1
|
Adenis V, Partouche E, Stahl P, Gnansia D, Huetz C, Edeline JM. Asymmetric pulses delivered by a cochlear implant allow a reduction in evoked firing rate and in spatial activation in the guinea pig auditory cortex. Hear Res 2024; 447:109027. [PMID: 38723386 DOI: 10.1016/j.heares.2024.109027] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 01/22/2024] [Revised: 04/30/2024] [Accepted: 05/02/2024] [Indexed: 05/25/2024]
Abstract
Despite that fact that the cochlear implant (CI) is one of the most successful neuro-prosthetic devices which allows hearing restoration, several aspects still need to be improved. Interactions between stimulating electrodes through current spread occurring within the cochlea drastically limit the number of discriminable frequency channels and thus can ultimately result in poor speech perception. One potential solution relies on the use of new pulse shapes, such as asymmetric pulses, which can potentially reduce the current spread within the cochlea. The present study characterized the impact of changing electrical pulse shapes from the standard biphasic symmetric to the asymmetrical shape by quantifying the evoked firing rate and the spatial activation in the guinea pig primary auditory cortex (A1). At a fixed charge, the firing rate and the spatial activation in A1 decreased by 15 to 25 % when asymmetric pulses were used to activate the auditory nerve fibers, suggesting a potential reduction of the spread of excitation inside the cochlea. A strong "polarity-order" effect was found as the reduction was more pronounced when the first phase of the pulse was cathodic with high amplitude. These results suggest that the use of asymmetrical pulse shapes in clinical settings can potentially reduce the channel interactions in CI users.
Collapse
Affiliation(s)
- V Adenis
- Paris-Saclay Institute of Neurosciences (Neuro-PSI), France; CNRS UMR 9197, 91405 Orsay cedex, France; Université Paris-Saclay, 91405 Orsay cedex, France
| | - E Partouche
- Paris-Saclay Institute of Neurosciences (Neuro-PSI), France; CNRS UMR 9197, 91405 Orsay cedex, France; Université Paris-Saclay, 91405 Orsay cedex, France
| | - P Stahl
- Oticon Medical, Vallauris, France
| | | | - C Huetz
- Paris-Saclay Institute of Neurosciences (Neuro-PSI), France; CNRS UMR 9197, 91405 Orsay cedex, France; Université Paris-Saclay, 91405 Orsay cedex, France
| | - J-M Edeline
- Paris-Saclay Institute of Neurosciences (Neuro-PSI), France; CNRS UMR 9197, 91405 Orsay cedex, France; Université Paris-Saclay, 91405 Orsay cedex, France.
| |
Collapse
|
2
|
Guérit F, Middlebrooks JC, Gransier R, Richardson ML, Wouters J, Carlyon RP. Exploring the Use of Interleaved Stimuli to Measure Cochlear-Implant Excitation Patterns. J Assoc Res Otolaryngol 2024; 25:201-213. [PMID: 38459245 PMCID: PMC11018570 DOI: 10.1007/s10162-024-00937-2] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/01/2023] [Accepted: 02/15/2024] [Indexed: 03/10/2024] Open
Abstract
PURPOSE Attempts to use current-focussing strategies with cochlear implants (CI) to reduce neural spread-of-excitation have met with only mixed success in human studies, in contrast to promising results in animal studies. Although this discrepancy could stem from between-species anatomical and aetiological differences, the masking experiments used in human studies may be insufficiently sensitive to differences in excitation-pattern width. METHODS We used an interleaved-masking method to measure psychophysical excitation patterns in seven participants with four masker stimulation configurations: monopolar (MP), partial tripolar (pTP), a wider partial tripolar (pTP + 2), and, importantly, a condition (RP + 2) designed to produce a broader excitation pattern than MP. The probe was always in partial-tripolar configuration. RESULTS We found a significant effect of stimulation configuration on both the amount of on-site masking (mask and probe on same electrode; an indirect indicator of sharpness) and the difference between off-site and on-site masking. Differences were driven solely by RP + 2 producing a broader excitation pattern than the other configurations, whereas monopolar and the two current-focussing configurations did not statistically differ from each other. CONCLUSION A method that is sensitive enough to reveal a modest broadening in RP + 2 showed no evidence for sharpening with focussed stimulation. We also showed that although voltage recordings from the implant accurately predicted a broadening of the psychophysical excitation patterns with RP + 2, they wrongly predicted a strong sharpening with pTP + 2. We additionally argue, based on our recent research, that the interleaved-masking method can usefully be applied to non-human species and objective measures of CI excitation patterns.
