1
|
Tseng HW, Karellas A, Vedantham S. Cone-beam breast CT using an offset detector: effect of detector offset and image reconstruction algorithm. Phys Med Biol 2022; 67. [PMID: 35316793 PMCID: PMC9045275 DOI: 10.1088/1361-6560/ac5fe1] [Citation(s) in RCA: 3] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/08/2021] [Accepted: 03/22/2022] [Indexed: 11/12/2022]
Abstract
Objective.A dedicated cone-beam breast computed tomography (BCT) using a high-resolution, low-noise detector operating in offset-detector geometry has been developed. This study investigates the effects of varying detector offsets and image reconstruction algorithms to determine the appropriate combination of detector offset and reconstruction algorithm.Approach.Projection datasets (300 projections in 360°) of 30 breasts containing calcified lesions that were acquired using a prototype cone-beam BCT system comprising a 40 × 30 cm flat-panel detector with 1024 × 768 detector pixels were reconstructed using Feldkamp-Davis-Kress (FDK) algorithm and served as the reference. The projection datasets were retrospectively truncated to emulate cone-beam datasets with sinograms of 768×768 and 640×768 detector pixels, corresponding to 5 cm and 7.5 cm lateral offsets, respectively. These datasets were reconstructed using the FDK algorithm with appropriate weights and an ASD-POCS-based Fast, total variation-Regularized, Iterative, Statistical reconstruction Technique (FRIST), resulting in a total of 4 offset-detector reconstructions (2 detector offsets × 2 reconstruction methods). Signal difference-to-noise ratio (SDNR), variance, and full-width at half-maximum (FWHM) of calcifications in two orthogonal directions were determined from all reconstructions. All quantitative measurements were performed on images in units of linear attenuation coefficient (1/cm).Results.The FWHM of calcifications did not differ (P > 0.262) among reconstruction algorithms and detector formats, implying comparable spatial resolution. For a chosen detector offset, the FRIST algorithm outperformed FDK in terms of variance and SDNR (P < 0.0001). For a given reconstruction method, the 5 cm offset provided better results.Significance.This study indicates the feasibility of using the compressed sensing-based, FRIST algorithm to reconstruct sinograms from offset-detectors. Among the reconstruction methods and detector offsets studied, FRIST reconstructions corresponding to a 30 cm × 30 cm with 5 cm lateral offset, achieved the best performance. A clinical prototype using such an offset geometry has been developed and installed for clinical trials.
Collapse
Affiliation(s)
- Hsin Wu Tseng
- Department of Medical Imaging, The University of Arizona, Tucson, AZ, United States of America
| | - Andrew Karellas
- Department of Medical Imaging, The University of Arizona, Tucson, AZ, United States of America
| | - Srinivasan Vedantham
- Department of Medical Imaging, The University of Arizona, Tucson, AZ, United States of America.,Department of Biomedical Engineering, The University of Arizona, Tucson, AZ, United States of America
| |
Collapse
|
2
|
Brombal L, Arana Peña LM, Arfelli F, Longo R, Brun F, Contillo A, Di Lillo F, Tromba G, Di Trapani V, Donato S, Menk RH, Rigon L. Motion artifacts assessment and correction using optical tracking in synchrotron radiation breast CT. Med Phys 2021; 48:5343-5355. [PMID: 34252212 PMCID: PMC9291820 DOI: 10.1002/mp.15084] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/18/2021] [Revised: 05/12/2021] [Accepted: 06/21/2021] [Indexed: 11/23/2022] Open
Abstract
Purpose The SYRMA‐3D collaboration is setting up a breast computed tomography (bCT) clinical program at the Elettra synchrotron radiation facility in Trieste, Italy. Unlike the few dedicated scanners available at hospitals, synchrotron radiation bCT requires the patient's rotation, which in turn implies a long scan duration (from tens of seconds to few minutes). At the same time, it allows the achievement of high spatial resolution. These features make synchrotron radiation bCT prone to motion artifacts. This article aims at assessing and compensating for motion artifacts through an optical tracking approach. Methods In this study, patients’ movements due to breathing have been first assessed on seven volunteers and then simulated during the CT scans of a breast phantom and a surgical specimen, by adding a periodic oscillatory motion (constant speed, 1 mm amplitude, 12 cycles/minute). CT scans were carried out at 28 keV with a mean glandular dose of 5 mGy. Motion artifacts were evaluated and a correction algorithm based on the optical tracking of fiducial marks was introduced. A quantitative analysis based on the structural similarity (SSIM) index and the normalized mean square error (nMSE) was performed on the reconstructed CT images. Results CT images reconstructed through the optical tracking procedure were found to be as good as the motionless reference image. Moreover, the analysis of SSIM and nMSE demonstrated that an uncorrected motion of the order of the system's point spread function (around 0.1 mm in the present case) can be tolerated. Conclusions Results suggest that a motion correction procedure based on an optical tracking system would be beneficial in synchrotron radiation bCT.
