D profile correction method and a 2D correction provided in Varian's PDPC Package for improving the accuracy of portal dosimetry images. J Appl Clin Med Phys 2015; 16:4973. [PMID: 26103173 PMCID: PMC5690095 DOI: 10.1120/jacmp.v16i2.4973] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/11/2014] [Revised: 12/10/2014] [Accepted: 12/07/2014] [Indexed: 11/23/2022] Open
Abstract
While commissioning Varian's Portal Dose Image Prediction (PDIP) algorithm for portal dosimetry, an asymmetric radial response in the portal imager due to backscatter from the support arm was observed. This asymmetric response led to differences on the order of 2%–3% for simple square fields (<20×20 cm2) when comparing the measured to predicted portal fluences. A separate problem was that discrepancies of up to 10% were seen in measured to predicted portal fluences at increasing off‐axis distance (>10 cm). We have modified suggested methods from the literature to provide a 1D correction for the off‐axis response problem which adjusts the diagonal profile used in the portal imager calibration. This inherently cannot fix the 2D problem since the PDIP algorithm assumes a radially symmetric response and will lead to some uncertainty in portal dosimetry results. Varian has recently released generic “2D correction” files with their Portal Dosimetry Pre‐configuration (PDPC) package, but no independent testing has been published. We present the comparison between QA results using the Varian correction method to results using our 1D profile correction method using the gamma passing rates with a 3%, 3 mm criterion. The average, minimum, and maximum gamma pass rates for nine fixed‐field IMRT fields at gantry 0° using our profile correction method were 98.1%, 93.7%, and 99.8%, respectively, while the results using the PDPC correction method were 98.4%, 93.1%, and 99.8%. For four RapidArc fields, the average, minimum, and maximum gamma pass rates using our correction method were 99.6%, 99.4%, and 99.9%, respectively, while the results using the PDPC correction method were 99.8%, 99.5%, and 99.9%. The average gamma pass rates for both correction methods are quite similar, but both show improvement over the uncorrected results. PACS numbers: 87.55.Qr, 87.55.N‐
Collapse
|
33
|
Fuangrod T, Rowshanfarzad P, Greer PB, Middleton RH. A cine-EPID based method for jaw detection and quality assurance for tracking jaw in IMRT/VMAT treatments. Phys Med 2015; 31:16-24. [DOI: 10.1016/j.ejmp.2014.11.002] [Citation(s) in RCA: 14] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 07/28/2014] [Revised: 10/30/2014] [Accepted: 11/01/2014] [Indexed: 10/24/2022] Open
|
34
|
Yip S, Rottmann J, Berbeco R. The impact of cine EPID image acquisition frame rate on markerless soft-tissue tracking. Med Phys 2015; 41:061702. [PMID: 24877797 DOI: 10.1118/1.4873322] [Citation(s) in RCA: 12] [Impact Index Per Article: 1.2] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/20/2022] Open
Abstract
PURPOSE Although reduction of the cine electronic portal imaging device (EPID) acquisition frame rate through multiple frame averaging may reduce hardware memory burden and decrease image noise, it can hinder the continuity of soft-tissue motion leading to poor autotracking results. The impact of motion blurring and image noise on the tracking performance was investigated. METHODS Phantom and patient images were acquired at a frame rate of 12.87 Hz with an amorphous silicon portal imager (AS1000, Varian Medical Systems, Palo Alto, CA). The maximum frame rate of 12.87 Hz is imposed by the EPID. Low frame rate images were obtained by continuous frame averaging. A previously validated tracking algorithm was employed for autotracking. The difference between the programmed and autotracked positions of a Las Vegas phantom moving in the superior-inferior direction defined the tracking error (δ). Motion blurring was assessed by measuring the area change of the circle with the greatest depth. Additionally, lung tumors on 1747 frames acquired at 11 field angles from four radiotherapy patients are manually and automatically tracked with varying frame averaging. δ was defined by the position difference of the two tracking methods. Image noise was defined as the standard deviation of the background intensity. Motion blurring and image noise are correlated with δ using Pearson correlation coefficient (R). RESULTS For both phantom and patient studies, the autotracking errors increased at frame rates lower than 4.29 Hz. Above 4.29 Hz, changes in errors were negligible withδ < 1.60 mm. Motion blurring and image noise were observed to increase and decrease with frame averaging, respectively. Motion blurring and tracking errors were significantly correlated for the phantom (R = 0.94) and patient studies (R = 0.72). Moderate to poor correlation was found between image noise and tracking error with R -0.58 and -0.19 for both studies, respectively. CONCLUSIONS Cine EPID image acquisition at the frame rate of at least 4.29 Hz is recommended. Motion blurring in the images with frame rates below 4.29 Hz can significantly reduce the accuracy of autotracking.