Collapse
Affiliation(s)
- François Guérit
- Cambridge Hearing Group, MRC Cognition & Brain Sciences Unit, University of Cambridge, Cambridge, England.
| | - John C Middlebrooks
- Department of Otolaryngology, University of California at Irvine, Irvine, CA, USA
- Department of Neurobiology and Behavior, University of California at Irvine, Irvine, CA, USA
- Department of Biomedical Engineering, University of California at Irvine, Irvine, CA, USA
| | - Robin Gransier
- Department of Neurosciences, ExpORL KU Leuven, Leuven, Belgium
- Leuven Brain Institute KU Leuven, Leuven, Belgium
| | - Matthew L Richardson
- Department of Otolaryngology, University of California at Irvine, Irvine, CA, USA
| | - Jan Wouters
- Department of Neurosciences, ExpORL KU Leuven, Leuven, Belgium
- Leuven Brain Institute KU Leuven, Leuven, Belgium
| | - Robert P Carlyon
- Cambridge Hearing Group, MRC Cognition & Brain Sciences Unit, University of Cambridge, Cambridge, England
| |
Collapse
|
3
|
Guérit F, Deeks JM, Arzounian D, Gransier R, Wouters J, Carlyon RP. Using Interleaved Stimulation and EEG to Measure Temporal Smoothing and Growth of the Sustained Neural Response to Cochlear-Implant Stimulation. J Assoc Res Otolaryngol 2023; 24:253-264. [PMID: 36754938 PMCID: PMC10121955 DOI: 10.1007/s10162-023-00886-2] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/18/2022] [Accepted: 01/04/2023] [Indexed: 02/10/2023] Open
Abstract
Two EEG experiments measured the sustained neural response to amplitude-modulated (AM) high-rate pulse trains presented to a single cochlear-implant (CI) electrode. Stimuli consisted of two interleaved pulse trains with AM rates F1 and F2 close to 80 and 120 Hz respectively, and where F2 = 1.5F1. Following Carlyon et al. (J Assoc Res Otolaryngol, 2021), we assume that such stimuli can produce a neural distortion response (NDR) at F0 = F2-F1 Hz if temporal dependencies ("smoothing") in the auditory system are followed by one or more neural nonlinearities. In experiment 1, the rate of each pulse train was 480 pps and the gap between pulses in the F1 and F2 pulse trains ranged from 0 to 984 µs. The NDR had a roughly constant amplitude for gaps between 0 and about 200-400 µs, and decreased for longer gaps. We argue that this result is consistent with a temporal dependency, such as facilitation, operating at the level of the auditory nerve and/or with co-incidence detection by cochlear-nucleus neurons. Experiment 2 first measured the NDR for stimuli at each listener's most comfortable level ("MCL") and for F0 = 37, 40, and 43 Hz. This revealed a group delay of about 42 ms, consistent with a thalamic/cortical source. We then showed that the NDR grew steeply with stimulus amplitude and, for most listeners, decreased by more than 12 dB between MCL and 75% of the listener's dynamic range. We argue that the NDR is a potentially useful objective estimate of MCL.
Collapse
Affiliation(s)
- François Guérit
- Cambridge Hearing Group, MRC Cognition & Brain Sciences Unit, University of Cambridge, Cambridge, England
| | - John M Deeks
- Cambridge Hearing Group, MRC Cognition & Brain Sciences Unit, University of Cambridge, Cambridge, England
| | - Dorothée Arzounian
- Cambridge Hearing Group, MRC Cognition & Brain Sciences Unit, University of Cambridge, Cambridge, England
| | - Robin Gransier
- ExpORL, Dept. of Neurosciences, KU Leuven, Leuven, Belgium
| | - Jan Wouters
- ExpORL, Dept. of Neurosciences, KU Leuven, Leuven, Belgium
| | - Robert P Carlyon
- Cambridge Hearing Group, MRC Cognition & Brain Sciences Unit, University of Cambridge, Cambridge, England.