Collapse
Affiliation(s)
- Luca Brombal
- Department of Physics, University of Trieste, Trieste, Italy.,Division of Trieste, Istituto Nazionale di Fisica Nucleare, Trieste, Italy
| | - Lucia Mariel Arana Peña
- Department of Physics, University of Trieste, Trieste, Italy.,Division of Trieste, Istituto Nazionale di Fisica Nucleare, Trieste, Italy
| | - Fulvia Arfelli
- Department of Physics, University of Trieste, Trieste, Italy.,Division of Trieste, Istituto Nazionale di Fisica Nucleare, Trieste, Italy
| | - Renata Longo
- Department of Physics, University of Trieste, Trieste, Italy.,Division of Trieste, Istituto Nazionale di Fisica Nucleare, Trieste, Italy
| | - Francesco Brun
- Division of Trieste, Istituto Nazionale di Fisica Nucleare, Trieste, Italy.,Department of Engineering and Architecture, University of Trieste, Trieste, Italy
| | | | | | | | - Vittorio Di Trapani
- Department of Physical sciences, Earth and environment, University of Siena, Siena, Italy.,Division of Pisa, Istituto Nazionale di Fisica Nucleare, Pisa, Italy
| | - Sandro Donato
- Department of Physics, University of Calabria, Arcavacata di Rende, Cosenza, Italy.,Division of Frascati, Istituto Nazionale di Fisca Nucleare, Frascati, Rome, Italy
| | - Ralf Hendrik Menk
- Division of Trieste, Istituto Nazionale di Fisica Nucleare, Trieste, Italy.,Elettra-Sincrotrone Trieste S.C.p.A., Trieste, Italy.,Department of Medical Imaging, University of Saskatchewan, Saskatoon, Canada
| | - Luigi Rigon
- Department of Physics, University of Trieste, Trieste, Italy.,Division of Trieste, Istituto Nazionale di Fisica Nucleare, Trieste, Italy
| |
Collapse
|
3
|
Tseng HW, Karellas A, Vedantham S. Radiation dosimetry of a clinical prototype dedicated cone-beam breast CT system with offset detector. Med Phys 2021; 48:1079-1088. [PMID: 33501686 DOI: 10.1002/mp.14688] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/30/2020] [Revised: 12/15/2020] [Accepted: 12/17/2020] [Indexed: 12/27/2022] Open
Abstract
PURPOSE A clinical-prototype, dedicated, cone-beam breast computed tomography (CBBCT) system with offset detector is undergoing clinical evaluation at our institution. This study is to estimate the normalized glandular dose coefficients ( DgN CT ) that provide air kerma-to-mean glandular dose conversion factors using Monte Carlo simulations. MATERIALS AND METHODS The clinical prototype CBBCT system uses 49 kV x-ray spectrum with 1.39 mm 1st half-value layer thickness. Monte Carlo simulations (GATE, version 8) were performed with semi-ellipsoidal, homogeneous breasts of various fibroglandular weight fractions ( f g = 0.01 , 0.15 , 0.5 , 1 ) , chest wall diameters ( d = 8 , 10 , 14 , 18 , 20 cm), and chest wall to nipple length ( l = 0.75 d ), aligned with the axis of rotation (AOR) located at 65 cm from the focal spot to determine the DgN CT . Three geometries were considered - 40 × 30 -cm detector with no offset that served as reference and corresponds to a clinical CBBCT system, 30 × 30 -cm detector with 5 cm offset, and a 30 × 30 -cm detector with 10 cm offset. RESULTS For 5 cm lateral offset, the DgN CT ranged 0.177 - 0.574 mGy/mGy and reduction in DgN CT with respect to reference geometry was observed only for 18 cm ( 6.4 % ± 0.23 % ) and 20 cm ( 9.6 % ± 0.22 % ) diameter breasts. For the 10 cm lateral offset, the DgN CT ranged 0.221 - 0.581 mGy/mGy and reduction in DgN CT was observed for all breast diameters. The reduction in DgN CT was 1.4 % ± 0.48 % , 7.1 % ± 0.13 % , 17.5 % ± 0.19 % , 25.1 % ± 0.15 % , and 27.7 % ± 0.08 % for 8, 10, 14, 18, and 20 cm diameter breasts, respectively. For a given breast diameter, the reduction in DgN CT with offset-detector geometries was not dependent on f g . Numerical fits of DgN CT d , l , f g were generated for each geometry. CONCLUSION The DgN CT and the numerical fit, D g N CT d , l , f g would be of benefit for current CBBCT systems using the reference geometry and for future generations using offset-detector geometry. There exists a potential for radiation dose reduction with offset-detector geometry, provided the same technique factors as the reference geometry are used, and the image quality is clinically acceptable.
