Shermock KM, Streiff MB, Pinto BL, Kraus P, Pronovost PJ. Novel analysis of clinically relevant diagnostic errors in point-of-care devices.
J Thromb Haemost 2011;
9:1769-75. [PMID:
21752186 DOI:
10.1111/j.1538-7836.2011.04439.x]
[Citation(s) in RCA: 11] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 08/31/2023]
Abstract
BACKGROUND
To ensure proper clinical decision-making and avoid preventable harm, the quality of point-of-care (POC) device measures is routinely assessed. Traditional analyses may not reveal clinically important diagnostic errors.
OBJECTIVES
To compare results between a novel analytic framework and traditional analyses.
METHODS
Patients in four anticoagulation clinics provided two measures of the International Normalized Ratio (INR) at the same visit as part of routine quality assurance: one via a venous sample and one fingerstick. These were assessed with Hemochron POC devices. Traditional, quarterly, quality assurance assessments emphasized correlation analysis. The novel analysis used enhanced graphics and a validated assessment of clinical decision-making.
RESULTS
1518 paired INRs were analyzed. The correlation between the POC and laboratory assessments ranged between 0.84 and 0.91. Traditional quality assurance showed that the Hemochron devices were acceptable for continued use in each quarterly analysis. Enhanced graphical analysis demonstrated that the Hemochron devices never reported seven common INR values. The Hemochron devices systematically inflated values < 3 and deflated values > 4, biasing results towards the target INR range. Consequently, the Hemochron devices lead to a different clinical decision than the clinical laboratory measure in 31% of cases (458/1466; 95% confidence interval [CI] 29-34). When the reference INR was low, the Hemochron devices would not result in appropriate dose increases in 52% of cases (95% CI 48-56), placing these patients at risk for a significant adverse drug event.
CONCLUSIONS
Our novel, clinically relevant analysis revealed previously undetected deficiencies in our POC INR devices, and our approach should be adopted by industry, regulators, and institutions.