| |
Collapse
|
4
|
Monaghan JJM, Carlyon RP, Deeks JM. Modulation Depth Discrimination by Cochlear Implant Users. J Assoc Res Otolaryngol 2022; 23:285-299. [PMID: 35080684 PMCID: PMC8964891 DOI: 10.1007/s10162-022-00834-6] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/23/2021] [Accepted: 12/30/2021] [Indexed: 11/29/2022] Open
Abstract
Cochlear implants (CIs) convey the amplitude envelope of speech by modulating high-rate pulse trains. However, not all of the envelope may be necessary to perceive amplitude modulations (AMs); the effective envelope depth may be limited by forward and backward masking from the envelope peaks. Three experiments used modulated pulse trains to measure which portions of the envelope can be effectively processed by CI users as a function of AM frequency. Experiment 1 used a three-interval forced-choice task to test the ability of CI users to discriminate less-modulated pulse trains from a fully modulated standard, without controlling for loudness. The stimuli in experiment 2 were identical, but a two-interval task was used in which participants were required to choose the less-modulated interval, ignoring loudness. Catch trials, in which judgements based on level or modulation depth would give opposing answers, were included. Experiment 3 employed novel stimuli whose modulation envelope could be modified below a variable point in the dynamic range, without changing the loudness of the stimulus. Overall, results showed that substantial portions of the envelope are not accurately encoded by CI users. In experiment 1, where loudness cues were available, participants on average were insensitive to changes in the bottom 30% of their dynamic range. In experiment 2, where loudness was controlled, participants appeared insensitive to changes in the bottom 50% of the dynamic range. In experiment 3, participants were insensitive to changes in the bottom 80% of the dynamic range. We discuss potential reasons for this insensitivity and implications for CI speech-processing strategies.
Collapse
Affiliation(s)
- Jessica J M Monaghan
- Macquarie University, The Australian Hearing Hub, NSW, 2109, Sydney, Australia.
- National Acoustic Laboratories, The Australian Hearing Hub, Sydney, NSW, 2109, Australia.
| | - Robert P Carlyon
- Cambridge Hearing Group, Medical Research Council Cognition and Brain Sciences Unit, University of Cambridge, 15 Chaucer Road, Cambridge, CB2 7EF, UK
| | - John M Deeks
- Cambridge Hearing Group, Medical Research Council Cognition and Brain Sciences Unit, University of Cambridge, 15 Chaucer Road, Cambridge, CB2 7EF, UK
| |
Collapse
|
5
|
Heshmat A, Sajedi S, Schrott-Fischer A, Rattay F. Polarity Sensitivity of Human Auditory Nerve Fibers Based on Pulse Shape, Cochlear Implant Stimulation Strategy and Array. Front Neurosci 2021; 15:751599. [PMID: 34955717 PMCID: PMC8692583 DOI: 10.3389/fnins.2021.751599] [Citation(s) in RCA: 3] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/01/2021] [Accepted: 11/08/2021] [Indexed: 11/28/2022] Open
Abstract
Neural health is of great interest to determine individual degeneration patterns for improving speech perception in cochlear implant (CI) users. Therefore, in recent years, several studies tried to identify and quantify neural survival in CI users. Among all proposed techniques, polarity sensitivity is a promising way to evaluate the neural status of auditory nerve fibers (ANFs) in CI users. Nevertheless, investigating neural health based on polarity sensitivity is a challenging and complicated task that involves various parameters, and the outcomes of many studies show contradictory results of polarity sensitivity behavior. Our computational study benefits from an accurate three-dimensional finite element model of a human cochlea with realistic human ANFs and determined ANF degeneration pattern of peripheral part with a diminishing of axon diameter and myelination thickness based on degeneration levels. In order to see how different parameters may impact the polarity sensitivity behavior of ANFs, we investigated polarity behavior under the application of symmetric and asymmetric pulse shapes, monopolar and multipolar CI stimulation strategies, and a perimodiolar and lateral CI array system. Our main findings are as follows: (1) action potential (AP) initiation sites occurred mainly in the peripheral site in the lateral system regardless of stimulation strategies, pulse polarities, pulse shapes, cochlear turns, and ANF degeneration levels. However, in the perimodiolar system, AP initiation sites varied between peripheral and central processes, depending on stimulation strategies, pulse shapes, and pulse polarities. (2) In perimodiolar array, clusters formed in threshold values based on cochlear turns and degeneration levels for multipolar strategies only when asymmetric pulses were applied. (3) In the perimodiolar array, a declining trend in polarity (anodic threshold/cathodic threshold) with multipolar strategies was observed between intact or slight degenerated cases and more severe degenerated cases, whereas in the lateral array, cathodic sensitivity was noticed for intact and less degenerated cases and anodic sensitivity for cases with high degrees of degeneration. Our results suggest that a combination of asymmetric pulse shapes, focusing more on multipolar stimulation strategies, as well as considering the distances to the modiolus wall, allows us to distinguish the degeneration patterns of ANFs across the cochlea.