Collapse
Affiliation(s)
- Stephen Yip
- Department of Radiation Oncology, Brigham and Women's Hospital, Dana-Farber Cancer Institute and Harvard Medical School, Boston, Massachusetts 02115
| | - Joerg Rottmann
- Department of Radiation Oncology, Brigham and Women's Hospital, Dana-Farber Cancer Institute and Harvard Medical School, Boston, Massachusetts 02115
| | - Ross Berbeco
- Department of Radiation Oncology, Brigham and Women's Hospital, Dana-Farber Cancer Institute and Harvard Medical School, Boston, Massachusetts 02115
| |
Collapse
|
35
|
Rowshanfarzad P, McGarry CK, Barnes MP, Sabet M, Ebert MA. An EPID-based method for comprehensive verification of gantry, EPID and the MLC carriage positional accuracy in Varian linacs during arc treatments. Radiat Oncol 2014; 9:249. [PMID: 25424471 PMCID: PMC4252011 DOI: 10.1186/s13014-014-0249-8] [Citation(s) in RCA: 27] [Impact Index Per Article: 2.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/02/2014] [Accepted: 11/04/2014] [Indexed: 11/10/2022] Open
Abstract
Background In modern radiotherapy, it is crucial to monitor the performance of all linac components including gantry, collimation system and electronic portal imaging device (EPID) during arc deliveries. In this study, a simple EPID-based measurement method has been introduced in conjunction with an algorithm to investigate the stability of these systems during arc treatments with the aim of ensuring the accuracy of linac mechanical performance. Methods The Varian EPID sag, gantry sag, changes in source-to-detector distance (SDD), EPID and collimator skewness, EPID tilt, and the sag in MLC carriages as a result of linac rotation were separately investigated by acquisition of EPID images of a simple phantom comprised of 5 ball-bearings during arc delivery. A fast and robust software package was developed for automated analysis of image data. Twelve Varian linacs of different models were investigated. Results The average EPID sag was within 1 mm for all tested linacs. All machines showed less than 1 mm gantry sag. Changes in SDD values were within 1.7 mm except for three linacs of one centre which were within 9 mm. Values of EPID skewness and tilt were negligible in all tested linacs. The maximum sag in MLC leaf bank assemblies was around 1 mm. The EPID sag showed a considerable improvement in TrueBeam linacs. Conclusion The methodology and software developed in this study provide a simple tool for effective investigation of the behaviour of linac components with gantry rotation. It is reproducible and accurate and can be easily performed as a routine test in clinics.
Collapse
Affiliation(s)
- Pejman Rowshanfarzad
- School of Physics, University of Western Australia, Crawley, WA, 6009, Australia.
| | - Conor K McGarry
- Radiotherapy Physics, Northern Ireland Cancer Centre, Belfast Health and Social Care Trust, Belfast, UK. .,Centre for Cancer Research and Cell Biology, Queen's University Belfast, Belfast, UK.
| | - Michael P Barnes
- Department of Radiation Oncology, Calvary Mater Newcastle Hospital, Newcastle, NSW, 2310, Australia.
| | - Mahsheed Sabet
- Department of Radiation Oncology, Sir Charles Gairdner Hospital, Nedlands, WA, 6009, Australia.
| | - Martin A Ebert
- School of Physics, University of Western Australia, Crawley, WA, 6009, Australia. .,Department of Radiation Oncology, Sir Charles Gairdner Hospital, Nedlands, WA, 6009, Australia.