Collapse
Affiliation(s)
- Hsin Wu Tseng
- Department of Medical Imaging, The University of Arizona, Tucson, AZ, USA
| | - Andrew Karellas
- Department of Medical Imaging, The University of Arizona, Tucson, AZ, USA
| | - Srinivasan Vedantham
- Department of Medical Imaging, The University of Arizona, Tucson, AZ, USA.,Department of Biomedical Engineering, The University of Arizona, Tucson, AZ, USA
| |
Collapse
|
4
|
Becker AE, Hernandez AM, Schwoebel PR, Boone JM. Cone beam CT multisource configurations: evaluating image quality, scatter, and dose using phantom imaging and Monte Carlo simulations. ACTA ACUST UNITED AC 2020; 65:235032. [DOI: 10.1088/1361-6560/abc306] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/12/2022]
|
5
|
Ghazi P. Reduction of scatter in breast CT yields improved microcalcification visibility. Phys Med Biol 2020; 65:235047. [PMID: 33274730 DOI: 10.1088/1361-6560/abae07] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
The inadequate visibility of microcalcifications-small calcium deposits that cue radiologists to early stages of cancer-is a major limitation in current designs of dedicated breast computed tomography (bCT). This limitation has previously been attributed to the constituent components, spatial resolution, and utilized dose. Scattered radiation has been considered an occurrence with low-frequency impacts that can be compensated for in post-processing. We hypothesized, however, that the acquisition of scattered radiation has a far more detrimental impact on clinically relevant features than has previously been understood. Critically, acquisition of scatter leads to the reduced visibility of microcalcifications. This hypothesis was investigated and supported via mathematical derivations and simulation studies. We conducted a series of comparative studies in which four bCT systems were simulated under iso-dose and iso-resolution conditions, characterizing the dependencies of microcalcification contrast on accumulated scatter. Included among the simulated systems is a novel bCT design-narrow beam bCT (NB-bCT)-that captures nearly zero scatter. We find that current bCT systems suffer from significant levels of scatter. As validated in theory, depending on the system and size of microcalcifications, between 25% and over 70% of contrast resolution is lost due to scatter. The results in NB-bCT, however, provide evidence that by removing scatter build-up in projections, the contrast of microcalcifications in a bCT image is preserved, regardless of their size or location in the breast.
Collapse
Affiliation(s)
- Peymon Ghazi
- Malcova LLC, 3000 Falls Rd Suite 400, Baltimore, MD 21211, United States of America
| |
Collapse
|
6
|
Aruleba K, Obaido G, Ogbuokiri B, Fadaka AO, Klein A, Adekiya TA, Aruleba RT. Applications of Computational Methods in Biomedical Breast Cancer Imaging Diagnostics: A Review. J Imaging 2020; 6:105. [PMID: 34460546 PMCID: PMC8321173 DOI: 10.3390/jimaging6100105] [Citation(s) in RCA: 11] [Impact Index Per Article: 2.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/11/2020] [Revised: 09/12/2020] [Accepted: 09/14/2020] [Indexed: 12/14/2022] Open
Abstract
With the exponential increase in new cases coupled with an increased mortality rate, cancer has ranked as the second most prevalent cause of death in the world. Early detection is paramount for suitable diagnosis and effective treatment of different kinds of cancers, but this is limited to the accuracy and sensitivity of available diagnostic imaging methods. Breast cancer is the most widely diagnosed cancer among women across the globe with a high percentage of total cancer deaths requiring an intensive, accurate, and sensitive imaging approach. Indeed, it is treatable when detected at an early stage. Hence, the use of state of the art computational approaches has been proposed as a potential alternative approach for the design and development of novel diagnostic imaging methods for breast cancer. Thus, this review provides a concise overview of past and present conventional diagnostics approaches in breast cancer detection. Further, we gave an account of several computational models (machine learning, deep learning, and robotics), which have been developed and can serve as alternative techniques for breast cancer diagnostics imaging. This review will be helpful to academia, medical practitioners, and others for further study in this area to improve the biomedical breast cancer imaging diagnosis.