Collapse
Affiliation(s)
- Amirreza Heshmat
- Institute for Analysis and Scientific Computing, Vienna University of Technology, Vienna, Austria.,Laboratory for Inner Ear Biology, Department of Otorhinolaryngology, Medical University of Innsbruck, Innsbruck, Austria
| | - Sogand Sajedi
- Institute for Analysis and Scientific Computing, Vienna University of Technology, Vienna, Austria
| | - Anneliese Schrott-Fischer
- Laboratory for Inner Ear Biology, Department of Otorhinolaryngology, Medical University of Innsbruck, Innsbruck, Austria
| | - Frank Rattay
- Institute for Analysis and Scientific Computing, Vienna University of Technology, Vienna, Austria
| |
Collapse
|
6
|
Navntoft CA, Landsberger DM, Barkat TR, Marozeau J. The Perception of Ramped Pulse Shapes in Cochlear Implant Users. Trends Hear 2021; 25:23312165211061116. [PMID: 34935552 PMCID: PMC8724057 DOI: 10.1177/23312165211061116] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/30/2022] Open
Abstract
The electric stimulation provided by current cochlear implants (CI) is not power
efficient. One underlying problem is the poor efficiency by which information
from electric pulses is transformed into auditory nerve responses. A novel
stimulation paradigm using ramped pulse shapes has recently been proposed to
remedy this inefficiency. The primary motivation is a better biophysical fit to
spiral ganglion neurons with ramped pulses compared to the rectangular pulses
used in most contemporary CIs. Here, we tested the hypotheses that ramped pulses
provide more efficient stimulation compared to rectangular pulses and that a
rising ramp is more efficient than a declining ramp. Rectangular, rising ramped
and declining ramped pulse shapes were compared in terms of charge efficiency
and discriminability, and threshold variability in seven CI listeners. The tasks
included single-channel threshold detection, loudness-balancing, discrimination
of pulse shapes, and threshold measurement across the electrode array. Results
showed that reduced charge, but increased peak current amplitudes, was required
at threshold and most comfortable levels with ramped pulses relative to
rectangular pulses. Furthermore, only one subject could reliably discriminate
between equally-loud ramped and rectangular pulses, suggesting variations in
neural activation patterns between pulse shapes in that participant. No
significant difference was found between rising and declining ramped pulses
across all tests. In summary, the present findings show some benefits of charge
efficiency with ramped pulses relative to rectangular pulses, that the direction
of a ramped slope is of less importance, and that most participants could not
perceive a difference between pulse shapes.