| |
Collapse
|
36
|
Podesta M, Persoon LCGG, Verhaegen F. A novel time dependent gamma evaluation function for dynamic 2D and 3D dose distributions. Phys Med Biol 2014; 59:5973-85. [DOI: 10.1088/0031-9155/59/20/5973] [Citation(s) in RCA: 15] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022]
|
37
|
Nelson CL, Mason BE, Robinson RC, Kisling KD, Kirsner SM. Commissioning results of an automated treatment planning verification system. J Appl Clin Med Phys 2014; 15:4838. [PMID: 25207567 PMCID: PMC5711088 DOI: 10.1120/jacmp.v15i5.4838] [Citation(s) in RCA: 22] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/02/2013] [Revised: 05/28/2014] [Accepted: 05/06/2014] [Indexed: 11/23/2022] Open
Abstract
A dose calculation verification system (VS) was acquired and commissioned as a second check on the treatment planning system (TPS). This system reads DICOM CT datasets, RT plans, RT structures, and RT dose from the TPS and automatically, using its own collapsed cone superposition/convolution algorithm, computes dose on the same CT dataset. The system was commissioned by extracting basic beam parameters for simple field geometries and dose verification for complex treatments. Percent depth doses (PDD) and profiles were extracted for field sizes using jaw settings 3 × 3 cm2 - 40 × 40 cm2 and compared to measured data, as well as our TPS model. Smaller fields of 1 × 1 cm2 and 2 × 2 cm2 generated using the multileaf collimator (MLC) were analyzed in the same fashion as the open fields. In addition, 40 patient plans consisting of both IMRT and VMAT were computed and the following comparisons were made: 1) TPS to the VS, 2) VS to measured data, and 3) TPS to measured data where measured data is both ion chamber (IC) and film measurements. Our results indicated for all field sizes using jaw settings PDD errors for the VS on average were less than 0.87%, 1.38%, and 1.07% for 6x, 15x, and 18x, respectively, relative to measured data. PDD errors for MLC field sizes were less than 2.28%, 1.02%, and 2.23% for 6x, 15x, and 18x, respectively. The infield profile analysis yielded results less than 0.58% for 6x, 0.61% for 15x, and 0.77% for 18x for the VS relative to measured data. Analysis of the penumbra region yields results ranging from 66.5% points, meeting the DTA criteria to 100% of the points for smaller field sizes for all energies. Analysis of profile data for field sizes generated using the MLC saw agreement with infield DTA analysis ranging from 68.8%-100% points passing the 1.5%/1.5 mm criteria. Results from the dose verification for IMRT and VMAT beams indicated that, on average, the ratio of TPS to IC and VS to IC measurements was 100.5 ± 1.9% and 100.4 ± 1.3%, respectively, while our TPS to VS was 100.1 ± 1.0%. When comparing the TPS and VS to film measurements, the average percentage pixels passing a 3%/3mm criteria based gamma analysis were 96.6 ± 4.2% and 97 ± 5.6%, respectively. When the VS was compared to the TPS, on average 98.1 ± 5.3% of pixels passed the gamma analysis. Based upon these preliminary results, the VS system should be able to calculate dose adequately as a verification tool of our TPS.
Collapse
|
38
|
Van Esch A, Basta K, Evrard M, Ghislain M, Sergent F, Huyskens DP. The Octavius1500 2D ion chamber array and its associated phantoms: Dosimetric characterization of a new prototype. Med Phys 2014; 41:091708. [DOI: 10.1118/1.4892178] [Citation(s) in RCA: 28] [Impact Index Per Article: 2.5] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
|
39
|
Podesta M, Nijsten SMJJG, Persoon LCGG, Scheib SG, Baltes C, Verhaegen F. Time dependent pre-treatment EPID dosimetry for standard and FFF VMAT. Phys Med Biol 2014; 59:4749-68. [PMID: 25088064 DOI: 10.1088/0031-9155/59/16/4749] [Citation(s) in RCA: 38] [Impact Index Per Article: 3.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/22/2022]
Abstract
Methods to calibrate Megavoltage electronic portal imaging devices (EPIDs) for dosimetry have been previously documented for dynamic treatments such as intensity modulated radiotherapy (IMRT) using flattened beams and typically using integrated fields. While these methods verify the accumulated field shape and dose, the dose rate and differential fields remain unverified. The aim of this work is to provide an accurate calibration model for time dependent pre-treatment dose verification using amorphous silicon (a-Si) EPIDs in volumetric modulated arc therapy (VMAT) for both flattened and flattening filter free (FFF) beams. A general calibration model was created using a Varian TrueBeam accelerator, equipped with an aS1000 EPID, for each photon spectrum 6 MV, 10 MV, 6 MV-FFF, 10 MV-FFF. As planned VMAT treatments use control points (CPs) for optimization, measured images are separated into corresponding time intervals for direct comparison with predictions. The accuracy of the calibration model was determined for a range of treatment conditions. Measured and predicted CP dose images were compared using a time dependent gamma evaluation using criteria (3%, 3 mm, 0.5 sec). Time dependent pre-treatment dose verification is possible without an additional measurement device or phantom, using the on-board EPID. Sufficient data is present in trajectory log files and EPID frame headers to reliably synchronize and resample portal images. For the VMAT plans tested, significantly more deviation is observed when analysed in a time dependent manner for FFF and non-FFF plans than when analysed using only the integrated field. We show EPID-based pre-treatment dose verification can be performed on a CP basis for VMAT plans. This model can measure pre-treatment doses for both flattened and unflattened beams in a time dependent manner which highlights deviations that are missed in integrated field verifications.