Collapse
Affiliation(s)
- Kehinde Aruleba
- School of Computer Science and Applied Mathematics, University of the Witwatersrand, Johannesburg 2001, South Africa; (K.A.); (G.O.); (B.O.)
| | - George Obaido
- School of Computer Science and Applied Mathematics, University of the Witwatersrand, Johannesburg 2001, South Africa; (K.A.); (G.O.); (B.O.)
| | - Blessing Ogbuokiri
- School of Computer Science and Applied Mathematics, University of the Witwatersrand, Johannesburg 2001, South Africa; (K.A.); (G.O.); (B.O.)
| | - Adewale Oluwaseun Fadaka
- Department of Biotechnology, Faculty of Natural Sciences, University of the Western Cape, Private Bag X17, Bellville, Cape Town 7535, South Africa;
| | - Ashwil Klein
- Department of Biotechnology, Faculty of Natural Sciences, University of the Western Cape, Private Bag X17, Bellville, Cape Town 7535, South Africa;
| | - Tayo Alex Adekiya
- Department of Pharmacy and Pharmacology, School of Therapeutic Science, Faculty of Health Sciences, University of the Witwatersrand, Johannesburg, 7 York Road, Parktown 2193, South Africa;
| | - Raphael Taiwo Aruleba
- Department of Molecular and Cell Biology, Faculty of Science, University of Cape Town, Cape Town 7701, South Africa
| |
Collapse
|
7
|
Clinical Feasibility of Reduced Field-of-View Diffusion-Weighted Magnetic Resonance Imaging with Computed Diffusion-Weighted Imaging Technique in Breast Cancer Patients. Diagnostics (Basel) 2020; 10:diagnostics10080538. [PMID: 32751723 PMCID: PMC7460410 DOI: 10.3390/diagnostics10080538] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/19/2020] [Revised: 07/20/2020] [Accepted: 07/28/2020] [Indexed: 11/29/2022] Open
Abstract
Background: We evaluated the feasibility of the reduced field-of-view (rFOV) diffusion-weighted imaging (DWI) with computed DWI technique by comparison and analysis of the inter-method agreement among acquired rFOV DWI (rFOVA), rFOV DWI with computed DWI technique (rFOVS), and dynamic contrast-enhanced (DCE) magnetic resonance imaging (MRI) in patients with breast cancer. Methods: A total of 130 patients with biopsy-proven breast cancers who underwent breast MRI from April 2017 to December 2017 were included in this study. The rFOVS were reformatted by calculation of the apparent diffusion coefficient curve obtained from rFOVA b = 0 s/mm2 and b = 500 s/mm2. Visual assessment of the image quality of rFOVA b = 1000 s/mm2, rFOVS, and DCE MRI was performed using a four-point grading system. Morphologic analyses of the index cancer was performed on rFOVA, rFOVS, and DCE MRI. The signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and contrast of tumor-to-parenchyma (TPC) were calculated. Results: Image quality scores with rFOVA, rFOVS, and DCE MRI were not significantly different (p = 0.357). Lesion analysis of shape, margin, and size of the index cancer also did not show significant differences among the three sequences (p = 0.858, p = 0.242, and p = 0.858, respectively). SNR, CNR, and TPC of DCE MRI were significantly higher than those of rFOVA and rFOVS (p < 0.001, p = 0.001, and p = 0.016, respectively). Significant differences were not found between the SNR, CNR, and TPC of rFOVA and those of rFOVS (p > 0.999, p > 0.999, and p > 0.999, respectively). Conclusion: The rFOVA and rFOVS showed nearly equivalent levels of image quality required for morphological analysis of the tumors and for lesion conspicuity compared with DCE MRI.
