Barin ES, Cejnar M, Nelson GI, Hunyor SN. Physical characteristics and clinical evaluation of a new disposable fibreoptic transducer-tipped catheter system.
Anaesth Intensive Care 1987;
15:323-9. [PMID:
3661966 DOI:
10.1177/0310057x8701500313]
[Citation(s) in RCA: 4] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/06/2023]
Abstract
A new disposable fibreoptic transducer-tipped catheter manometer system was evaluated to assess its accuracy, stability of accuracy under prolonged simulated intra-arterial conditions, and dynamic characteristics. Maximum errors observed in the measurement of static pressure using a sample of five catheters (with one display unit) were 2 mmHg at 0 mmHg reference pressure, 2 at 20 mmHg, 4 at 40 mmHg, 4 at 100 mmHg and 9 at 200 mmHg. An immersion artifact caused a shift in baseline of up to 2 mmHg. Exposure of the transducer to 24 hours of simulated intra-arterial conditions (pulsatile pressure at 40 degrees C) resulted in errors of up to 7 mmHg for pressures up to 100 mmHg, and 11 mmHg for 200 mmHg, which were largely attributable to a drift in baseline pressure (up 6 mmHg by 24 hours). Consistent overestimation by the system suggested inappropriate gain setting within the display unit which, however, is not user-adjustable. The system exhibited uniform frequency response up to 33 Hz.