Collapse
Affiliation(s)
- Charlotte Amalie Navntoft
- Hearing Systems Group, Department of Health Technology, 5205Technical University of Denmark, Kgs. Lyngby, Denmark.,Brain and Sound Lab, Department of Biomedicine, 27209Basel University, Basel, Switzerland
| | - David M Landsberger
- Department of Otolaryngology, 12296New York University School of Medicine, New York, USA
| | - Tania Rinaldi Barkat
- Brain and Sound Lab, Department of Biomedicine, 27209Basel University, Basel, Switzerland
| | - Jeremy Marozeau
- Hearing Systems Group, Department of Health Technology, 5205Technical University of Denmark, Kgs. Lyngby, Denmark
| |
Collapse
|
7
|
The Effects of GJB2 or SLC26A4 Gene Mutations on Neural Response of the Electrically Stimulated Auditory Nerve in Children. Ear Hear 2021; 41:194-207. [PMID: 31124793 DOI: 10.1097/aud.0000000000000744] [Citation(s) in RCA: 16] [Impact Index Per Article: 5.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/26/2022]
Abstract
OBJECTIVES This study aimed to (1) investigate the effect of GJB2 and SLC26A4 gene mutations on auditory nerve function in pediatric cochlear implant users and (2) compare their results with those measured in implanted children with idiopathic hearing loss. DESIGN Participants included 20 children with biallelic GJB2 mutations, 16 children with biallelic SLC26A4 mutations, and 19 children with idiopathic hearing loss. All subjects except for two in the SLC26A4 group had concurrent Mondini malformation and enlarged vestibular aqueduct. All subjects used Cochlear Nucleus devices in their test ears. For each subject, electrophysiological measures of the electrically evoked compound action potential (eCAP) were recorded using both anodic- and cathodic-leading biphasic pulses. Dependent variables (DVs) of interest included slope of eCAP input/output (I/O) function, the eCAP threshold, and eCAP amplitude measured at the maximum comfortable level (C level) of the anodic-leading stimulus (i.e., the anodic C level). Slopes of eCAP I/O functions were estimated using statistical modeling with a linear regression function. These DVs were measured at three electrode locations across the electrode array. Generalized linear mixed effect models were used to evaluate the effects of study group, stimulus polarity, and electrode location on each DV. RESULTS Steeper slopes of eCAP I/O function, lower eCAP thresholds, and larger eCAP amplitude at the anodic C level were measured for the anodic-leading stimulus compared with the cathodic-leading stimulus in all subject groups. Children with GJB2 mutations showed steeper slopes of eCAP I/O function and larger eCAP amplitudes at the anodic C level than children with SLC26A4 mutations and children with idiopathic hearing loss for both the anodic- and cathodic-leading stimuli. In addition, children with GJB2 mutations showed a smaller increase in eCAP amplitude when the stimulus changed from the cathodic-leading pulse to the anodic-leading pulse (i.e., smaller polarity effect) than children with idiopathic hearing loss. There was no statistically significant difference in slope of eCAP I/O function, eCAP amplitude at the anodic C level, or the size of polarity effect on all three DVs between children with SLC26A4 mutations and children with idiopathic hearing loss. These results suggested that better auditory nerve function was associated with GJB2 but not with SLC26A4 mutations when compared with idiopathic hearing loss. In addition, significant effects of electrode location were observed for slope of eCAP I/O function and the eCAP threshold. CONCLUSIONS GJB2 and SLC26A4 gene mutations did not alter polarity sensitivity of auditory nerve fibers to electrical stimulation. The anodic-leading stimulus was generally more effective in activating auditory nerve fibers than the cathodic-leading stimulus, despite the presence of GJB2 or SLC26A4 mutations. Patients with GJB2 mutations appeared to have better functional status of the auditory nerve than patients with SLC26A4 mutations who had concurrent Mondini malformation and enlarged vestibular aqueduct and patients with idiopathic hearing loss.
Collapse
|
8
|
Xu L, Skidmore J, Luo J, Chao X, Wang R, Wang H, He S. The Effect of Pulse Polarity on Neural Response of the Electrically Stimulated Cochlear Nerve in Children With Cochlear Nerve Deficiency and Children With Normal-Sized Cochlear Nerves. Ear Hear 2021; 41:1306-1319. [PMID: 32141933 PMCID: PMC7879579 DOI: 10.1097/aud.0000000000000854] [Citation(s) in RCA: 14] [Impact Index Per Article: 4.7] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/26/2022]
Abstract
OBJECTIVE This study aimed to (1) investigate the effect of pulse polarity on neural response of the electrically stimulated cochlear nerve in children with cochlear nerve deficiency (CND) and children with normal-sized cochlear nerves and (2) compare the size of the pulse polarity effect between these two subject groups. DESIGN The experimental and control group included 31 children with CND and 31 children with normal-sized cochlear nerves, respectively. For each study participant, evoked compound action potential (eCAP) input/output (I/O) functions for anodic-leading and cathodic-leading biphasic stimuli were measured at three electrode locations across the electrode array. The dependent variables of interest included the eCAP amplitude measured at the maximum comfortable level of the anodic stimulus, the lowest