Collapse
|
8
|
Ghazi P, Hernandez AM, Abbey C, Yang K, Boone JM. Shading artifact correction in breast CT using an interleaved deep learning segmentation and maximum-likelihood polynomial fitting approach. Med Phys 2019; 46:3414-3430. [PMID: 31102462 DOI: 10.1002/mp.13599] [Citation(s) in RCA: 10] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/06/2018] [Revised: 05/09/2019] [Accepted: 05/12/2019] [Indexed: 12/19/2022] Open
Abstract
PURPOSE The purpose of this work was twofold: (a) To provide a robust and accurate method for image segmentation of dedicated breast CT (bCT) volume data sets, and (b) to improve Hounsfield unit (HU) accuracy in bCT by means of a postprocessing method that uses the segmented images to correct for the low-frequency shading artifacts in reconstructed images. METHODS A sequential and iterative application of image segmentation and low-order polynomial fitting to bCT volume data sets was used in the interleaved correction (IC) method. Image segmentation was performed through a deep convolutional neural network (CNN) with a modified U-Net architecture. A total of 45 621 coronal bCT images from 111 patient volume data sets were segmented (using a previously published segmentation algorithm) and used for neural network training, validation, and testing. All patient data sets were selected from scans performed on four different prototype breast CT systems. The adipose voxels for each patient volume data set, segmented using the proposed CNN, were then fit to a three-dimensional low-order polynomial. The polynomial fit was subsequently used to correct for the shading artifacts introduced by scatter and beam hardening in a method termed "flat fielding." An interleaved utilization of image segmentation and flat fielding was repeated until a convergence criterion was satisfied. Mathematical and physical phantom studies were conducted to evaluate the dependence of the proposed algorithm on breast size and the distribution of fibroglandular tissue. In addition, a subset of patient scans (not used in the CNN training, testing or validation) were used to investigate the accuracy of the IC method across different scanner designs and beam qualities. RESULTS The IC method resulted in an accurate classification of different tissue types with an average Dice similarity coefficient > 95%, precision > 97%, recall > 95%, and F1-score > 96% across all tissue types. The flat fielding correction of bCT images resulted in a significant reduction in either cupping or capping artifacts in both mathematical and physical phantom studies as measured by the integral nonuniformity metric with an average reduction of 71% for cupping and 30% for capping across different phantom sizes, and the Uniformity Index with an average reduction of 53% for cupping and 34% for capping. CONCLUSION The validation studies demonstrated that the IC method improves Hounsfield Units (HU) accuracy and effectively corrects for shading artifacts caused by scatter contamination and beam hardening. The postprocessing approach described herein is relevant to the broad scope of bCT devices and does not require any modification in hardware or existing scan protocols. The trained CNN parameters and network architecture are available for interested users.
Collapse
Affiliation(s)
| | - Andrew M Hernandez
- Department of Radiology, University of California Davis, Sacramento, CA, 95817, USA
| | - Craig Abbey
- Department of Psychological & Brain Sciences, University of California Santa Barbara, Santa Barbara, CA, 93106, USA
| | - Kai Yang
- Department of Radiology, Massachusetts General Hospital, Boston, MA, 2114, USA
| | - John M Boone
- Department of Radiology, University of California Davis, Sacramento, CA, 95817, USA
| |
Collapse
|
9
|
Stayman JW, Capostagno S, Gang GJ, Siewerdsen JH. Task-driven source-detector trajectories in cone-beam computed tomography: I. Theory and methods. J Med Imaging (Bellingham) 2019; 6:025002. [PMID: 31065569 PMCID: PMC6497008 DOI: 10.1117/1.jmi.6.2.025002] [Citation(s) in RCA: 25] [Impact Index Per Article: 4.2] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/12/2018] [Accepted: 03/29/2019] [Indexed: 11/14/2022] Open
Abstract
We develop a mathematical framework for the design of orbital trajectories that are optimal to a particular imaging task (or tasks) in advanced cone-beam computed tomography systems that have the capability of general source-detector positioning. The framework allows various parameterizations of the orbit as well as constraints based on imaging system capabilities. To accommodate nonstandard system geometries, a model-based iterative reconstruction method is applied. Such algorithms generally complicate the assessment and prediction of reconstructed image properties; however, we leverage efficient implementations of analytical predictors of local noise and spatial resolution that incorporate dependencies of the reconstruction algorithm on patient anatomy, x-ray technique, and geometry. These image property predictors serve as inputs to a task-based performance metric defined by detectability index, which is optimized with respect to the orbital parameters of data acquisition. We investigate the framework of the task-driven trajectory design in several examples to examine the dependence of optimal source-detector trajectories on the imaging task (or tasks), including location and spatial-frequency dependence. A variety of multitask objectives are also investigated, and the advantages to imaging performance are quantified in simulation studies.
