Alvi SN, dgither SA, Al-Odaib A. Method Development and Clinical Utility for Simultaneous Measurement of 21-Deoxycortisol, 17-Hydroxyprogesterone, Cortisol, and Cortisone Levels in Human Plasma Using UHPLC-MS/MS.
Adv Pharmacol Pharm Sci 2025;
2025:3859670. [PMID:
39866748 PMCID:
PMC11759567 DOI:
10.1155/adpp/3859670]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/28/2024] [Accepted: 12/30/2024] [Indexed: 01/28/2025] Open
Abstract
A simple and efficient validated assay for quantifying 21-deoxycortisol (21-DOC), 17-hydroxyprogesterone (17-OHP), cortisol, and cortisone in human plasma has been developed using ultra-high performance liquid chromatography coupled with tandem mass spectrometry (UHPLC-MS/MS). Analysis of plasma samples were performed on Atlantis dC18 (3 μm) column using a mobile phase of 20.0 mM ammonium acetate and acetonitrile (50:50, v : v) that was delivered at isocratic flow rate 0.3 mL/minute. After addition of d4-cortisol as an internal standard (IS), plasma samples containing 21-DOC, 17-OHP, cortisol, and cortisone were extracted with mixture of dichloromethane and tert-butylmethyl ether 1:2 (v/v). Analytes were detected and quantified in the positive ion mode of electrospray ionization using multiple reaction monitoring transition set at mass to charge (m/z): 347.17 ⟶ 311.12, 331.17 ⟶ 96.97, 363.11 ⟶ 121.00, 361.18 ⟶ 163.11, and 367.19 ⟶ 121.24 for 21-DOC and 17-OHP, cortisol, cortisone, and cortisol-d4 (IS), respectively. The relationship between concentration and peak area response (analyte/IS) were linear over the range of 0.25-50, 0.5-100, 1-200, and 2-400 ng/mL for 21-DOC, 17-OHP, cortisone, and cortisol, respectively. The mean extraction recovery of the analytes was in the range of 83%-96%. The coefficient of variation within and between days was less than 13.6%, and the bias ranged from -9.2% to 12%. The measured level of cortisol, cortisone, and 17-OHP was in the range of 21.9-110, 4.33-12.71, and 0.37-1.4 ng/mL, respectively. Furthermore, the measured value of cortisone-cortisol ratio was in the range of 0.08-0.21.
Collapse