Kuschel CA, Bach KP, Webster NJ, Page B, Groves AM, Battin MR. The reliability of 2D and colour Doppler ultrasound in localising longline position.
J Paediatr Child Health 2008;
44:483-7. [PMID:
18557803 DOI:
10.1111/j.1440-1754.2008.01333.x]
[Citation(s) in RCA: 8] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 11/29/2022]
Abstract
BACKGROUND
The position of percutaneously inserted central venous catheters (longlines) in neonates is critical, as malpositioned longlines are associated with potentially fatal complications.
AIM
To determine if cardiac ultrasound (two-dimensional (2D) and colour Doppler) is useful in evaluating longline position, when compared with the position identified by contrast radiography.
SETTING
Single level 3 neonatal unit.
PARTICIPANTS
Forty-four neonates undergoing insertion of 24-gauge silastic longlines between July 2004 and September 2005.
METHODS
Infants who had a longline inserted underwent echocardiography by a novice and an experienced operator. Operators identified longline position using a 2D then colour Doppler echocardiography during a rapid bolus infusion of saline. The position was identified from contrast radiography by two independent observers.
RESULTS
Using 2D echocardiography, the novice and experienced operators could identify 41 and 59% of longlines, respectively. However, only 34% of longlines were identified by both operators. In 15 infants whose longline positions were identified by both operators, there was agreement in only eight infants (53%). Colour Doppler improved the experienced operator's success but did not assist the novice operator. For radiographs, there was 68% agreement on longline position between observers. The experienced echocardiographer located three (7%) longlines within the heart that from radiographs were thought to be in a proximal central vessel.
CONCLUSIONS
This technique is experience-dependent and complements rather than replaces the use of contrast radiography. However, some infants with an apparently acceptable longline position on contrast radiography have longlines located within the heart on echocardiography.
Collapse