Collapse
Affiliation(s)
- Mark Podesta
- Department of Radiation Oncology (MAASTRO), Maastricht University Medical Center (MUMC), Maastricht, the Netherlands
| | | | | | | | | | | |
Collapse
|
40
|
Hayashi N, Malmin RL, Watanabe Y. Dosimetric verification for intensity-modulated arc therapy plans by use of 2D diode array, radiochromic film and radiosensitive polymer gel. JOURNAL OF RADIATION RESEARCH 2014; 55:541-552. [PMID: 24449714 PMCID: PMC4014162 DOI: 10.1093/jrr/rrt139] [Citation(s) in RCA: 10] [Impact Index Per Article: 0.9] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Figures] [Subscribe] [Scholar Register] [Received: 11/27/2012] [Revised: 05/30/2013] [Accepted: 11/01/2013] [Indexed: 06/03/2023]
Abstract
Several tools are used for the dosimetric verification of intensity-modulated arc therapy (IMAT) treatment delivery. However, limited information is available for composite on-line evaluation of these tools. The purpose of this study was to evaluate the dosimetric verification of IMAT treatment plans using a 2D diode array detector (2D array), radiochromic film (RCF) and radiosensitive polymer gel dosimeter (RPGD). The specific verification plans were created for IMAT for two prostate cancer patients by use of the clinical treatment plans. Accordingly, the IMAT deliveries were performed with the 2D array on a gantry-mounting device, RCF in a cylindrical acrylic phantom, and the RPGD in two cylindrical phantoms. After the irradiation, the planar dose distributions from the 2D array and the RCFs, and the 3D dose distributions from the RPGD measurements were compared with the calculated dose distributions using the gamma analysis method (3% dose difference and 3-mm distance-to-agreement criterion), dose-dependent dose difference diagrams, dose difference histograms, and isodose distributions. The gamma passing rates of 2D array, RCFs and RPGD for one patient were 99.5%, 96.5% and 93.7%, respectively; the corresponding values for the second patient were 97.5%, 92.6% and 92.9%. Mean percentage differences between the RPGD measured and calculated doses in 3D volumes containing PTVs were -0.29 ± 7.1% and 0.97 ± 7.6% for the two patients, respectively. In conclusion, IMAT prostate plans can be delivered with high accuracy, although the 3D measurements indicated less satisfactory agreement with the treatment plans, mainly due to the dosimetric inaccuracy in low-dose regions of the RPGD measurements.