Collapse
Affiliation(s)
- J. Webster Stayman
- Johns Hopkins University, Department of Biomedical Engineering, Baltimore, Maryland, United States
| | - Sarah Capostagno
- Johns Hopkins University, Department of Biomedical Engineering, Baltimore, Maryland, United States
| | - Grace J. Gang
- Johns Hopkins University, Department of Biomedical Engineering, Baltimore, Maryland, United States
| | - Jeffrey H. Siewerdsen
- Johns Hopkins University, Department of Biomedical Engineering, Baltimore, Maryland, United States
- Johns Hopkins University, Department of Radiology and Radiological Science, Baltimore, Maryland, United States
| |
Collapse
|
10
|
Raylman RR, Van Kampen W, Stolin AV, Gong W, Jaliparthi G, Martone PF, Smith MF, Sarment D, Clinthorne NH, Perna M. A dedicated breast-PET/CT scanner: Evaluation of basic performance characteristics. Med Phys 2018; 45:1603-1613. [PMID: 29389017 DOI: 10.1002/mp.12780] [Citation(s) in RCA: 8] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/25/2017] [Revised: 01/05/2018] [Accepted: 01/05/2018] [Indexed: 12/29/2022] Open
Abstract
PURPOSE Application of advanced imaging techniques, such as PET and x ray CT, can potentially improve detection of breast cancer. Unfortunately, both modalities have challenges in the detection of some lesions. The combination of the two techniques, however, could potentially lead to an overall improvement in diagnostic breast imaging. The purpose of this investigation is to test the basic performance of a new dedicated breast-PET/CT. METHODS The PET component consists of a rotating pair of detectors. Its performance was evaluated using the NEMA NU4-2008 protocols. The CT component utilizes a pulsed x ray source and flat panel detector mounted on the same gantry as the PET scanner. Its performance was assessed using specialized phantoms. The radiation dose to a breast during CT imaging was explored by the measurement of free-in-air kerma and air kerma measured at the center of a 16 cm-diameter PMMA cylinder. Finally, the combined capabilities of the system were demonstrated by imaging of a micro-hot-rod phantom. RESULTS Overall, performance of the PET component is comparable to many pre-clinical and other dedicated breast-PET scanners. Its spatial resolution is 2.2 mm, 5 mm from the center of the scanner using images created with the single-sliced-filtered-backprojection algorithm. Peak NECR is 24.6 kcps; peak sensitivity is 1.36%; the scatter fraction is 27%. Spatial resolution of the CT scanner is 1.1 lp/mm at 10% MTF. The free-in-air kerma is 2.33 mGy, while the PMMA-air kerma is 1.24 mGy. Finally, combined imaging of a micro-hot-rod phantom illustrated the potential utility of the dual-modality images produced by the system. CONCLUSION The basic performance characteristics of a new dedicated breast-PET/CT scanner are good, demonstrating that its performance is similar to current dedicated PET and CT scanners. The potential value of this system is the capability to produce combined duality-modality images that could improve detection of breast disease. The next stage in development of this system is testing with more advanced phantoms and human subjects.
Collapse
Affiliation(s)
- Raymond R Raylman
- Center for Advanced Imaging, Department of Radiology, 1 Medical Center Dr., West Virginia University, Morgantown, WV, 26506, USA
| | - Will Van Kampen
- Xoran Technologies Inc., 5210 S State Rd., Ann Arbor, MI, 48108, USA
| | - Alexander V Stolin
- Center for Advanced Imaging, Department of Radiology, 1 Medical Center Dr., West Virginia University, Morgantown, WV, 26506, USA
| | - Wenbo Gong
- Xoran Technologies Inc., 5210 S State Rd., Ann Arbor, MI, 48108, USA
| | - Gangadhar Jaliparthi
- Center for Advanced Imaging, Department of Radiology, 1 Medical Center Dr., West Virginia University, Morgantown, WV, 26506, USA
| | - Peter F Martone
- Center for Advanced Imaging, Department of Radiology, 1 Medical Center Dr., West Virginia University, Morgantown, WV, 26506, USA
| | - Mark F Smith
- Department of Diagnostic Radiology and Nuclear Medicine, University of Maryland School of Medicine, 655 W. Baltimore Street, Baltimore, MD, 21201, USA
| | - David Sarment
- Xoran Technologies Inc., 5210 S State Rd., Ann Arbor, MI, 48108, USA
| | | | - Mark Perna
- Perna Health Physics, Inc., 705 Augusta Dr, Bridgeville, PA, 15017, USA
| |
Collapse
|
11
|
Shah JP, Mann SD, McKinley RL, Tornai MP. Characterization of CT Hounsfield Units for 3D acquisition trajectories on a dedicated breast CT system. JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY 2018; 26:535-551. [PMID: 29689765 PMCID: PMC6102078 DOI: 10.3233/xst-17350] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/10/2023]
Abstract
Hounsfield Units (HU) are used clinically in differentiating tissue types in a reconstructed CT image, and therefore the HU accuracy of a system is important, especially when using multiple sources, novel detector and non-traditional trajectories. Dedicated clinical breast CT (BCT) systems therefore should be similarly evaluated. In this study, uniform cylindrical phantoms filled with various uniform density fluids were used to characterize differences in HU values between simple circular and complex 3D (saddle) orbits. Based on ACR recommendations, the HU accuracy, center-to-edge variability within a slice, and overall variability within the reconstructed volume were characterized for simple and complex acquisitions possible on a single versatile BCT system. Results illustrate the statistically significantly better performance of the saddle orbit, especially close to the chest and nipple regions of what would clinically be a pendant breast volume. The incomplete cone beam acquisition of a simple circular orbit causes shading artifacts near the nipple, due to insufficient sampling, rendering a major portion of the scanned phantom unusable, whereas the saddle orbit performs exceptionally well and provides a tighter distribution of HU values throughout the reconstructed volumes. This study further establishes the advantages of using 3D acquisition trajectories for breast CT as well as other applications by demonstrating the robustness of HU values throughout large reconstructed volumes.