level that could evoke an eCAP (i.e., the eCAP threshold), the slope of the eCAP I/O function estimated based on linear regression, the negative-peak (i.e., N1) latency of the eCAP, as well as the size of the pulse polarity effect on these eCAP measurements. Generalized linear mixed effect models were used to compare the eCAP amplitude, the eCAP threshold, the slope of the eCAP I/O function, and the N1 latency evoked by the anodic-leading stimulus with those measured for the cathodic-leading stimulus for children with CND and children with normal-sized cochlear nerves. Generalized linear mixed effect models were also used to compare the size of the pulse polarity effect on the eCAP between these two study groups. The one-tailed Spearman correlation test was used to assess the potential correlation between the pulse phase duration and the difference in N1 latency measured for different pulse polarities. RESULTS Compared with children who had normal-sized cochlear nerves, children with CND had reduced eCAP amplitudes, elevated eCAP thresholds, flatter eCAP I/O functions, and prolonged N1 latencies. The anodic-leading stimulus led to higher eCAP amplitudes, lower eCAP thresholds, and shorter N1 latencies than the cathodic-leading stimulus in both study groups. Steeper eCAP I/O functions were recorded for the anodic-leading stimulus than those measured for the cathodic-leading stimulus in children with CND, but not in children with normal-sized cochlear nerves. Group differences in the size of the pulse polarity effect on the eCAP amplitude, the eCAP threshold, or the N1 latency were not statistically significant. CONCLUSIONS Similar to the normal-sized cochlear nerve, the hypoplastic cochlear nerve is more sensitive to the anodic-leading than to the cathodic-leading stimulus. Results of this study do not provide sufficient evidence for proving the idea that the pulse polarity effect can provide an indication for local neural health.
Collapse
Affiliation(s)
- Lei Xu
- Department of Auditory Implantation, Shandong Provincial ENT Hospital Affiliated to Shandong University, Duanxing W. Rd, Huaiyin, Jinan, Shandong Province, P.R. China 250022
| | - Jeffrey Skidmore
- Department of Otolaryngology – Head and Neck Surgery, The Ohio State University, 915 Olentangy River Road, Columbus, OH 43212
| | - Jianfen Luo
- Department of Auditory Implantation, Shandong Provincial ENT Hospital Affiliated to Shandong University, Duanxing W. Rd, Huaiyin, Jinan, Shandong Province, P.R. China 250022
| | - Xiuhua Chao
- Department of Auditory Implantation, Shandong Provincial ENT Hospital Affiliated to Shandong University, Duanxing W. Rd, Huaiyin, Jinan, Shandong Province, P.R. China 250022
| | - Ruijie Wang
- Department of Auditory Implantation, Shandong Provincial ENT Hospital Affiliated to Shandong University, Duanxing W. Rd, Huaiyin, Jinan, Shandong Province, P.R. China 250022
| | - Haibo Wang
- Department of Auditory Implantation, Shandong Provincial ENT Hospital Affiliated to Shandong University, Duanxing W. Rd, Huaiyin, Jinan, Shandong Province, P.R. China 250022
| | - Shuman He
- Department of Otolaryngology – Head and Neck Surgery, The Ohio State University, 915 Olentangy River Road, Columbus, OH 43212
- Department of Audiology, Nationwide Children’s Hospital, 700 Children’s Drive, Columbus, OH 43205
| |
Collapse
|
9
|
Carlyon RP, Guérit F, Deeks JM, Harland A, Gransier R, Wouters J, de Rijk SR, Bance M. Using Interleaved Stimulation to Measure the Size and Selectivity of the Sustained Phase-Locked Neural Response to Cochlear Implant Stimulation. J Assoc Res Otolaryngol 2021; 22:141-159. [PMID: 33492562 PMCID: PMC7943679 DOI: 10.1007/s10162-020-00783-y] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/26/2020] [Accepted: 12/21/2020] [Indexed: 02/08/2023] Open
Abstract
We measured the sustained neural response to electrical stimulation by a cochlear implant (CI). To do so, we interleaved two stimuli with frequencies F1 and F2 Hz and recorded a neural distortion response (NDR) at F2-F1 Hz. We show that, because any one time point contains only the F1 or F2 stimulus, the instantaneous nonlinearities typical of electrical artefact should not produce distortion at this frequency. However, if the stimulus is smoothed, such as by charge integration at the nerve membrane, subsequent (neural) nonlinearities can produce a component at F2-F1 Hz. We stimulated a single CI electrode with interleaved sinusoids or interleaved amplitude-modulated pulse trains such that F2 = 1.5F1, and found no evidence for an NDR when F2-F1 was between 90 and 120 Hz. However, interleaved amplitude-modulated pulse trains with F2-F1~40 Hz revealed a substantial NDR with a group delay of about 45 ms, consistent with a thalamic and/or cortical response. The NDR could be measured even from recording electrodes adjacent to the implant and at the highest pulse rates (> 4000 pps) used clinically. We then measured the selectivity of this sustained response by presenting F1 and F2 to different electrodes and at different between-electrode distances. This revealed a broad tuning that, we argue, reflects the overlap between the excitation elicited by the two electrodes. Our results also provide a glimpse of the neural nonlinearity in the auditory system, unaffected by the biomechanical cochlear nonlinearities that accompany acoustic stimulation. Several potential clinical applications of our findings are discussed.