Collapse
Affiliation(s)
- Naoki Hayashi
- Faculty of Radiological Technology, School of Health Sciences, Fujita Health University, 1-98, Dengakugakubo, Kutsukake-cho, Toyoake, Aichi, 470-1192, Japan
| | - Ryan L. Malmin
- Ridgeview Regional Radiation Oncology, 560 South Maple Street, Suite 10, Waconia, MN 55387, USA
| | - Yoichi Watanabe
- Department of Radiation Oncology, 420 Delaware St. SE, MMC494, Minneapolis, MN 55455, USA
| |
Collapse
|
41
|
Abdellatif A, Gaede S. Control point analysis comparison for 3 different treatment planning and delivery complexity levels using a commercial 3-dimensional diode array. Med Dosim 2014; 39:174-9. [PMID: 24480374 DOI: 10.1016/j.meddos.2013.12.005] [Citation(s) in RCA: 9] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/28/2013] [Revised: 11/20/2013] [Accepted: 12/04/2013] [Indexed: 11/27/2022]
Abstract
To investigate the use of "Control Point Analysis" (Sun Nuclear Corporation, Melbourne, FL) to analyze and compare delivered volumetric-modulated arc therapy (VMAT) plans for 3 different treatment planning complexity levels. A total of 30 patients were chosen and fully anonymized for the purpose of this study. Overall, 10 lung stereotactic body radiotherapy (SBRT), 10 head-and-neck (H&N), and 10 prostate VMAT plans were generated on Pinnacle(3) and delivered on a Varian linear accelerator (LINAC). The delivered dose was measured using ArcCHECK (Sun Nuclear Corporation, Melbourne, FL). Each plan was analyzed using "Sun Nuclear Corporation (SNC) Patient 6" and "Control Point Analysis." Gamma passing percentage was used to assess the differences between the measured and planned dose distributions and to assess the role of various control point binning combinations. Of the different sites considered, the prostate cases reported the highest gamma passing percentages calculated with "SNC Patient 6" (97.5% to 99.2% for the 3%, 3mm) and "Control Point Analysis" (95.4% to 98.3% for the 3%, 3mm). The mean percentage of passing control point sectors for the prostate cases increased from 51.8 ± 7.8% for individual control points to 70.6 ± 10.5% for 5 control points binned together to 87.8 ± 11.0% for 10 control points binned together (2%, 2-mm passing criteria). Overall, there was an increasing trend in the percentage of sectors passing gamma analysis with an increase in the number of control points binned together in a sector for both the gamma passing criteria (2%, 2mm and 3%, 3mm). Although many plans passed the clinical quality assurance criteria, plans involving the delivery of high Monitor Unit (MU)/control point (SBRT) and plans involving high degree of modulation (H&N) showed less delivery accuracy per control point compared with plans with low MU/control point and low degree of modulation (prostate).
Collapse
Affiliation(s)
- Ady Abdellatif
- Department of Physics and Engineering, London Regional Cancer Program, London, Ontario, Canada.
| | - Stewart Gaede
- Department of Physics and Engineering, London Regional Cancer Program, London, Ontario, Canada; Department of Oncology, Western University, London, Ontario, Canada; Department of Medical Biophysics, Western University, London, Ontario, Canada
| |
Collapse
|
42
|
McCowan PM, Rickey DW, Rowshanfarzad P, Greer PB, Ansbacher W, McCurdy BM. An investigation of gantry angle data accuracy for cine-mode EPID images acquired during arc IMRT. J Appl Clin Med Phys 2014; 15:4507. [PMID: 24423849 PMCID: PMC5711235 DOI: 10.1120/jacmp.v15i1.4507] [Citation(s) in RCA: 20] [Impact Index Per Article: 1.8] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/29/2013] [Revised: 09/28/2013] [Accepted: 08/29/2013] [Indexed: 12/03/2022] Open
Abstract
EPID images acquired in cine mode during arc therapy have inaccurate gantry angles recorded in their image headers. In this work, methods were developed to assess the accuracy of the gantry potentiometer for linear accelerators. As well, assessments of the accuracy of other, more accessible, sources of gantry angle information (i.e., treatment log files, analysis of EPID image headers) were investigated. The methods used in this study are generally applicable to any linear accelerator unit, and have been demonstrated here with Clinac/Trilogy systems. Gantry angle data were simultaneously acquired using three methods: i) a direct gantry potentiometer measurement, ii) an incremental rotary encoder, and iii) a custom‐made radiographic gantry‐angle phantom which produced unique wire intersections as a function of gantry angle. All methods were compared to gantry angle data from the EPID image header and the linac MLC DynaLog file. The encoder and gantry‐angle phantom were used to validate the accuracy of the linac's potentiometer. The EPID image header gantry angles and the DynaLog file gantry angles were compared to the potentiometer. The encoder and gantry‐angle phantom mean angle differences with the potentiometer were 0.13∘±0.14∘ and 0.10∘±0.30∘, respectively. The EPID image header angles analyzed in this study were within ±1∘ of the potentiometer angles only 35% of the time. In some cases, EPID image header gantry angles disagreed by as much as 3° with the potentiometer. A time delay in frame acquisition was determined using the continuous acquisition mode of the EPID. After correcting for this time delay, 75% of the header angles, on average, were within ±1∘ of the true gantry angle, compared to an average of only 35% without the correction. Applying a boxcar smoothing filter to the corrected gantry angles further improved the accuracy of the header‐derived gantry angles to within ±1∘ for almost all images (99.4%). An angle accuracy of 0.11∘±0.04∘ was determined using a point‐by‐point comparison of the gantry angle data in the MLC DynaLog file and the potentiometer data. These simple correction methods can be easily applied to individual treatment EPID images in order to more accurately define the gantry angle. PACS numbers: 87.53.Kn, 87.55.T‐, 87.56.bd, 87.59.‐e