Collapse
Affiliation(s)
- Jainil P. Shah
- CivaTech Oncology, Durham, NC, USA
- Multi Modality Imaging Lab, Dept. of Radiology, Duke University Medical Center, Durham, NC, USA
- Department of Biomedical Engineering, Duke University, Durham, NC, USA
- Corresponding author: Jainil P. Shah,
| | - Steve D. Mann
- Clinical Imaging Physics Group, Duke University Health System, Durham, NC, USA
| | | | - Martin P. Tornai
- Multi Modality Imaging Lab, Dept. of Radiology, Duke University Medical Center, Durham, NC, USA
- Department of Biomedical Engineering, Duke University, Durham, NC, USA
- Medical Physics Graduate Program, Duke University Medical Center, Durham, NC, USA
| |
Collapse
|
12
|
Ge G, Zhang J, Winkler M, Lumby C, Cong W, Wang G. Clinical validation of CT image reconstruction with interior tomography. JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY 2018; 26:303-309. [PMID: 29562569 DOI: 10.3233/xst-17329] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/08/2023]
Abstract
Active x-ray collimation is well adopted in radiography and fluoroscopy for radiation dose reduction and image quality improvement. The application of this concept in computed tomography (CT) is significantly limited due to the truncation of projection data. Generally, an internal field of view (FOV) inside an imaging object cannot be exactly reconstructed only from the truncated projection data. Recent research shows that given some prior information of the FOV image, interior tomography can provide a unique and stable solution for image reconstruction of an internal FOV. The objective of this study is to evaluate the performance of interior reconstruction based on patient datasets obtained from a clinical CT scanner with dual x-ray tubes, which simultaneously gives full projections and truncated projections. Image reconstructions are performed from full and truncated projection data for the comparison of image quality, respectively. The reconstructed CT images were reviewed by a radiologist and a resident. The evaluation results of two observers showed that CT images reconstructed with truncated projections met clinically diagnostic requirements and were comparable to clinical images. This study demonstrates that with the development of interior tomography, active x-ray collimation in the imaging plane can be readily employed in CT imaging to further reduce patient radiation and improve image quality.
Collapse
Affiliation(s)
- Gary Ge
- Department of Radiation Medicine, University of Kentucky, Lexington, KY, USA
| | - Jie Zhang
- Department of Radiology, University of Kentucky, Lexington, KY, USA
| | - Michael Winkler
- Department of Radiology, University of Kentucky, Lexington, KY, USA
| | - Cynthia Lumby
- Department of Radiology, University of Kentucky, Lexington, KY, USA
| | - Wenxiang Cong
- Biomedical Imaging Center, Rensselaer Polytechnic Institute, Troy, NY, USA
| | - Ge Wang
- Biomedical Imaging Center, Rensselaer Polytechnic Institute, Troy, NY, USA
| |
Collapse
|
13
|
Mettivier G, Costa M, Lanconelli N, Ianiro A, Pugliese M, Quarto M, Russo P. Evaluation of Dose Homogeneity in Cone-Beam Breast Computed Tomography. RADIATION PROTECTION DOSIMETRY 2017; 175:473-481. [PMID: 28074018 DOI: 10.1093/rpd/ncw375] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 07/26/2016] [Accepted: 12/05/2016] [Indexed: 06/06/2023]
Abstract
The aim of this study is to determine, via measurements on phantoms and Monte Carlo (MC) simulations, the dose distribution of absorbed dose in a cone-beam breast computed tomography scan. The absorbed dose volume distribution was measured inside a polyethylene cylindrical phantom, simulating adipose breast tissue, using LiF:Mg,Ti thermoluminescence dosimeters. A reasonable agreement (between 2 and 8%) between the simulated and measured data was observed. The 3D distribution of absorbed dose was evaluated at 40, 60 and 80 kV in a phantom simulating a pendant breast. MC simulations indicate a significantly lower spread of volume dose than in mammography. The dose variation along the radial distance in the simulated phantoms was in the range of 4‒14%. These findings might be useful when devising models for breast imaging dose assessment that take into account the uneven distribution of the glandular mass in the breast volume.