Collapse
Affiliation(s)
- Robert P Carlyon
- Cambridge Hearing Group, MRC Cognition & Brain Sciences Unit, University of Cambridge, 15 Chaucer Rd, Cambridge, CB2 7EF, England.
| | - François Guérit
- Cambridge Hearing Group, MRC Cognition & Brain Sciences Unit, University of Cambridge, 15 Chaucer Rd, Cambridge, CB2 7EF, England
| | - John M Deeks
- Cambridge Hearing Group, MRC Cognition & Brain Sciences Unit, University of Cambridge, 15 Chaucer Rd, Cambridge, CB2 7EF, England
| | - Andrew Harland
- Cambridge Hearing Group, MRC Cognition & Brain Sciences Unit, University of Cambridge, 15 Chaucer Rd, Cambridge, CB2 7EF, England
| | - Robin Gransier
- Dept. of Neurosciences, ExpORL, KU Leuven, Herestraat 49 box 721, 3000, Leuven, Belgium
| | - Jan Wouters
- Dept. of Neurosciences, ExpORL, KU Leuven, Herestraat 49 box 721, 3000, Leuven, Belgium
| | - Simone R de Rijk
- Cambridge Hearing Group, Dept. Clinical Neurosciences, University of Cambridge, Cambridge Biomedical Campus, Cambridge, CB2 0QQ, England
| | - Manohar Bance
- Cambridge Hearing Group, Dept. Clinical Neurosciences, University of Cambridge, Cambridge Biomedical Campus, Cambridge, CB2 0QQ, England
| |
Collapse
|
10
|
Guérit F, Marozeau J, Epp B, Carlyon RP. Effect of the Relative Timing between Same-Polarity Pulses on Thresholds and Loudness in Cochlear Implant Users. J Assoc Res Otolaryngol 2020; 21:497-510. [PMID: 32833160 PMCID: PMC7644659 DOI: 10.1007/s10162-020-00767-y] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/06/2020] [Accepted: 07/31/2020] [Indexed: 12/21/2022] Open
Abstract
The effect of the relative timing between pairs of same-polarity monophasic pulses has been studied extensively in single-neuron animal studies and has revealed fundamental properties of the neurons. For human cochlear implant listeners, the requirement to use charge-balanced stimulation and the typical use of symmetric, biphasic pulses limits such measures, because currents of opposite polarities interact at the level of the neural membrane. Here, we propose a paradigm to study same-polarity summation of currents while keeping the stimulation charge-balanced within a short time window. We used pairs of mirrored pseudo-monophasic pulses (a long-low phase followed by a short-high phase for the first pulse and a short-high phase followed by a long-low phase for the second pulse). We assumed that most of the excitation would stem from the two adjacent short-high phases, which had the same polarity. The inter-pulse interval between the short-high phases was varied from 0 to 345 μs. The inter-pulse interval had a significant effect on the perceived loudness, and this effect was consistent with both passive (membrane-related) and active (ion-channel-related) neuronal mechanisms contributing to facilitation. Furthermore, the effect of interval interacted with the polarity of the pulse pairs. At threshold, there was an effect of polarity, but, surprisingly, no effect of interval nor an interaction between the two factors. We discuss possible peripheral origins of these results.
Collapse
Affiliation(s)
- François Guérit
- Hearing Systems Group, Department of Health Technology, Technical University of Denmark, 352 Ørsteds Plads, 2800, Kgs. Lyngby, Denmark.