Collapse
|
43
|
Jin H, Jesseph FB, Ahmad S. A Comparison Study of Volumetric Modulated Arc Therapy Quality Assurances Using Portal Dosimetry and MapCHECK 2. ACTA ACUST UNITED AC 2014. [DOI: 10.14316/pmp.2014.25.2.65] [Citation(s) in RCA: 14] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/18/2022]
Affiliation(s)
- Hosang Jin
- Department of Radiation Oncology, Stephenson Cancer Center, University of Oklahoma Health Sciences Center, Oklahoma, USA
| | - Fredrick B. Jesseph
- Department of Radiation Oncology, Stephenson Cancer Center, University of Oklahoma Health Sciences Center, Oklahoma, USA
| | - Salahuddin Ahmad
- Department of Radiation Oncology, Stephenson Cancer Center, University of Oklahoma Health Sciences Center, Oklahoma, USA
| |
Collapse
|
44
|
Fuangrod T, Woodruff HC, Rowshanfarzad P, O'Connor DJ, Middleton RH, Greer PB. An independent system for real-time dynamic multileaf collimation trajectory verification using EPID. Phys Med Biol 2013; 59:61-81. [PMID: 24334552 DOI: 10.1088/0031-9155/59/1/61] [Citation(s) in RCA: 13] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
A new tool has been developed to verify the trajectory of dynamic multileaf collimators (MLCs) used in advanced radiotherapy techniques using only the information provided by the electronic portal imaging devices (EPID) measured image frames. The prescribed leaf positions are resampled to a higher resolution in a pre-processing stage to improve the verification precision. Measured MLC positions are extracted from the EPID frames using a template matching method. A cosine similarity metric is then applied to synchronise measured and planned leaf positions for comparison. Three additional comparison functions were incorporated to ensure robust synchronisation. The MLC leaf trajectory error detection was simulated for both intensity modulated radiation therapy (IMRT) (prostate) and volumetric modulated arc therapy (VMAT) (head-and-neck) deliveries with anthropomorphic phantoms in the beam. The overall accuracy for MLC positions automatically extracted from EPID image frames was approximately 0.5 mm. The MLC leaf trajectory verification system can detect leaf position errors during IMRT and VMAT with a tolerance of 3.5 mm within 1 s.
Collapse
Affiliation(s)
- Todsaporn Fuangrod
- Faculty of Engineering and Built Environment, School of Electrical Engineering and Computer Science, the University of Newcastle, NSW 2308, Australia
| | | | | | | | | | | |
Collapse
|
45
|
Mooslechner M, Mitterlechner B, Weichenberger H, Huber S, Sedlmayer F, Deutschmann H. Analysis of a free-running synchronization artifact correction for MV-imaging with aSi:H flat panels. Med Phys 2013; 40:031906. [PMID: 23464321 DOI: 10.1118/1.4790564] [Citation(s) in RCA: 7] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/23/2023] Open
Abstract
PURPOSE Solid state flat panel electronic portal imaging devices (EPIDs) are widely used for megavolt (MV) photon imaging applications in radiotherapy. In addition to their original purpose in patient position verification, they are convenient to use in quality assurance and dosimetry to verify beam geometry and dose deposition or to perform linear accelerator (linac) calibration procedures. However, native image frames from amorphous silicon (aSi:H) detectors show a range of artifacts which have to be eliminated by proper correction algorithms. When a panel is operated in free-running frame acquisition mode, moving vertical stripes (periodic synchronization artifacts) are a disturbing feature in image frames. Especially for applications in volumetric intensity modulated arc therapy (VMAT) or motion tracking, the synchronization (sync) artifacts are the limiting factor for potential and accuracy since they become even worse at higher frame rates and at lower dose rates, i.e., linac pulse repetition frequencies (PRFs). METHODS The authors introduced a synchronization correction method which is based on a theoretical model describing the interferences of the panel's readout clocking with the linac's dose pulsing. Depending on the applied PRF, a certain number of dose pulses is captured per frame which is readout columnwise, sequentially. The interference of the PRF with the panel readout is responsible for the period and the different gray value levels of the sync stripes, which can be calculated analytically. Sync artifacts can then be eliminated multiplicatively in precorrected frames without additional information about radiation pulse timing. RESULTS For the analysis, three aSi:H EPIDs of various types were investigated with 6 and 15 MV photon beams at varying PRFs of 25, 50, 100, 200, and 400 pulses per second. Applying the sync correction at panels with gadolinium oxysulfide scintillators improved single frame flood field image quality drastically [improvement of the signal-to-noise ratio (SNR) up to 66.1 dB for 6 MV and 66.0 dB for 15 MV]. Also for the EPID with a caesium iodide scintillator, the noise for the lower PRFs could be reduced (SNR at 6 MV of up to 56.3 dB and at 15 MV up to 46.7 dB). However, the simplistic readout interference model fails at higher PRFs, where image lag and ghosting effects due to trapped charges in the thin film transistor and scintillator postglowing require additional corrections. CONCLUSIONS The presented free-running sync correction method improves SNR of single frames and enables imaging applications, like low-dose rate imaging at increased image frame rates (e.g., to track moving gold fiducials in the lung). Adaptive image guided radiotherapy protocols become even feasible in VMAT plans. Also simultaneous kilovolt and MV imaging applications can benefit from new possibilities of MV scatter removal in x-ray images.
