Fujinaga Y, Kitou Y, Ohya A, Adachi Y, Tamaru N, Shiobara A, Ueda H, Nickel MD, Maruyama K, Kadoya M. Advantages of radial volumetric breath-hold examination (VIBE) with k-space weighted image contrast reconstruction (KWIC) over Cartesian VIBE in liver imaging of volunteers simulating inadequate or no breath-holding ability.
Eur Radiol 2015;
26:2790-7. [PMID:
26601972 DOI:
10.1007/s00330-015-4103-7]
[Citation(s) in RCA: 12] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/16/2015] [Revised: 11/03/2015] [Accepted: 11/10/2015] [Indexed: 11/30/2022]
Abstract
OBJECTIVES
To investigate the superiority of radial volumetric breath-hold examination (r-VIBE) with k-space weighted image contrast reconstruction (KWIC) over Cartesian VIBE (c-VIBE) for reducing motion artefacts.
METHODS
We acquired r-VIBE-KWIC and c-VIBE images in 10 healthy volunteers. Each acquisition lasted 24 seconds. The volunteers held their breath for decreasing lengths of time during the acquisitions, from 24 to 0 seconds (protocols A-E). Magnetic resonance images at the level of the right portal vein and confluence of hepatic veins were assessed by two readers using a five-point scale with a higher number indicating a better study.
RESULTS
The mean scores for the complete r-VIBE-KWIC series (r-VIBEfull) and first r-VIBE-KWIC series (r-VIBE1) were not significantly lower than those for c-VIBE in any protocols. The mean scores for c-VIBE were lower than those for r-VIBEfull and r-VIBE1 in protocols C and D. The mean score for c-VIBE was lower than that for r-VIBEfull in protocol E. The mean score for the eighth r-VIBE-KWIC series (r-VIBE8) was lower than that for c-VIBE only in protocol B.
CONCLUSION
r-VIBE-KWIC minimised artefacts relative to c-VIBE at any slice location. The r-VIBE-KWIC's sub-frame images during the breath-holding period were hardly affected by another failed breath-holding period.
KEY POINTS
• A two-reader study revealed r-VIBE-KWIC's advantages over c-VIBE • The image quality of r-VIBE-KWIC's sub-frame images was maintained during breath holding • Full-frame r-VIBE-KWIC images minimized motion artefacts caused by breathing • A complete breath holding over half the acquisition time is recommended for c-VIBE • c-VIBE was susceptible to respiratory motion especially in the subphrenic region.
Collapse