- Cambridge Hearing Group, MRC Cognition and Brain Sciences Unit, University of Cambridge, 15 Chaucer Road, Cambridge, CB2 7EF, United Kingdom.
| | - Jeremy Marozeau
- Hearing Systems Group, Department of Health Technology, Technical University of Denmark, 352 Ørsteds Plads, 2800, Kgs. Lyngby, Denmark
| | - Bastian Epp
- Hearing Systems Group, Department of Health Technology, Technical University of Denmark, 352 Ørsteds Plads, 2800, Kgs. Lyngby, Denmark
| | - Robert P Carlyon
- Cambridge Hearing Group, MRC Cognition and Brain Sciences Unit, University of Cambridge, 15 Chaucer Road, Cambridge, CB2 7EF, United Kingdom
| |
Collapse
|
11
|
Konerding W, Arenberg JG, Kral A, Baumhoff P. Late electrically-evoked compound action potentials as markers for acute micro-lesions of spiral ganglion neurons. Hear Res 2020; 413:108057. [PMID: 32883545 DOI: 10.1016/j.heares.2020.108057] [Citation(s) in RCA: 7] [Impact Index Per Article: 1.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 03/24/2020] [Revised: 07/13/2020] [Accepted: 08/11/2020] [Indexed: 12/20/2022]
Abstract
Cochlear implants (CIs) are the treatment of choice for profoundly hearing impaired people. It has been proposed that speech perception in CI users is influenced by the neural health (deafferentation, demyelination and degeneration) of the cochlea, which may be heterogeneous along an individual cochlea. Several options have been put forward to account for these local differences in neural health when fitting the speech processor settings, however with mixed results. The interpretation of the results is hampered by the fact that reliable markers of locally restricted changes in spiral ganglion neuron (SGN) health are lacking. The aim of the study was (i) to establish mechanical micro-lesions in the guinea pig as a model of heterogeneous SGN deafferentation and degeneration and (ii) to assess potential electrophysiological markers that can also be used in human subjects. First, we defined the extent of micro-lesions in normal hearing animals using acoustically-evoked compound action potentials (aCAPs); second, we measured electrically-evoked CAPs (eCAPs) before and after focal lesioning in neomycin-deafened and implanted animals. Therefore, we inserted guinea pig adjusted 6-contact CIs through a cochleostomy in the scala tympani. The eCAP was recorded from a ball electrode at the round window niche in response to monopolar or bipolar, 50 µs/phase biphasic pulses of alternating anodic- and cathodic-leading polarity. To exclude the large electrical artifact from the analysis, we focused on the late eCAP component. We systematically isolated the eCAP parameter that showed local pre- versus post-lesion changes and lesion-target specificity. Histological evaluation of the cleared cochleae revealed focal damage of an average size of 0.0036 mm3 with an apical-basal span of maximal 440 µm. We found that the threshold of the late N2P2 eCAP component was significantly elevated after lesioning when stimulating at basal (near the lesion), but not apical (distant to the lesion) CI contacts. To circumvent the potentially conflicting influence of the apical-basal gradient in eCAP thresholds, we used the polarity effect (PE=cathodic-anodic) as a relative measure. During monopolar stimulation, but not bipolar stimulation, the PE was sensitive to the lesion target and showed significantly better cathodic than anodic thresholds after soma lesions. We conclude that the difference in N2P2 thresholds in response to cathodic versus anodic-leading monopolar stimulation corresponds to the presence of SGN soma damage, and may therefore be a marker for SGN loss. We consider this electrophysiological estimate of local neural health a potentially relevant tool for human applications because of the temporal separation from the stimulation artifact and possible implementation into common eCAP measurements.
Collapse
Affiliation(s)
- Wiebke Konerding
- Department of Experimental Otology, Hannover Medical School, Stadtfelddamm 34, 30625 Hannover, Germany.
| | - Julie G Arenberg
- Massachusetts Eye and Ear Infirmary, Harvard Medical School, 25 Shattuck Street, Boston, MA 02115, USA.
| | - Andrej Kral
- Department of Experimental Otology, Hannover Medical School, Stadtfelddamm 34, 30625 Hannover, Germany; Cluster of Excellence "Hearing4all", Germany.
| | - Peter Baumhoff
- Department of Experimental Otology, Hannover Medical School, Stadtfelddamm 34, 30625 Hannover, Germany.
| |
Collapse
|