Collapse
Affiliation(s)
- Michaela Mooslechner
- Institute for Research and Development on Advanced Radiation Technologies (radART), Paracelsus Medical University, Salzburg 5020, Austria.
| | | | | | | | | | | |
Collapse
|
46
|
Liu B, Adamson J, Rodrigues A, Zhou F, Yin FF, Wu Q. A novel technique for VMAT QA with EPID in cine mode on a Varian TrueBeam linac. Phys Med Biol 2013; 58:6683-700. [DOI: 10.1088/0031-9155/58/19/6683] [Citation(s) in RCA: 31] [Impact Index Per Article: 2.6] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
|
47
|
McGarry CK, O'Connell BF, Grattan MWD, Agnew CE, Irvine DM, Hounsell AR. Octavius 4D characterization for flattened and flattening filter free rotational deliveries. Med Phys 2013; 40:091707. [PMID: 24007140 DOI: 10.1118/1.4817482] [Citation(s) in RCA: 30] [Impact Index Per Article: 2.5] [Reference Citation Analysis] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Affiliation(s)
- Conor K McGarry
- Radiotherapy Physics, Northern Ireland Cancer Centre, Belfast Health and Social Care Trust, Belfast, BT9 7AB Northern Ireland.
| | | | | | | | | | | |
Collapse
|
48
|
Fuangrod T, Woodruff HC, van Uytven E, McCurdy BMC, Kuncic Z, O'Connor DJ, Greer PB. A system for EPID-based real-time treatment delivery verification during dynamic IMRT treatment. Med Phys 2013; 40:091907. [DOI: 10.1118/1.4817484] [Citation(s) in RCA: 32] [Impact Index Per Article: 2.7] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
|
49
|
Woodruff HC, Fuangrod T, Rowshanfarzad P, McCurdy BMC, Greer PB. Gantry-angle resolved VMAT pretreatment verification using EPID image prediction. Med Phys 2013; 40:081715. [DOI: 10.1118/1.4816384] [Citation(s) in RCA: 40] [Impact Index Per Article: 3.3] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
|
50
|
Tatsumi D, Nakada R, Ienaga A, Yomoda A, Inoue M, Ichida T, Hosono M. [Electronic portal image device dosimetry for volumetric modulated arc therapy]. Nihon Hoshasen Gijutsu Gakkai Zasshi 2013; 69:11-8. [PMID: 23358333 DOI: 10.6009/jjrt.2013_jsrt_69.1.11] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
Recently electronic portal image devices (EPIDs) have been widely used for quality assurance and dose verification. However there are no reports describing EPID dosimetry for Elekta volumetric modulated arc therapy (VMAT). We have investigated EPID dosimetry during VMAT delivery using a commercial software EPIDose with an Elekta Synergy linac. Dose rate dependence and the linac system sag during gantry rotation were measured. Gamma indices were calculated between measured doses using an EPID and calculation made by a treatment planning system for prostate VMAT test plans. The results were also compared to gamma indices using films and a two-dimensional detector array, MapCHECK2. The pass rates of the gamma analysis with a criterion of 3% and 2 mm for the three methods were over 96% with good consistency. Our results have showed that EPID dosimetry is feasible for Elekta VMAT.
Collapse
|