Collapse
Affiliation(s)
- Giovanni Mettivier
- Dipartimento di Fisica 'Ettore Pancini', Università di Napoli Federico II, Napoli, Italy
- Istituto Nazionale di Fisica Nucleare, Sezione di Napoli, Napoli, Italy
| | - Matilde Costa
- Dipartimento di Fisica, Università di Trieste, Trieste, Italy
- Alma Mater Studiorum, Dipartimento di Fisica e Astronomia, Università di Bologna, Bologna, Italy
| | - Nico Lanconelli
- Alma Mater Studiorum, Dipartimento di Fisica e Astronomia, Università di Bologna, Bologna, Italy
- Istituto Nazionale di Fisica Nucleare, Sezione di Bologna, Bologna, Italy
| | - Anna Ianiro
- Dipartimento di Fisica 'Ettore Pancini', Università di Napoli Federico II, Napoli, Italy
| | - Mariagabriella Pugliese
- Dipartimento di Fisica 'Ettore Pancini', Università di Napoli Federico II, Napoli, Italy
- Istituto Nazionale di Fisica Nucleare, Sezione di Napoli, Napoli, Italy
| | - Maria Quarto
- Dipartimento di Fisica 'Ettore Pancini', Università di Napoli Federico II, Napoli, Italy
- Istituto Nazionale di Fisica Nucleare, Sezione di Napoli, Napoli, Italy
| | - Paolo Russo
- Dipartimento di Fisica 'Ettore Pancini', Università di Napoli Federico II, Napoli, Italy
- Istituto Nazionale di Fisica Nucleare, Sezione di Napoli, Napoli, Italy
| |
Collapse
|
14
|
Shah JP, Mann SD, McKinley RL, Tornai MP. Implementation and CT sampling characterization of a third-generation SPECT-CT system for dedicated breast imaging. J Med Imaging (Bellingham) 2017; 4:033502. [PMID: 28924570 PMCID: PMC5536183 DOI: 10.1117/1.jmi.4.3.033502] [Citation(s) in RCA: 9] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/19/2017] [Accepted: 07/14/2017] [Indexed: 11/14/2022] Open
Abstract
Stand-alone cone beam computed tomography (CT) and single-photon emission computed tomography (SPECT) systems capable of complex acquisition trajectories have previously been developed for breast imaging. Fully three-dimensional (3-D) motions of SPECT systems provide views into the chest wall and throughout the entire volume. The polar tilting capability of the CBCT system has shown improvement in sampling close to the chest wall, while eliminating cone beam artifacts. Here, a single hybrid SPECT-CT system, with each individual modality capable of independently traversing complex trajectories around a common pendant breast volume, was developed. We present the practical implementation of this design and preliminary results of the CT system. The fully 3-D SPECT was nested inside the suspended CT gantry and oriented perpendicular to the CT source-detector pair. Both subsystems were positioned on a rotation stage, with the combined polar and azimuthal motions enabling spherical trajectories. Six trajectories were used for initial evaluation of the tilt capable CT system. The developed system can achieve polar tilt angles with a [Formula: see text] positioning error and no hysteresis. Initial imaging results demonstrate that additional off-axis projection views of various geometric resolution phantoms facilitate more complete sampling, more consistent attenuation value recovery, and markedly improved reconstructions. This system could have various applications in diagnostic or therapeutic breast imaging.
Collapse
Affiliation(s)
- Jainil P. Shah
- Duke University, Department of Biomedical Engineering, Durham, North Carolina, United States
- Duke University Medical Center, Multi Modality Imaging Lab, Department of Radiology, Durham, North Carolina, United States
| | - Steve D. Mann
- Duke University Medical Center, Multi Modality Imaging Lab, Department of Radiology, Durham, North Carolina, United States
- Duke University Medical Center, Medical Physics Graduate Program, Durham, North Carolina, United States
| | | | - Martin P. Tornai
- Duke University, Department of Biomedical Engineering, Durham, North Carolina, United States
- Duke University Medical Center, Multi Modality Imaging Lab, Department of Radiology, Durham, North Carolina, United States
- Duke University Medical Center, Medical Physics Graduate Program, Durham, North Carolina, United States
| |
Collapse
|